Repeatability and agreement between a reference Scheimpflug tomographer and a low-cost Scheimpflug system

被引:6
|
作者
Xu, Wenjun [1 ]
Zhai, Changbin [1 ]
Yusufu, Mayinuer [2 ,3 ]
Gan, Jiahe [1 ]
Li, Shiming [1 ]
Zheng, Yan [1 ]
Kang, Mengtian [1 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China
[2] Royal Victorian Eye & Ear Hosp, Ctr Eye Res Australia, East Melbourne, Australia
[3] Univ Melbourne, Dept Surg Ophthalmol, Melbourne, Vic, Australia
来源
基金
中国国家自然科学基金;
关键词
OPTICAL COHERENCE TOMOGRAPHY; CORNEAL THICKNESS; PENTACAM HR; REPRODUCIBILITY; RELIABILITY; ACCURACY; ECTASIA;
D O I
10.1097/j.jcrs.0000000000001168
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the repeatability and agreement between Scheimpflug-based corneal topographers, Scansys and Pentacam, in measuring ocular parameters in myopic eyes. Setting: Grade-A tertiary hospital in Beijing, China. Design: Cross-sectional study. Methods: The following measurements were performed in 204 subjects undergoing preoperative examinations for refractive surgery from the corneal curvature, astigmatism, corneal thickness, corneal volume, corneal asphericity, anterior chamber depth ( ACD) and volume (ACV), and pupil diameters. Repeatability was determined using intraclass correlation coefficient with 95% CI, Cronbach a, coefficient of variation, within- subject SD, test-retest repeatability, and 1-way analysis of variance (1-way analysis of variance). The interdevice agreement was determined using paired t test and BlandAltman plots. Results: Scansys showed good repeatability in the anterior and posterior corneal flattest meridian (Kf), steepest meridian (Ks), and mean (Km); axis of anterior corneal Ks; anterior corneal astigmatism; pupil central corneal thickness (CCT); CCT; corneal apex thickness; thinnest corneal thickness; corneal volume; photopic pupil diameter; ACD; and ACV. In the agreement study, Bland-Altman plots showed that 95% limit of agreement of corneal curvature, pupil CCT, corneal apex thickness, thinnest corneal thickness, and corneal volume generated by Scansys and Pentacam were narrow. There were no statistically significant differences in the anterior corneal Kf and Km. Conclusions: Scansys showed good repeatability in measuring corneal curvature of anterior and posterior surfaces, anterior corneal astigmatism, corneal thickness, corneal volume, photopic pupil diameter, ACV, and ACD. Scansys and Pentacam can only be used interchangeably in measuring anterior corneal curvature. Copyright (c) 2023 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:614 / 619
页数:6
相关论文
共 50 条
  • [41] Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    Kawamorita, Takushi
    Uozato, Hiroshi
    Kamiya, Kazutaka
    Bax, Leon
    Tsutsui, Kenta
    Aizawa, Daisuke
    Shimizu, Kimiya
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01): : 127 - 133
  • [42] Reproducibility, Repeatability, and Correlation of Central Corneal Thickness Measurement with the Pentacam Scheimpflug System and Ultrasound Pachymetry
    Macias-Rodriguez, Yolanda
    Ramos-Davila, Eugenia M.
    Ruiz-Lozano, Raul E.
    Reyes-Arena, Jose, V
    Jocelyn Rivera-Alvarado, I
    Hernandez-Camarena, Julio C.
    Rodriguez-Garcia, Alejandro
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2024, 241 (11) : 1238 - 1244
  • [43] Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
    Gharieb, Hesham Mohamed
    Shalaby, Hisham Samy
    Othman, Ihab Saad
    CLINICAL OPHTHALMOLOGY, 2020, 14 : 3801 - 3810
  • [44] Agreement between Scheimpflug photography and A-scan ultrasonography in anterior segment ocular measurements in children
    Tong, L
    Wong, EM
    Chan, YH
    Balakrishnan, V
    OPTOMETRY AND VISION SCIENCE, 2003, 80 (07) : 529 - 534
  • [45] Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer
    Tana-Rivero, Pedro
    Aguilar-Corcoles, Salvador
    Tello-Elordi, Cristina
    Pastor-Pascual, Francisco
    Montes-Mico, Robert
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (04): : 488 - 495
  • [46] Intrasubject repeatability in keratoconus-eye measurements obtained with a new Scheimpflug photography-based system
    Montalban, Raul
    Alio, Jorge L.
    Javaloy, Jaime
    Pinero, David P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (02): : 211 - 218
  • [47] A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug plus Placido topographer and rotating prism auto-keratorefractor
    Prakash, Gaurav
    Srivastava, Dhruv
    Choudhuri, Sounak
    INTERNATIONAL OPHTHALMOLOGY, 2015, 35 (06) : 869 - 880
  • [48] Interchangeability between Placido disc and Scheimpflug system: quantitative and qualitative analysis
    De Stefano, Vinicius Silbiger
    Soares Melo Junior, Luiz Alberto
    Mallmann, Felipe
    Schor, Paulo
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2010, 73 (04) : 363 - 366
  • [49] A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor
    Gaurav Prakash
    Dhruv Srivastava
    Sounak Choudhuri
    International Ophthalmology, 2015, 35 : 869 - 880
  • [50] Scheimpflug vs Scanning-Slit Corneal Tomography: Comparison of Corneal and Anterior Chamber Tomography Indices for Repeatability and Agreement in Healthy Eyes
    Kanellopoulos, Anastasios John
    CLINICAL OPHTHALMOLOGY, 2020, 14 : 2583 - 2592