Functional MRI of the Lungs Using Single Breath-Hold and Self-Navigated Ultrashort Echo Time Sequences

被引:11
|
作者
Heidenreich, Julius F. [1 ]
Veldhoen, Simon [1 ]
Metz, Corona [1 ]
Pereira, Lenon Mendes [1 ]
Benkert, Thomas [2 ]
Pfeuffer, Josef [2 ]
Bley, Thorsten A. [1 ]
Kostler, Herbert [1 ]
Weng, Andreas M. [1 ]
机构
[1] Univ Hosp Wurzburg, Dept Diagnost & Intervent Radiol, Oberdurrbacher Str 6, D-97080 Wurzburg, Germany
[2] Siemens Healthcare, Dept Applicat Dev, Erlangen, Germany
来源
RADIOLOGY-CARDIOTHORACIC IMAGING | 2020年 / 2卷 / 03期
关键词
D O I
10.1148/ryct.2020190162
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate three-dimensional (3D) ultrashort echo time (UTE) MRI regarding image quality and suitability for functional image analysis using gradient-echo sequences in breath-hold and with self-navigation. Materials and Methods: In this prospective exploratory study, 10 patients (mean age, 21 years; age range, 558 years; five men) and 10 healthy control participants (mean age, 25 years; age range, 1039 years; five men) underwent 3D UTE MRI at 3.0 T. Imaging was performed with a prototypical stack-of-spirals 3D UTE sequence during single breath holds (echo time [TE], 0.05 msec) and with a self-navigated Koosh ball 3D UTE sequence at free breathing (TE, 0.03 msec). Image quality was rated on a four-point Likert scale. Edge sharpness was calculated. After semiautomated segmentation, fractional ventilation was calculated from MRI signal intensity (FVSI) and volume change (FVVol). The air volume fraction (AVF) was estimated from relative signal intensity (aortic blood signal intensity was used as a reference). Means were compared between techniques and participants. The Wilcoxon signed rank test and Spearman rank correlation were used for statistical analyses. Results: Image quality ratings were equal for both techniques. However, stack-of-spirals breath-hold UTE was more susceptible to motion and aliasing artifacts. Mean FVSI was higher during breath hold than at free breathing (mean +/- standard deviation in milliliters of gas per milliliters of parenchyma, 0.17 mL/mL +/- 0.06 [minimum, 0.07; maximum, 0.34] vs 0.11 mL/mL +/- 0.03 [minimum, 0.06; maximum, 0.17], P = .016). Mean FVSI and FVVol were in good agreement (mean difference: at breath hold, -0.008 [95% confidence interval {CI}: 0.007, -0.024]; rho = 0.97 vs free breathing, -0.004 [95% CI: 0.007, -0.016]; rho = 0.91). AVF correlated between both techniques (rho = 0.94). Conclusion: Breath-hold and self-navigated 3D UTE sequences yield proton densityweighted images of the lungs that are similar in quality, and both techniques are suitable for functional image analysis. Summary Three-dimensional ultrashort echo time gradient-echo sequences are promising tools for acquisition of proton density-weighted lung MR images within a single breath hold or at free breathing, providing high spatial resolution and allowing for functional image analysis such as ventilation or air volume fraction.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Analysis of a Deep Learning-Based Superresolution Algorithm Tailored to Partial Fourier Gradient Echo Sequences of the Abdomen at 1.5 T Reduction of Breath-Hold Time and Improvement of Image Quality
    Afat, Saif
    Wessling, Daniel
    Afat, Carmen
    Nickel, Dominik
    Arberet, Simon
    Herrmann, Judith
    Othman, Ahmed E.
    Gassenmaier, Sebastian
    INVESTIGATIVE RADIOLOGY, 2022, 57 (03) : 157 - 162
  • [42] Contrast-enhanced MRI of the menisci of the knee using ultrashort echo time (UTE) pulse sequences: imaging of the red and white zones
    Gatehouse, PD
    He, T
    Puri, BK
    Thomas, RD
    Resnick, D
    Bydder, GM
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (920): : 641 - 647
  • [43] Comparison between gradient and spin-echo (GRASE) and compressed sensing sequences for single breath-hold three-dimensional magnetic resonance cholangiopancreatography in patients with T1 hyperintense bile
    Morimoto-Ishikawa, Daisuke
    Hyodo, Tomoko
    Takenaka, Mamoru
    Matsukubo, Yuko
    Numoto, Isao
    Itoh, Makoto
    Ohmi, Masato
    Kamata, Ken
    Ueda, Yu
    Wakana, Miyuki
    Kudo, Masatoshi
    Saito, Shigeyoshi
    Ishii, Kazunari
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 150
  • [44] Three-dimensional Magnetic Resonance Imaging Using Single Breath-hold k-t BLAST for Assessment of Global Left Ventricular Functional Parameters
    Jaroni, Jennifer
    Meier, Reinhard
    Beer, Ambros
    Herrmann, Ken
    Settles, Marcus
    Rummeny, Ernst J.
    Huber, Armin
    ACADEMIC RADIOLOGY, 2013, 20 (08) : 987 - 994
  • [45] MR cholangiopancreatography (MRCP) using respiratory triggered three-dimensional turbo spin echo (3d TSE RT) and breath-hold single shot turbo spin echo sequences (SSTSE BH) before and after blueberry juice (BBJ) administration
    Papanikolaou, N
    Karantanas, AH
    Heracleous, H
    Gourtsoyiannis, NC
    RADIOLOGY, 1998, 209P : 484 - 484
  • [46] Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm
    Bernd J. Wintersperger
    Konstantin Nikolaou
    Olaf Dietrich
    Johannes Rieber
    Matthias Nittka
    Maximilian F. Reiser
    Stefan O. Schoenberg
    European Radiology, 2003, 13 : 1931 - 1936
  • [47] Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm
    Wintersperger, BJ
    Nikolaou, K
    Dietrich, O
    Rieber, J
    Nittka, M
    Reiser, MF
    Schoenberg, SO
    EUROPEAN RADIOLOGY, 2003, 13 (08) : 1931 - 1936
  • [48] Liver T2-weighted MR imaging: Comparison of fast and conventional half-Fourier single-shot turbo spin-echo, breath-hold turbo spin-echo, and respiratory-triggered turbo spin-echo sequences
    Tang, Y
    Yamashita, Y
    Namimoto, T
    Abe, Y
    Takahashi, M
    RADIOLOGY, 1997, 203 (03) : 766 - 772
  • [49] T2-weighted MR imaging for focal hepatic lesion detection: Supplementary value of breath-hold imaging with half-Fourier single-shot fast spin-echo and multishot spin-echo echoplanar sequences
    Matsuo, M
    Kanematsu, M
    Murakami, T
    Kim, T
    Hori, M
    Kondo, H
    Nakamura, H
    Hoshi, H
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 12 (03) : 444 - 452
  • [50] Qualitative and quantitative analysis of regional left ventricular wall dynamics using real-time magnetic resonance imaging: Comparison with conventional breath-hold gradient echo acquisition in volunteers and patients
    Plein, S
    Smith, WHT
    Ridgway, JP
    Kassner, A
    Beacock, DJ
    Bloomer, TN
    Sivananthan, MU
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (01) : 23 - 30