Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisition

被引:112
|
作者
McKerracher, C [1 ]
Thwaites, DI [1 ]
机构
[1] Western Gen Hosp, Edinburgh EH4 2XU, Midlothian, Scotland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1999年 / 44卷 / 09期
关键词
D O I
10.1088/0031-9155/44/9/303
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Two new detectors (0.015 cm(3) ion chamber from PTW, 0.6 mm diameter diode from Scanditronix AB) designed specifically for use in small stereotactic fields were compared against similar, more routine, detectors (0.125 cm(3) ion chamber; parallel plate chamber, shielded and unshielded diodes and film). Percentage depth doses, tissue maximum ratios, off-axis ratios and relative output factors were compared for circular fields in the 40-12.5 mm diameter range, with a view to identifying the optimum detector for stereotactic beam data acquisition. Practical suggestions for beam data collection and analysis are made, with an emphasis on what is achievable practically in radiotherapy departments where the primary demand is to provide a routine service. No single detector was found to be ideal, and neither of the two new measurement devices had any significant advantages over more routine devices, in the situations measured. Although the new 0.015 cm(3) ion chamber was an improvement on a 0.125 cm(3) ion chamber in the measurement of profiles, it was still too large when compared with a diode. The new small diode had a low signal to noise ratio which made reliable data difficult to extract and its only advantage is possibly improved resolution in fields smaller than the range tested. The use of a larger unshielded diode is recommended for all measurements, with the additional cross-checking of data against at least one small ion chamber and film. A simple method of obtaining reliable output data from the detectors used is explained.
引用
收藏
页码:2143 / 2160
页数:18
相关论文
共 49 条
  • [41] A THEORETICAL COMPARISON OF THE DATA-ACQUISITION TIME CHARACTERISTICS OF THE TIME OF FLIGHT VOLTAGE CONTRAST DETECTOR WITH RETARDING-FIELD DETECTORS
    KHURSHEED, A
    DINNIS, AR
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1990, 8 (06): : 1841 - 1847
  • [42] Preliminary Evaluation of a Reference Detector Measuring Leakage Radiation for Use in Small Field Beam Data Acquisition
    Gopal, A.
    Laugeman, E.
    Kim, J.
    Dumas, M.
    Huang, Y.
    Wen, N.
    Shi, J.
    Simon, B.
    Chetty, I.
    MEDICAL PHYSICS, 2017, 44 (06) : 2950 - 2950
  • [43] A comparative study of small field total scatter factors and dose profiles using plastic scintillation detectors and other stereotactic dosimeters: The case of the CyberKnife
    Morin, Jonathan
    Beliveau-Nadeau, Dominic
    Chung, Eunah
    Seuntjens, Jan
    Theriault, Dany
    Archambault, Louis
    Beddar, Sam
    Beaulieu, Luc
    MEDICAL PHYSICS, 2013, 40 (01)
  • [44] Determination of small-field output factors for beam-matched linear accelerators using various detectors and comparison of detector-specific output correction factors using IAEA Technical Report Series 483 protocol
    Kannan, Mageshraja
    Saminathan, Sathiyan
    Chandraraj, Varatharaj
    Raj, D. Gowtham
    Ganesh, K. M.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2023, 28 (02) : 241 - 254
  • [45] A new reference-type ionization chamber with direction-independent response for use in small-field photon-beam dosimetry - An experimental and Monte Carlo study
    Delfs, Bjoern
    Kapsch, Ralf-Peter
    Chofor, Ndimofor
    Looe, Hui Khee
    Harder, Dietrich
    Poppe, Bjoern
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (01): : 39 - 48
  • [46] Detector Selection Impact on Small-Field Dosimetry of Collecting Beam Data Measurements among Versa HD 6 MV FFF Beams: A Multi-Institutional Variability Analysis
    Ceylan, Cemile
    Gungor, Boran
    Yesil, Abdullah
    Gungor, Sinem
    Inal, Serpil Yondem
    Kara, Esil
    Konuk, Aykut
    Ozden, Ozcan
    Kara, Ozgur
    Avci, Volkan
    Baser, Tamer
    Ugur, Timur
    Saglam, Yucel
    Can, Sumeyra
    Durmus, Ismail Faruk
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2024, 39 (01): : 1 - 18
  • [47] Comparison Between the HYPERSCINT RP200 Scintillation Detector and Other Small Field Detectors for 10MV FFF SRS Beam Modelling On a VersaHD Linear Accelerator
    Morin, J.
    Cabana, J. F.
    Goulet, M.
    Theriault, D.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [48] MEASURING PARAMETERS OF SLOW-NEUTRON FIELD IN CHANNEL OF A VVR-M-TYPE REACTOR BY MEANS OF DETECTORS CALIBRATED AGAINST STATE STANDARD FOR UNITS OF THERMAL-NEUTRON FLUX-DENSITY
    YARITSYNA, IA
    KUCHERYAVENKO, EP
    KUTEEVA, TM
    KHARITONOV, IA
    VASILEV, GY
    KONOPLEV, KA
    MEASUREMENT TECHNIQUES, 1978, 21 (01) : 114 - 118
  • [49] A new peak recognition algorithm for detection of ultra-small nano-particles by single particle ICP-MS using rapid time resolved data acquisition on a sector-field mass spectrometer
    Tuoriniemi, Jani
    Cornelis, Geert
    Hassellov, Martin
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (08) : 1723 - 1729