Repeatability and agreement between a reference Scheimpflug tomographer and a low-cost Scheimpflug system

被引:6
|
作者
Xu, Wenjun [1 ]
Zhai, Changbin [1 ]
Yusufu, Mayinuer [2 ,3 ]
Gan, Jiahe [1 ]
Li, Shiming [1 ]
Zheng, Yan [1 ]
Kang, Mengtian [1 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China
[2] Royal Victorian Eye & Ear Hosp, Ctr Eye Res Australia, East Melbourne, Australia
[3] Univ Melbourne, Dept Surg Ophthalmol, Melbourne, Vic, Australia
来源
基金
中国国家自然科学基金;
关键词
OPTICAL COHERENCE TOMOGRAPHY; CORNEAL THICKNESS; PENTACAM HR; REPRODUCIBILITY; RELIABILITY; ACCURACY; ECTASIA;
D O I
10.1097/j.jcrs.0000000000001168
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the repeatability and agreement between Scheimpflug-based corneal topographers, Scansys and Pentacam, in measuring ocular parameters in myopic eyes. Setting: Grade-A tertiary hospital in Beijing, China. Design: Cross-sectional study. Methods: The following measurements were performed in 204 subjects undergoing preoperative examinations for refractive surgery from the corneal curvature, astigmatism, corneal thickness, corneal volume, corneal asphericity, anterior chamber depth ( ACD) and volume (ACV), and pupil diameters. Repeatability was determined using intraclass correlation coefficient with 95% CI, Cronbach a, coefficient of variation, within- subject SD, test-retest repeatability, and 1-way analysis of variance (1-way analysis of variance). The interdevice agreement was determined using paired t test and BlandAltman plots. Results: Scansys showed good repeatability in the anterior and posterior corneal flattest meridian (Kf), steepest meridian (Ks), and mean (Km); axis of anterior corneal Ks; anterior corneal astigmatism; pupil central corneal thickness (CCT); CCT; corneal apex thickness; thinnest corneal thickness; corneal volume; photopic pupil diameter; ACD; and ACV. In the agreement study, Bland-Altman plots showed that 95% limit of agreement of corneal curvature, pupil CCT, corneal apex thickness, thinnest corneal thickness, and corneal volume generated by Scansys and Pentacam were narrow. There were no statistically significant differences in the anterior corneal Kf and Km. Conclusions: Scansys showed good repeatability in measuring corneal curvature of anterior and posterior surfaces, anterior corneal astigmatism, corneal thickness, corneal volume, photopic pupil diameter, ACV, and ACD. Scansys and Pentacam can only be used interchangeably in measuring anterior corneal curvature. Copyright (c) 2023 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:614 / 619
页数:6
相关论文
共 50 条
  • [21] The Repeatability of Axial Length Measurements Using a Scheimpflug-based System
    Sardari, Sara
    Khabazkhoob, Mehdi
    Jafarzadehpur, Ebrahim
    Fotouhi, Akbar
    JOURNAL OF OPHTHALMIC & VISION RESEARCH, 2023, 18 (04) : 396 - 404
  • [22] Repeatability and agreement of a Scheimpflug-Placido tomography and a Partial Coherence Interferometer in keratoconus and control eyes
    Ozkan, Yaprak
    Gokcinar, Nesrin Buyuktortop
    INTERNATIONAL OPHTHALMOLOGY, 2025, 45 (01)
  • [23] Pachymetric measurements with a new Scheimpflug photography-based system Intraobserver repeatability and agreement with optical coherence tomography pachymetry
    Milla, Myriam
    Pinero, David P.
    Amparo, Francisco
    Alio, Jorge L.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02): : 310 - 316
  • [24] Agreement Between Placido Topography and Scheimpflug Tomography for Corneal Astigmatism Assessment
    Delrivo, Marianella
    Ruisenor Vazquez, Pablo R.
    David Galletti, Jonatan
    Garibotto, Martin
    Fuentes Bonthoux, Fernando
    Pfoertner, Tomas
    Gaston Galletti, Jeremias
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (01) : 49 - 53
  • [25] Repeatability and Agreement of Three Scheimpflug-Based Imaging Systems for Measuring Anterior Segment Parameters in Keratoconus
    Shetty, Rohit
    Arora, Vishal
    Jayadev, Chaitra
    Nuijts, Rudy M. M. A.
    Kumar, Mukesh
    Puttaiah, Narendra K.
    Kummelil, Mathew Kurian
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (08) : 5263 - 5268
  • [26] Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: Repeatability, reproducibility, and agreement
    Wang, Qinmei
    Ding, Xixia
    Savini, Giacomo
    Chen, Hao
    Feng, Yifan
    Pan, Chao
    Hua, Yanjun
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (01): : 178 - 185
  • [27] Agreement and repeatability of corneal tomography in healthy eyes using a new swept-source OCT, a rotating Scheimpflug camera, and a dual Scheimpflug-Placido system (vol 48, pg 190, 2022)
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (05): : 646 - 646
  • [28] Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras
    Savini, Giacomo
    Noeser, Kristian
    Schiano-Lomoriello, Domenico
    Ducoli, Pietro
    CORNEA, 2017, 36 (04) : 463 - 469
  • [29] Agreement between swept-source OCT and Scheimpflug biometers in ocular measurements
    Mclintock, Cameron
    Mckelvie, James
    Niyazmand, Hamed
    Uprety, Samir
    BMC OPHTHALMOLOGY, 2025, 25 (01)
  • [30] Repeatability and agreement of two swept-source optical coherence tomographers and scheimpflug imaging for measurements of corneal curvature
    Cheng, Shiming
    Zhang, Jiasheng
    Zhou, Kaijing
    Li, Tiantian
    Lin, Junhai
    Yu, A. -Yong
    BMC OPHTHALMOLOGY, 2024, 24 (01)