Accuracy in contouring of small and low contrast lesions: Comparison between diagnostic quality computed tomography scanner and computed tomography simulation scanner-A phantom study

被引:2
|
作者
Ho, Yick Wing [1 ]
Wong, Wing Kei Rebecca [1 ]
Yu, Siu Ki [1 ]
Lam, Wai Wang [1 ]
Geng, Hui [1 ]
机构
[1] Hong Kong Sanat & Hosp, Med Phys & Res Dept, Happy Valley, Hong Kong, Peoples R China
关键词
Computed tomography; Delineation accuracy; RADIOSURGERY; RADIOTHERAPY; DELINEATION;
D O I
10.1016/j.meddos.2012.03.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
To evaluate the accuracy in detection of small and low-contrast regions using a high-definition diagnostic computed tomography (CT) scanner compared with a radiotherapy CT simulation scanner. A custom-made phantom with cylindrical holes of diameters ranging from 2-9 mm was filled with 9 different concentrations of contrast solution. The phantom was scanned using a 16-slice multidetector CT simulation scanner (LightSpeed RT16, General Electric Healthcare, Milwaukee, WI) and a 64-slice high-definition diagnostic CT scanner (Discovery CT750 HD, General Electric Healthcare). The low-contrast regions of interest (ROIs) were delineated automatically upon their full width at half maximum of the CT number profile in Hounsfield units on a treatment planning workstation. Two conformal indexes, CIin, and CIout, were calculated to represent the percentage errors of underestimation and overestimation in the automated contours compared with their actual sizes. Summarizing the conformal indexes of different sizes and contrast concentration, the means of CIin, and CIout for the CT simulation scanner were 33.7% and 60.9%, respectively, and 10.5% and 41.5% were found for the diagnostic CT scanner. The mean differences between the 2 scanners' CIin, and CIout were shown to be significant with p < 0.001. A descending trend of the index values was observed as the ROI size increases for both scanners, which indicates an improved accuracy when the ROI size increases, whereas no observable trend was found in the contouring accuracy with respect to the contrast levels in this study. Images acquired by the diagnostic CT scanner allow higher accuracy on size estimation compared with the CT simulation scanner in this study. We recommend using a diagnostic CT scanner to scan patients with small lesions (<1 cm in diameter) for radiotherapy treatment planning, especially for those pending for stereotactic radiosurgery in which accurate delineation of small-sized, low-contrast regions is important for dose calculation. (C) 2012 American Association of Medical Dosimetrists.
引用
收藏
页码:401 / 405
页数:5
相关论文
共 50 条
  • [31] Comparison of defect detectability between Computed Tomography inspection and CT simulation using a calibrated defect phantom
    Galleguillos, Carlos
    Senck, Sascha
    Vilches, Celia
    Rendl, Sarah
    Lasagni, Fernando
    e-Journal of Nondestructive Testing, 2022, 27 (03):
  • [32] FOCAL LIVER-LESIONS - COMPARATIVE-STUDY COMPUTED-TOMOGRAPHY, CONVENTIONAL ULTRASOUND, AUTOMATED WATERPATH SCANNER (OCTOSON)
    FIEGLER, W
    SCHUTZ, H
    CLAUSSEN, C
    SCHORNER, W
    FELIX, R
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1984, 141 (01): : 71 - 74
  • [33] Influence of Heart Rate and Innovative Motion-Correction Algorithm on Coronary Artery Image Quality and Measurement Accuracy Using 256-Detector Row Computed Tomography Scanner: Phantom Study
    Park, Jeong Bin
    Jeong, Yeon Joo
    Lee, Geewon
    Lee, Nam Kyung
    Kim, Jin You
    Lee, Ji Won
    KOREAN JOURNAL OF RADIOLOGY, 2019, 20 (01) : 94 - 101
  • [34] Comparison of Diagnostic Quality of Multidetector Computed Tomography and Cone-Beam Computed RESEARCH ARTICLE Tomography in High and Low Resolution Modes for Assessment of the Nasal Cavity
    Shokri, Abbas
    Foroozandeh, Maryam
    Irani, Amin Doosti
    Asalian, Shadi
    WORLD JOURNAL OF PLASTIC SURGERY, 2023, 12 (02) : 77 - 89
  • [35] Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality
    Alshamari, Muhammed
    Geijer, Mats
    Norrman, Eva
    Geijer, Hakan
    ACTA RADIOLOGICA, 2014, 55 (07) : 824 - 832
  • [36] Accuracy of Pulmonary Nodule Volumetry at Different Exposure Parameters in Low-Dose Computed Tomography: A Phantom Study
    Liu, Jieke
    Qing, Haomiao
    Luo, Hongbing
    He, Changjiu
    Wang, Chunhua
    Ren, Jing
    Zhou, Peng
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2019, 43 (06) : 926 - 930
  • [37] Dual-energy computed tomography pulmonary angiography with ultra-low dose contrast administration: Comparison of image quality with standard computed tomography pulmonary angiography
    Celtikci, Pinar
    Hekimoglu, Koray
    Kahraman, Gokhan
    Haberal, Kemal Murat
    Kilic, Dalokay
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2022, 30 (04): : 549 - 556
  • [38] Evaluation of Diagnostic Accuracy of Contrast-enhanced Computed Tomography in Acute Abdomen: A Cross-sectional Study
    Madhesia, Ajeet Kumar
    Panda, Sangram
    Mohanty, Sudhansu Sekhar
    Suma, Matadha Kumaraswamy
    Sen, Kamal Kumar
    Kolluru, Radha Krishna
    Swaraj, Sunny
    Rajesh, Yalamanchi
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2023, 17 (05) : TC6 - TC9
  • [39] COMPARISON OF DIAGNOSTIC CONSISTENCY AND DIAGNOSTIC ACCURACY BETWEEN SURVEY RADIOGRAPHY AND COMPUTED TOMOGRAPHY OF THE SKULL IN 30 RABBITS WITH DENTAL DISEASE
    Capello, Vittorio
    Cauduro, Alberto
    JOURNAL OF EXOTIC PET MEDICINE, 2016, 25 (02) : 115 - 127
  • [40] In-Line Phase Contrast Mammography, Phase Contrast Digital Breast Tomosynthesis, and Phase Contrast Breast Computed Tomography With a Dedicated CT Scanner and a Microfocus X-Ray Tube: Experimental Phantom Study
    Mettivier, Giovanni
    di Franco, Francesca
    Sarno, Antonio
    Castriconi, Roberta
    Di Lillo, Francesca
    Bliznakova, Kristina
    Russo, Paolo
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (06) : 793 - 806