Quench protection design of a 9.4 T whole-body MRI superconducting magnet

被引:5
|
作者
Chen, Shunzhong [1 ]
Li, Yi [1 ]
Dai, Yinming [1 ]
Lei, Yuanzhong [1 ]
Yan, Luguang [1 ]
机构
[1] Chinese Acad Sci, Inst Elect Engn, Key Lab Appl Superconduct, Beijing 100190, Peoples R China
关键词
MRI; Quench; Protection; Heater;
D O I
10.1016/j.physc.2013.11.001
中图分类号
O59 [应用物理学];
学科分类号
摘要
A 9.4 T MRI superconducting magnet with a 800 mm clear warm bore in diameter is designed and fabricated for bioscience research. The superconducting magnet consisting of five coaxial solenoid coils is fabricated with NbTi Wire-in-Channel (WIC) conductor where the ratios of copper to non-copper are from 5 to 10. The four compensation solenoid coils are with rectangular NbTi/Cu strand wires. The magnet will be operated in a relative low nominal current of 224.5 A with a high level of stored energy, about 138 MJ. A protection method with the cold diodes and resistors in series across the subdivided sections and active trigger heater to accelerate quench is adopted to avoid the damage of the magnet. In the paper, the quench simulation results of currents, voltages and hot-spot temperatures based on the protection scheme are analyzed in details. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:49 / 53
页数:5
相关论文
共 50 条
  • [31] Quench Protection Design and Simulation of a 7 Tesla Superconducting Magnet
    Guo, Beilei
    Ma, Lizhen
    Wu, Wei
    Yao, Qinggao
    Huang, Wenxue
    Wu, Xi
    He, Yuan
    Han, Shaofei
    Yuan, Ping
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 2087 - 2090
  • [32] SUPERCONDUCTING MAGNET QUENCH PROTECTION FOR ISABELLE
    ROBINS, KE
    SAMPSON, WB
    THOMAS, MG
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (03) : 1318 - 1319
  • [33] SUPERCONDUCTING SWITCHES AND JOINTS IN 1-T WHOLE-BODY MRI MAGNETS
    LIN, MH
    ZHOU, XT
    CRYOGENICS, 1990, 30 : 626 - 629
  • [34] The Iseult/Inumac whole body 11.7 T MRI magnet design
    Schild, Th.
    Aubert, G.
    Berriaud, C.
    Bredy, Ph.
    Juster, F. P.
    Meuris, C.
    Nunio, F.
    Quettier, L.
    Rey, J. M.
    Vedrine, P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2008, 18 (02) : 904 - 907
  • [35] Mechanical Design of the Iseult 11.7 T Whole Body MRI Magnet
    Nunio, F.
    Berriaud, C.
    Bredy, Ph.
    Schild, Th.
    Scola, L.
    Tellier, O.
    Vedrine, P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 760 - 763
  • [36] Design and characteristic analysis of active quench protection system for superconducting magnet
    Chu, Yong
    Kim, Ho Min
    Jang, Mi Hye
    Kim, Kee Man
    Ko, Tae Kuk
    Physica C: Superconductivity and its Applications, 2000, 341-348
  • [37] Design and characteristic analysis of active quench protection system for superconducting magnet
    Chu, Y
    Kim, HM
    Jang, MH
    Kim, KM
    Ko, TK
    PHYSICA C, 2000, 341 : 2605 - 2606
  • [38] SERIES SUPERCONDUCTING MAGNET QUENCH PROTECTION FOR ISABELLE
    ROBINS, KE
    SAMPSON, WB
    THOMAS, MG
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1977, 22 (02): : 139 - 140
  • [39] Monofilament MgB2 Wire for a Whole-Body MRI Magnet: Superconducting Joints and Test Coils
    Ling, Jiayin
    Voccio, John
    Kim, Youngjae
    Hahn, Seungyong
    Bascunan, Juan
    Park, Dongkeun K.
    Iwasa, Yukikazu
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)
  • [40] Design and Quench Protection Analysis of a 9 T NbTi Superconducting Magnet for a Large-Bore EMPS
    Jang, Jae Young
    Hwang, Young Jin
    Kim, Myung Su
    Lee, Wooseung
    Choi, Yeon Suk
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2020, 30 (04)