Evaluation of a HbA1c point-of-care analyzer

被引:7
|
作者
Menendez-Valladares, Paloma [1 ]
Fernandez-Riejos, Patricia [1 ]
Sanchez-Mora, Catalina [1 ]
Perez-Perez, Antonio [1 ]
Sanchez-Margalet, Victor [1 ]
Gonzalez-Rodriguez, Concepcion [1 ]
机构
[1] Virgen Macarena Univ Hosp, Dept Clin Biochem, Seville 41009, Spain
关键词
POCT; HbA1c; Diabetes; Evaluation; GLYCATED HEMOGLOBIN; INSTRUMENTS; CAPILLARY; HBA(1C);
D O I
10.1016/j.clinbiochem.2015.03.016
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: A better glycemic monitoring of diabetic patients and avoiding complications of poorly controlled diabetes could be possible with point-of-care testing technology (POCT) for HbA1c determination. B-Analyst (R) was studied to check whether it complied with the quality requirements for this purpose. Design and methods: We evaluated the B-Analyst (R) (Menarini Diagnostics), which is based in the principle of latex agglutination immunoturbidimetry, to assess the validity of the technique of HbA1c. We carried out the method comparison with the HA-8180 (R) (Menarini Diagnostics) as a reference method [High Performance Liquid Chromatography (HPLC)]. We assessed the analytical quality of the B-Analyst (R) studying the accuracy: inter-assay variability and intra-assay study. Furthermore, possible interferences by hemoglobinopathies were studied. Results: Regression analysis of the data for the method comparison between HA-8180 (R) and B-Analyst (R) showed a slope of 1.0085 and an intercept of 0.1208. The Pearson's correlation coefficient was 0.9958 (p < 0.0001). Bias study showed a mean difference from B-Analyst (R) with respect to HA-8180 (R) of 0.1872 with a 95% confidence interval. The standard error of the estimate (S-yx) was 0.2091. The concordance correlation coefficient to assess accuracy was 0.9922 (0.9891-0.9945). The CV for the inter-assay study was 1.4%. For the intra-assay study we analyzed 3 samples with different HbA1c % whose CV were 1.03% [4.7% HbA1c (28 mmol/mol)], 0.46% [6.4% HbA1c (46 mmol/mol)] and 0.78% [8.1% HbA1c (65 mmol/mol)]. Conclusion: The B-Analyst (R) evaluated not only showed good correlation with HA-8180 (R), but also it presented a great accuracy both in the inter-assay and in the intra-assay studies. The 8-Analyst (R) complies with quality specifications required for monitoring of diabetic patients. (C) 2015 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:686 / 689
页数:4
相关论文
共 50 条
  • [41] Evaluation of a new hemoglobin A1c analyzer for point-of-care testing
    Zhou, Rui
    Wang, Wei
    Song, Zhi-Xin
    Tong, Qing
    Wang, Qing-Tao
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (01)
  • [42] A new Point-of-Care Analyzer for Rapid and Precise Determination of Lipid Profile and HbA1c.
    Joergensen, B.
    Zidek, A. Sais
    CEREBROVASCULAR DISEASES, 2014, 37 : 343 - 343
  • [43] The value proposition for point-of-care testing in healthcare: HbA1c for monitoring in diabetes management as an exemplar
    Price, Christopher P.
    St John, Andrew
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2019, 79 (05): : 298 - 304
  • [44] Point-of-Care Amperometric Testing of Diabetic Marker ( HbA1c) Using Specific Electroactive Antibodies
    Chopra, Adity
    Rawat, Sanjay
    Bhalla, Vijayender
    Suri, C. Raman
    ELECTROANALYSIS, 2014, 26 (03) : 469 - 472
  • [45] Systematic Diabetes Screening Using Point-of-Care HbA1c Testing Facilitates Identification of Prediabetes
    Whitley, Heather P.
    Hanson, Courtney
    Parton, Jason M.
    ANNALS OF FAMILY MEDICINE, 2017, 15 (02) : 162 - 164
  • [46] Evaluation of the Afinion AS100 Point-of-Care Analyzer for Hemoglobin A(1c)
    Arabadjief, Melissa
    Nichols, James H.
    POINT OF CARE, 2009, 8 (01): : 11 - 15
  • [47] Evaluation of the Alere Afinion HbA1c Dx Test: Can Point of Care HbA1c Testing Deliver the Performance Needed for Diagnostic Use?
    Swensen, Monica H.
    Stivers, David
    Arnold, William
    Amar, Meera
    Horowitz, Barry
    Little, Randie
    DIABETES, 2016, 65 : A380 - A381
  • [48] Evaluation of the Siemens DCA Vantage hemoglobin A1c point-of-care analyzer
    Charnot-Katsikos, A.
    Hui, S. R.
    Yeo, K. T.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (03) : 481 - 481
  • [49] Assessment of an affinity based point of care system for HbA1c
    Mosca, Andrea
    Paleari, Renata
    Passerini, Gabriella
    Ceriotti, Ferruccio
    DIABETES, 2008, 57 : A119 - A119
  • [50] USE OF HBA1C POINT-OF-CARE DEVICE AT BASIC HEALTH UNITS IN BRAZIL, AS PART OF HEALTHRISE BRAZIL
    Nicolaevna Kochergin, C.
    Soares, D.
    Medeiros, D.
    Louzado Andrade, J.
    Silva, K.
    Oliveira, M.
    Barone, M.
    Cortes, M.
    Mistro, S.
    Bezerra, V.
    Amorim Wildes, W.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2019, 21 : A155 - A155