Dosimetry comparison between TG-43 and Monte Carlo calculations using the Freiburg flap for skin high-dose-rate brachytherapy

被引:18
|
作者
Vijande, Javier [1 ,2 ]
Ballester, Facundo [1 ]
Ouhib, Zoubir [3 ]
Granero, Domingo [4 ]
Carmen Pujades-Claumarchirant, M. [5 ]
Perez-Calatayud, Jose [5 ]
机构
[1] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[2] Univ Valencia, CSIC, IFIC, Paterna, Valencia, Spain
[3] Boca Raton Reg Hosp, Lynn Canc Inst, Boca Raton, FL USA
[4] Hosp Gen Univ, ERESA, Dept Radiat Phys, Valencia, Spain
[5] La Fe Univ Hosp, Dept Radiat Oncol, Valencia, Spain
关键词
Ir-192; Brachytherapy; Dosimetry; Penelope2008; Freiburg flap; PARAMETERS; IR-192; APPLICATORS;
D O I
10.1016/j.brachy.2011.11.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work was to evaluate whether the delivered dose to the skin surface and at the prescription depth when using a Freiburg flap applicator is in agreement with the one predicted by the treatment planning system (TPS) using the TG-43 dose-calculation formalism. METHODS AND MATERIALS: Monte Carlo (MC) simulations and radiochromic film measurements have been performed to obtain dose distributions with the source located at the center of one of the spheres and between two spheres. Primary and scatter dose contributions were evaluated to understand the role played by the scatter component. A standard treatment plan was generated using MC- and TG-43-based TPS applying the superposition principle. RESULTS: The MC model has been validated by performing additional simulations in the same conditions but transforming air and Freiburg flap materials into water to match TG-43 parameters. Both dose distributions differ less than 1%. Scatter defect compared with TG-43 data is up to 15% when the source is located at the center of the sphere and up to 25% when the source is between two spheres. Maximum deviations between TPS- and MC-based distributions are of 5%. CONCLUSIONS: The deviations in the TG-43-based dose distributions for a standard treatment plan with respect to the MC dose distribution calculated taking into account the composition and shape of the applicator and the surrounding air are lower than 5%. Therefore, this study supports the validity of the TPS used in clinical practice. (C) 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:528 / 535
页数:8
相关论文
共 50 条
  • [41] Monte Carlo Evaluation for the Effect of Positional Inaccuracy of Source on Patient's Dose in High-Dose-Rate Brachytherapy
    Lee, Kyung Ja
    An, So Hyun
    Lee, Rena
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (06) : 2361 - 2365
  • [42] THUBrachy: fast Monte Carlo dose calculation tool accelerated by heterogeneous hardware for high-dose-rate brachytherapy
    Hu, An-Kang
    Qiu, Rui
    Liu, Huan
    Wu, Zhen
    Li, Chun-Yan
    Zhang, Hui
    Li, Jun-Li
    Yang, Rui-Jie
    NUCLEAR SCIENCE AND TECHNIQUES, 2021, 32 (03)
  • [43] THUBrachy: fast Monte Carlo dose calculation tool accelerated by heterogeneous hardware for high-dose-rate brachytherapy
    An-Kang Hu
    Rui Qiu
    Huan Liu
    Zhen Wu
    Chun-Yan Li
    Hui Zhang
    Jun-Li Li
    Rui-Jie Yang
    NuclearScienceandTechniques, 2021, 32 (03) : 109 - 121
  • [44] THUBrachy: fast Monte Carlo dose calculation tool accelerated by heterogeneous hardware for high-dose-rate brachytherapy
    An-Kang Hu
    Rui Qiu
    Huan Liu
    Zhen Wu
    Chun-Yan Li
    Hui Zhang
    Jun-Li Li
    Rui-Jie Yang
    Nuclear Science and Techniques, 2021, 32
  • [45] Comparison of high-dose-rate intracavitary brachytherapy dosimetry with and without anesthesia in patients with cervical carcinoma
    Sharma, Daya N.
    Chaudhari, Pritee
    Sharma, Seema
    Gupta, Leena
    Jagadesan, Pandjatcharam
    Rath, Goura K.
    Julka, Pramod K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 60 - 66
  • [46] Monte Carlo dosimetric characterization of the Flexisource Co-60 high-dose-rate brachytherapy source using PENELOPE
    Almansa, Julio F.
    Guerrero, Rafael
    Torres, Javier
    Lallena, Antonio M.
    BRACHYTHERAPY, 2017, 16 (05) : 1073 - 1080
  • [47] Monte Carlo dosimetry of a sonolucent vaginal cylinder for three-dimensional transvaginal ultrasound guidance in high-dose-rate interstitial gynecologic brachytherapy
    Van Elburg, Devin
    Morrison, Hali
    Rodgers, Jessica
    Fenster, Aaron
    Roumeliotis, Michael
    Meyer, Tyler
    MEDICAL PHYSICS, 2019, 46 (11) : 5397 - 5397
  • [48] Validating a 192Ir-Based Small Animal Irradiation Apparatus Using a 3D-Printed Applicator: Comparison Between TG-43, Monte Carlo and Films Dosimetry
    Nie, K.
    Fekete, C. Collins
    Pinnaduwage, D.
    Cunha, J. A. M.
    Mellis, K.
    Descovich, M.
    Beaulieu, L.
    Pouliot, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [49] Monte Carlo dose calculations of clinical post-implant cases: A comparison to the TG43 standard
    Beaulieu, L.
    Carrier, J-F.
    D'Amours, M.
    Martin, A-G.
    Vigneault, E.
    RADIOTHERAPY AND ONCOLOGY, 2006, 80 : S42 - S43
  • [50] Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy
    Naseri, Alireza
    Mesbahi, Asghar
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2009, 14 (06) : 200 - 204