Absolute linearity measurements on a PbS detector in the infrared

被引:26
|
作者
Theocharous, E [1 ]
机构
[1] Natl Phys Lab, Div Qual Life, Opt Radiat Team, Teddington TW11 0LW, Middx, England
关键词
D O I
10.1364/AO.45.002381
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The nonlinearity characteristics of a commercially available thin-film photoconductive PbS detector were experimentally investigated in the infrared using the National Physical Laboratory detector linearity characterization facility. The deviation from linearity of this detector was shown to be significant even for relatively low values of radiant power incident on the active area of the detector. For example, the linearity factor was approximately 0.8 when 0.6 mu W of radiant power at a wavelength of 2.2 mu m was illuminating a spot of 1 mm in diameter on the active area of the PbS detector. These figures demonstrate the poor linearity characteristics of this detector and provide a warning to other users of PbS detection systems. The deviation from linearity was shown to be a function of the size of the spot being illuminated on the detector active area, as well as the wavelength of the incident radiation. The deviation from linearity was shown to be a function of irradiance illuminating the detector for irradiance values lower than 1 mu W mm(-2). (c) 2006 Optical Society of America.
引用
收藏
页码:2381 / 2386
页数:6
相关论文