Repeatability and agreement between a reference Scheimpflug tomographer and a low-cost Scheimpflug system

被引:6
|
作者
Xu, Wenjun [1 ]
Zhai, Changbin [1 ]
Yusufu, Mayinuer [2 ,3 ]
Gan, Jiahe [1 ]
Li, Shiming [1 ]
Zheng, Yan [1 ]
Kang, Mengtian [1 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China
[2] Royal Victorian Eye & Ear Hosp, Ctr Eye Res Australia, East Melbourne, Australia
[3] Univ Melbourne, Dept Surg Ophthalmol, Melbourne, Vic, Australia
来源
基金
中国国家自然科学基金;
关键词
OPTICAL COHERENCE TOMOGRAPHY; CORNEAL THICKNESS; PENTACAM HR; REPRODUCIBILITY; RELIABILITY; ACCURACY; ECTASIA;
D O I
10.1097/j.jcrs.0000000000001168
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the repeatability and agreement between Scheimpflug-based corneal topographers, Scansys and Pentacam, in measuring ocular parameters in myopic eyes. Setting: Grade-A tertiary hospital in Beijing, China. Design: Cross-sectional study. Methods: The following measurements were performed in 204 subjects undergoing preoperative examinations for refractive surgery from the corneal curvature, astigmatism, corneal thickness, corneal volume, corneal asphericity, anterior chamber depth ( ACD) and volume (ACV), and pupil diameters. Repeatability was determined using intraclass correlation coefficient with 95% CI, Cronbach a, coefficient of variation, within- subject SD, test-retest repeatability, and 1-way analysis of variance (1-way analysis of variance). The interdevice agreement was determined using paired t test and BlandAltman plots. Results: Scansys showed good repeatability in the anterior and posterior corneal flattest meridian (Kf), steepest meridian (Ks), and mean (Km); axis of anterior corneal Ks; anterior corneal astigmatism; pupil central corneal thickness (CCT); CCT; corneal apex thickness; thinnest corneal thickness; corneal volume; photopic pupil diameter; ACD; and ACV. In the agreement study, Bland-Altman plots showed that 95% limit of agreement of corneal curvature, pupil CCT, corneal apex thickness, thinnest corneal thickness, and corneal volume generated by Scansys and Pentacam were narrow. There were no statistically significant differences in the anterior corneal Kf and Km. Conclusions: Scansys showed good repeatability in measuring corneal curvature of anterior and posterior surfaces, anterior corneal astigmatism, corneal thickness, corneal volume, photopic pupil diameter, ACV, and ACD. Scansys and Pentacam can only be used interchangeably in measuring anterior corneal curvature. Copyright (c) 2023 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:614 / 619
页数:6
相关论文
共 50 条
  • [1] Repeatability and agreement of topometric parameters between a topographer, scheimpflug tomographer, and optical coherence tomographer in Keratoconus
    Angelo, Lize
    Gokul, Akilesh
    Wadhwa, Himanshu
    McGhee, Charles
    Ziaei, Mohammed
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2022, 50 (08): : 904 - 905
  • [2] Corneal aberrations measured with a high-resolution Scheimpflug tomographer: repeatability and reproducibility
    McAlinden, Colm
    Schwiegerling, Jim
    Khadka, Jyoti
    Pesudovs, Konrad
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (04): : 581 - 590
  • [3] Agreement of anterior ocular biometric measurements with a new optical biometer and a Scheimpflug tomographer
    McAlinden, Colm
    Bao, Fangjun
    Guo, Yan
    Yu, Xinxin
    Lu, Weicong
    Chen, Hao
    Wang, Qinmei
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (05): : 679 - 684
  • [4] Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer
    Kim, Eric J.
    de Oca, Ildamaris Montes
    Wang, Li
    Weikert, Mitchell P.
    Koch, Douglas D.
    Khandelwal, Sumitra S.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (12): : 2731 - 2738
  • [5] Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (04): : 471 - 478
  • [6] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    EYE AND VISION, 2022, 9 (01)
  • [7] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Robert Herber
    Janine Lenk
    Lutz E. Pillunat
    Frederik Raiskup
    Eye and Vision, 9
  • [8] Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreement
    Chen, Wuhe
    McAlinden, Colm
    Pesudovs, Konrad
    Wang, Qinmei
    Lu, Fan
    Feng, Yifan
    Chen, Jie
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (09): : 1626 - 1632
  • [9] Evaluation of the Reliability and Repeatability of Scheimpflug System Measurement in Keratoconus
    de Luis Eguileor, Beatriz
    Escudero Argaluza, Julene
    Pijoan Zubizarreta, Jose Ignacio
    Santamaria Carro, Alaitz
    Etxebarria Ecenarro, Jaime
    CORNEA, 2018, 37 (02) : 177 - 181
  • [10] The repeatability and agreement of biometric measurements by dual Scheimpflug device with integrated optical biometer
    Hassan Hashemi
    Sara Sardari
    Abbasali Yekta
    Mehdi Khabazkhoob
    Scientific Reports, 12