Evaluation of a HbA1c point-of-care analyzer

被引:7
|
作者
Menendez-Valladares, Paloma [1 ]
Fernandez-Riejos, Patricia [1 ]
Sanchez-Mora, Catalina [1 ]
Perez-Perez, Antonio [1 ]
Sanchez-Margalet, Victor [1 ]
Gonzalez-Rodriguez, Concepcion [1 ]
机构
[1] Virgen Macarena Univ Hosp, Dept Clin Biochem, Seville 41009, Spain
关键词
POCT; HbA1c; Diabetes; Evaluation; GLYCATED HEMOGLOBIN; INSTRUMENTS; CAPILLARY; HBA(1C);
D O I
10.1016/j.clinbiochem.2015.03.016
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: A better glycemic monitoring of diabetic patients and avoiding complications of poorly controlled diabetes could be possible with point-of-care testing technology (POCT) for HbA1c determination. B-Analyst (R) was studied to check whether it complied with the quality requirements for this purpose. Design and methods: We evaluated the B-Analyst (R) (Menarini Diagnostics), which is based in the principle of latex agglutination immunoturbidimetry, to assess the validity of the technique of HbA1c. We carried out the method comparison with the HA-8180 (R) (Menarini Diagnostics) as a reference method [High Performance Liquid Chromatography (HPLC)]. We assessed the analytical quality of the B-Analyst (R) studying the accuracy: inter-assay variability and intra-assay study. Furthermore, possible interferences by hemoglobinopathies were studied. Results: Regression analysis of the data for the method comparison between HA-8180 (R) and B-Analyst (R) showed a slope of 1.0085 and an intercept of 0.1208. The Pearson's correlation coefficient was 0.9958 (p < 0.0001). Bias study showed a mean difference from B-Analyst (R) with respect to HA-8180 (R) of 0.1872 with a 95% confidence interval. The standard error of the estimate (S-yx) was 0.2091. The concordance correlation coefficient to assess accuracy was 0.9922 (0.9891-0.9945). The CV for the inter-assay study was 1.4%. For the intra-assay study we analyzed 3 samples with different HbA1c % whose CV were 1.03% [4.7% HbA1c (28 mmol/mol)], 0.46% [6.4% HbA1c (46 mmol/mol)] and 0.78% [8.1% HbA1c (65 mmol/mol)]. Conclusion: The B-Analyst (R) evaluated not only showed good correlation with HA-8180 (R), but also it presented a great accuracy both in the inter-assay and in the intra-assay studies. The 8-Analyst (R) complies with quality specifications required for monitoring of diabetic patients. (C) 2015 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:686 / 689
页数:4
相关论文
共 50 条
  • [21] Reducing analytical variation between point-of-care and laboratory HbA1c testing
    Malkani, Samir
    Korpi-Steiner, Nichole
    Rao, Lokinendi V.
    JOURNAL OF DIABETES, 2013, 5 (02) : 192 - 196
  • [22] Point-of-care HbA1c Measurements in Oral Cancer and Control Patients in Hungary
    Vegh, Adam
    Vegh, Daniel
    Banyai, Dorottya
    Kammerhofer, Gabor
    Biczo, Zita
    Voros, Balazs
    Ujpal, Marta
    Francisco Pena-Cardelles, Juan
    Yonel, Zehra
    Joob-Fancsaly, Arpad
    Hermann, Peter
    Nemeth, Zsolt
    IN VIVO, 2022, 36 (05): : 2248 - 2254
  • [23] Comparison of point-of-care HbA1c test versus standardized laboratory testing
    Sicard, DA
    Taylor, JR
    ANNALS OF PHARMACOTHERAPY, 2005, 39 (06) : 1024 - 1028
  • [24] Different strategies for detection of HbA1c emphasizing on biosensors and point-of-care analyzers
    Kaur, Jagjit
    Jiang, Cheng
    Liu, Guozhen
    BIOSENSORS & BIOELECTRONICS, 2019, 123 : 85 - 100
  • [25] Analytical performance of point-of-care device QuikRead go in HbA1c analysis
    Linko-Parvinen, Anna
    Dahlstrom, Johanna
    Pallari, Hanna-Mari
    JOURNAL OF LABORATORY AND PRECISION MEDICINE, 2023, 8
  • [26] Evaluation of the In2it® analyzer for HbA1c determination
    Martin, M.
    Leroy, N.
    Sulmont, V.
    Gillery, P.
    DIABETES & METABOLISM, 2010, 36 (02) : 158 - 164
  • [27] Evaluation of the DCA Vantage analyzer for HbA1c assay
    Szymezak, Jean
    Leroy, Nathalie
    Lavalard, Emmanuelle
    Gillery, Philippe
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (08) : 1195 - 1198
  • [28] Analytical performance evaluation of the GreenCare A1c and Cera-Stat HbA1c point-of-care testing assays
    Lee, Joon Hee
    Jun, Sun-Hee
    Lee, Jikyo
    Song, Sang Hoon
    Lee, Kyunghoon
    ANNALS OF CLINICAL BIOCHEMISTRY, 2025, 62 (01) : 46 - 56
  • [29] HBA1C point-of-care testing for diabetes control in a low-income population: A before and after study and cost-parity analysis HbA1c point-of-care testing for diabetes control
    Andrade, Monica Viegas
    Noronha, Kenya Valeria Micaela de Souza
    Santos, Andre Soares
    Maia, Junia Xavier
    Nogueira, Lucas Tavares
    Cimini, Christiane Correa Rodrigues
    Furtado, Maria Eduarda
    Coelho, Laryssa
    Marcolino, Milena Soriano
    Ribeiro, Antonio Luiz Pinho
    PRIMARY CARE DIABETES, 2023, 17 (05) : 447 - 453
  • [30] Point-of-care testing of HbA1c and blood glucose in a remote Aboriginal Australian community
    Martin, DD
    Shephard, MDS
    Freeman, H
    Jones, TW
    Davis, EA
    Maguire, GP
    MEDICAL JOURNAL OF AUSTRALIA, 2005, 182 (10) : 524 - 527