Analytical Performance of a Standardized Single-Platform MHC Tetramer Assay for the Identification and Enumeration of CMV-Specific CD8+T Lymphocytes

被引:18
|
作者
Brooimans, Rik A. [1 ]
Boyce, Christopher S. [2 ]
Popma, Janice [3 ]
Broyles, Dennis A. [2 ]
Gratama, Jan W. [1 ]
Southwick, Paula C. [2 ]
Keeney, Michael [3 ]
机构
[1] Erasmus MC, Dr Daniel Den Hoed Canc Ctr, Dept Internal Oncol, NL-3008 AE Rotterdam, Netherlands
[2] Beckman Coulter Inc, San Diego, CA USA
[3] London Hlth Sci, Dept Hematol Flow Cytometry, London, ON, Canada
关键词
cytomegalovirus; CD8 T lymphocytes; tetramer assay; flow cytometry;
D O I
10.1002/cyto.a.20641
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Major histocompatibility complex (MHC) multimers that identify antigen- specific T cells, coupled with flow cytometry, have made a major impact on immunological research. HLA Class I multimers detect T cells directed against viral, tumor, and transplantation antigens with exquisite sensitivity. This technique has become an important standard for the quantification of a T cell immune response. The utility of this method in multicenter studies, however, is dependant on reproducibility between laboratories. As part of a clinical study using a standardized two-tube three-color single-platform method, we monitored and characterized performance across multiple sites using tetramers against the T cell receptors (TCR) specific for MHC Class I, A*0101-VTEHDTLLY, A*0201-NLVPMVATV and B*0702-TPRVTGGGAM CMV peptides. We studied the analytical performance of this method, focusing on reducing background, maximizing signal intensity, and ensuring that sufficient cells are enumerated to provide meaningful statistics. Inter and intra-assay performance were assessed, which included inherent variability introduced by shipping, type of flow cytometer used, protocol adherence, and analytical interpretation across a range of multiple sample levels and specificities under routine laboratory testing conditions. Using the described protocol, it is possible to obtain intra- and interlab CV's of <20%, with a functional sensitivity for absolute tetramer counts of 1 cell/mu L and 0.2% tetramer+ percent for A*0101, A*0201, and B*0702 alleles. The standardized single-platform MHC tetramer assay is simple, rapid, reproducible, and useful for assessing CMV-specific T cells, and will allow for reasonable comparisons of clinical evaluations across multiple centers at clinically relevant thresholds (2.0-10.0 cells/mu L). (C) 2008 international Society for Advancement of Cytometry
引用
收藏
页码:992 / 1000
页数:9
相关论文
共 50 条
  • [1] Enumeration of antigen-specific CD8+T cells by single-platform, HLA tetramer-based flow cytometry: A European multicenter study
    Heijnen, IAFM
    Gratama, JW
    CYTOMETRY PART A, 2004, 59A (01): : 65 - 65
  • [2] SCREENING AND ENUMERATION OF VIRUS- AND TUMOR-ASSOCIATED ANTIGEN-SPECIFIC CD8+T LYMPHOCYTES USING STREPTAMER TECHNOLOGY AND SINGLE-PLATFORM FLOWCYTOMETRY ASSAY
    Matko, S.
    Teichert, M.
    Odendahl, M.
    Wohsmann, M.
    Tonn, T.
    CYTOTHERAPY, 2014, 16 (04) : S26 - S27
  • [3] Enumeration of antigen-specific CD8+ T lymphocytes by single-platform, HLA tetramer-based flow cytometry:: A European multicenter evaluation
    Heijnen, IAFM
    Barnett, D
    Arroz, MJ
    Barry, SM
    Bonneville, M
    Brando, B
    D'hautcourt, JL
    Kern, F
    Tötterman, TH
    Marijt, EWA
    Bossy, D
    Preijers, FWMB
    Rothe, G
    Gratama, JW
    CYTOMETRY PART B-CLINICAL CYTOMETRY, 2004, 62B (01) : 1 - 13
  • [4] Characterization of CMV-specific CD8+T-cells using MHC tetramers
    Tötterman, TH
    Engstrand, M
    Korsgren, O
    Zweigberg-Wirgart, B
    Eriksson, BM
    Wadstrom, J
    Romagné, F
    Bonneville, M
    CYTOMETRY, 2000, 42 (02): : 138 - 138
  • [5] Flow cytometric enumeration of Class IHLA-restricted, peptide-specific CD8+ T lymphocytes using tetramer technology and single-platform absolute T-cell counting
    Kraan, J
    Arroz, M
    Keeney, M
    Freire, M
    Weir, K
    Heijnen, I
    Gratama, JW
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2003, 17 (03): : 268 - 278
  • [6] Enumeration of WT1-specific CD8+ T cells using an MHC streptamer based no-wash single-platform flow-cytometric assay
    Matko, S.
    Teichert, M.
    Tunger, A.
    Schmitz, M.
    Bornhaeuser, M.
    Tonn, T.
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2017, 44 : 74 - 75
  • [7] MESENCHYMAL STROMAL CELLS INHIBIT PROLIFERATION OF CMV-SPECIFIC CD8+T CELLS
    Malcherek, G.
    Jin, N.
    Wuchter, P.
    Diehlmann, A.
    Schmitt, A.
    Dreger, P.
    Ho, A. D.
    Michael, S.
    BONE MARROW TRANSPLANTATION, 2014, 49 : S192 - S192
  • [8] CMV-specific CD8+T cells from RA patients contribute to autoimmune disease
    Rothe, K.
    Quandt, D.
    Schubert, K.
    Klingner, M.
    Baerwald, C.
    Wagner, U.
    ZEITSCHRIFT FUR RHEUMATOLOGIE, 2013, 72 : 40 - 41
  • [9] Flow cytometric detection of degranulation reveals phenotypic heterogeneity of degranulating CMV-specific CD8+T lymphocytes in rhesus macaques
    Chan, Kenneth S.
    Kaur, Amitinder
    JOURNAL OF IMMUNOLOGICAL METHODS, 2007, 325 (1-2) : 20 - 34
  • [10] Effect of Type of Dialysis on CMV-Specific CD8+T Cells in Kidney Transplant Candidates
    Ramon Vidal-Castineira, Jose
    Corte-Iglesias, Viviana
    Sobrino-Diaz, Lucia
    Perez-Fernandez, Sonia
    Melon, Santiago
    Lopez-Larrea, Carlos
    Diaz-Corte, Carmen
    FRONTIERS IN IMMUNOLOGY, 2019, 10 : 1680