Quench protection design of a 9.4 T whole-body MRI superconducting magnet

被引:5
|
作者
Chen, Shunzhong [1 ]
Li, Yi [1 ]
Dai, Yinming [1 ]
Lei, Yuanzhong [1 ]
Yan, Luguang [1 ]
机构
[1] Chinese Acad Sci, Inst Elect Engn, Key Lab Appl Superconduct, Beijing 100190, Peoples R China
关键词
MRI; Quench; Protection; Heater;
D O I
10.1016/j.physc.2013.11.001
中图分类号
O59 [应用物理学];
学科分类号
摘要
A 9.4 T MRI superconducting magnet with a 800 mm clear warm bore in diameter is designed and fabricated for bioscience research. The superconducting magnet consisting of five coaxial solenoid coils is fabricated with NbTi Wire-in-Channel (WIC) conductor where the ratios of copper to non-copper are from 5 to 10. The four compensation solenoid coils are with rectangular NbTi/Cu strand wires. The magnet will be operated in a relative low nominal current of 224.5 A with a high level of stored energy, about 138 MJ. A protection method with the cold diodes and resistors in series across the subdivided sections and active trigger heater to accelerate quench is adopted to avoid the damage of the magnet. In the paper, the quench simulation results of currents, voltages and hot-spot temperatures based on the protection scheme are analyzed in details. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:49 / 53
页数:5
相关论文
共 50 条
  • [21] SUPERCONDUCTING MAGNET QUENCH PROTECTION ANALYSIS AND DESIGN
    SALASOO, L
    IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 1908 - 1911
  • [22] Protection Circuit and Quench Simulation of a 7 T Superconducting Magnet for Animal MRI
    Chen, Jiangbo
    Jiang, Xiaohua
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)
  • [23] Globally Optimal Algorithm for Design of 0.7 T Actively Shielded Whole-Body Open MRI Superconducting Magnet System
    Ni, Zhipeng
    Hu, Geli
    Li, Lankai
    Yu, Gongqi
    Wang, Qiuliang
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)
  • [24] Optimization in design of actively shielded whole-body open high-field superconducting MRI magnet
    Feng Zhong-Kui
    Hu Ge-Li
    Xu Ying
    Zhu Guang
    Zhou Feng
    Dai Yin-Ming
    Wang Qiu-Liang
    ACTA PHYSICA SINICA, 2013, 62 (23)
  • [25] Quench detection and protection system design and analysis of the 7 T superconducting magnet
    龙风
    施毅
    刘方
    倪志鹏
    中国物理C, 2010, 34 (04) : 492 - 495
  • [26] Quench detection and protection system design and analysis of the 7 T superconducting magnet
    Long Feng
    Shi Yi
    Liu Fang
    Ni Zhi-Peng
    CHINESE PHYSICS C, 2010, 34 (04) : 492 - 495
  • [27] Analysis of Damage by Quench and Improvements in Rewinding for a 9.4-T Superconducting NMR Magnet
    Li, Yi
    Wang, Qiuliang
    Song, Shousen
    Lei, Yuanzhong
    Cheng, Junsheng
    Chen, Shunzhong
    Zhao, Baozhi
    Dai, Yinming
    Ni, Zhipeng
    Hu, Xinning
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (01)
  • [28] Thermal and Electrical Behaviour of the Persistent Current Switch for a Whole-Body Superconducting MRI Magnet
    Nandawadekar, Ajit
    Soni, V
    Suman, N.
    Ram, Sankar T.
    Kumar, R.
    Saini, S. K.
    Sharma, R. G.
    Singh, Mukhtiar
    Kar, Soumen
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2021, 31 (05)
  • [29] Decoupling Design of Z2 Superconducting Shim Coils for 9.4-T MRI Superconducting Magnet
    Li, Yi
    Wang, Qiuliang
    Zhu, Xuchen
    Ni, Zhipeng
    Wang, Lei
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)
  • [30] A Superconducting Magnet System for Whole-Body Metabolism Imaging
    Wang, Q.
    Dai, Y.
    Zhao, B.
    Song, S.
    Wang, C.
    Li, L.
    Cheng, J.
    Chen, S.
    Wang, H.
    Ni, Z.
    Li, Y.
    Cui, C.
    Hu, X.
    Wang, H.
    Lei, Y.
    Chan, K.
    Yan, L.
    Wen, C.
    Hui, G.
    Yang, W.
    Liu, F.
    Zhuo, Y.
    Zhou, X.
    Yan, Z.
    Chen, J.
    Xu, T.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)