Dosimetry comparison between TG-43 and Monte Carlo calculations using the Freiburg flap for skin high-dose-rate brachytherapy

被引:18
|
作者
Vijande, Javier [1 ,2 ]
Ballester, Facundo [1 ]
Ouhib, Zoubir [3 ]
Granero, Domingo [4 ]
Carmen Pujades-Claumarchirant, M. [5 ]
Perez-Calatayud, Jose [5 ]
机构
[1] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[2] Univ Valencia, CSIC, IFIC, Paterna, Valencia, Spain
[3] Boca Raton Reg Hosp, Lynn Canc Inst, Boca Raton, FL USA
[4] Hosp Gen Univ, ERESA, Dept Radiat Phys, Valencia, Spain
[5] La Fe Univ Hosp, Dept Radiat Oncol, Valencia, Spain
关键词
Ir-192; Brachytherapy; Dosimetry; Penelope2008; Freiburg flap; PARAMETERS; IR-192; APPLICATORS;
D O I
10.1016/j.brachy.2011.11.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work was to evaluate whether the delivered dose to the skin surface and at the prescription depth when using a Freiburg flap applicator is in agreement with the one predicted by the treatment planning system (TPS) using the TG-43 dose-calculation formalism. METHODS AND MATERIALS: Monte Carlo (MC) simulations and radiochromic film measurements have been performed to obtain dose distributions with the source located at the center of one of the spheres and between two spheres. Primary and scatter dose contributions were evaluated to understand the role played by the scatter component. A standard treatment plan was generated using MC- and TG-43-based TPS applying the superposition principle. RESULTS: The MC model has been validated by performing additional simulations in the same conditions but transforming air and Freiburg flap materials into water to match TG-43 parameters. Both dose distributions differ less than 1%. Scatter defect compared with TG-43 data is up to 15% when the source is located at the center of the sphere and up to 25% when the source is between two spheres. Maximum deviations between TPS- and MC-based distributions are of 5%. CONCLUSIONS: The deviations in the TG-43-based dose distributions for a standard treatment plan with respect to the MC dose distribution calculated taking into account the composition and shape of the applicator and the surrounding air are lower than 5%. Therefore, this study supports the validity of the TPS used in clinical practice. (C) 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:528 / 535
页数:8
相关论文
共 50 条
  • [31] COMPARISON OF TG43-BASED SKIN DOSE CALCULATIONS TO INVIVO SKIN DOSIMETRY FOR PERMANENT BREAST PD-103 SEED IMPLANT BRACHYTHERAPY
    Villarreal-Barajas, Jose Eduardo
    Watt, Elizabeth
    Long, Karen
    Karchewski-Welter, Ruth
    Roumeliotis, Michael
    Husain, Siraj
    Meyer, Tyler
    RADIOTHERAPY AND ONCOLOGY, 2016, 120 : S47 - S48
  • [32] Pediatric peri-operative fractionated high-dose-rate brachytherapy for recurrent Wilms' tumor using a reconstructed Freiburg flap
    Flower, Emily
    Tran, Kathy Ngoc
    Zanjani, Salman
    Smith, Wayne
    Karpelowsky, Jonathan
    Summerhayes, Katie
    Estoesta, Edgar
    Chard, Jennifer
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (05) : 429 - 433
  • [33] DEEP LEARNING-BASED TRANSFER OF TG-43 BASED DOSE TO WATER TO DOSE TO MEDIUM IN MEDIUM IN HIGH DOSE RATE BRACHYTHERAPY
    Quetin, Sebastien
    Bahoric, Boris
    Maleki, Farhad
    Enger, Shirin Abbasinejad
    RADIOTHERAPY AND ONCOLOGY, 2023, 186 : S75 - S75
  • [34] Sub-second high dose rate brachytherapy Monte Carlo dose calculations with bGPUMCD
    Hissoiny, Sami
    D'Amours, Michel
    Ozell, Benoit
    Despres, Philippe
    Beaulieu, Luc
    MEDICAL PHYSICS, 2012, 39 (07) : 4559 - 4567
  • [35] Dosimetric comparison of AcurosBV with AAPM TG43 dose calculation formalism in cervical intraductal high-dose-rate brachytherapy using three different applicators
    Bi, Su-yan
    Chen, Zhi-jian
    Sun, Xing-ru
    Dai, Zhi-tao
    PRECISION RADIATION ONCOLOGY, 2022, 6 (03): : 234 - 242
  • [36] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Ahn, Woo Sang
    Choi, Wonsik
    Shin, Seong Soo
    Jung, Jinhong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (04) : 557 - 564
  • [37] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Woo Sang Ahn
    Wonsik Choi
    Seong Soo Shin
    Jinhong Jung
    Journal of the Korean Physical Society, 2014, 65 : 557 - 564
  • [38] Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    Daskalov, GM
    Loffler, E
    Williamson, JF
    MEDICAL PHYSICS, 1998, 25 (11) : 2200 - 2208
  • [39] Monte Carlo Dosimetry of the 60Co Bebig High Dose Rate Source for Brachytherapy
    Campos, L. T.
    de Almeida, C. E. V.
    MEDICAL PHYSICS, 2016, 43 (06) : 3464 - 3464
  • [40] Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    MEDICAL PHYSICS, 2000, 27 (11) : 2521 - 2527