Improving adult pulsatile minimal invasive extracorporeal circulation in a mock circulation

被引:1
|
作者
Duerr, Anke [1 ]
Weber, Elena [1 ]
Eisenmann, Lisa [1 ]
Albrecht, Guenter [1 ]
Liebold, Andreas [1 ]
Hoenicka, Markus [1 ]
机构
[1] Ulm Univ, Dept Cardiothorac & Vasc Surg, Med Ctr, Ulm, Germany
关键词
cannula; diagonal pump; minimal invasive extracorporeal circulation; oxygenator; pulsatile perfusion; surplus hemodynamic energy; PERFUSION; FLOW; PUMP; METAANALYSIS;
D O I
10.1111/aor.14439
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background Pulsatile extracorporeal circulation (ECC) may improve perfusion of critical organs during cardiac surgery. This study analyzed the influence of the components of a minimal invasive ECC (MiECC) on the transfer of pulsatile energy into the pseudo-patient of a mock circulation. Methods An aortic model with human-like geometry and compliance was perfused by a diagonal pump. Surplus hemodynamic energy (SHE) was determined from flow and pressure data. Five adult-size oxygenator models and three sizes of cannulas were compared. Pulsatile pump settings were optimized, and parallel dual-pump configurations were evaluated. Results Oxygenator models showed up to twofold differences in pressure gradients and influenced SHE at flow rates up to 2.0 L min(-1). Adjustments of frequency, systole duration, and rotational speed gain significantly improved SHE compared with empirical settings, with SHE above 21% of mean arterial pressure at flow rates of 1.0 L min(-1) to 1.5 L min(-1) and SHE above 5% at 3.5 L min(-1). Small diameter cannula (15 Fr) limited SHE compared with larger cannula (21 Fr and 23 Fr). Two diagonal pumps did not provide higher SHE than a single pump, but permitted additional control over pulse pressure and SHE by varying the total fraction of pulsatile flow and the fraction of flow bypassing the oxygenator. Conclusions Proper selection of components and optimizations of pump settings significantly improved pulse pressure and SHE of pulsatile MiECC. Surplus hemodynamic energy depended on flow rate with a maximum at 1.0 L min(-1)-1.5 L min(-1). Pulsatile MiECC may specifically assist organ perfusion during phases of low flow.
引用
收藏
页码:828 / 839
页数:12
相关论文
共 50 条
  • [41] Impact of Minimal Invasive Extracorporeal Circulation in Coronary Artery Bypass Surgery: Meta-Analysis
    Al Wssawi, Ameer Fadhil Abbas
    Gumera, Armand
    Al-Shammari, Ali Saad
    Abdelsamad, Hasnaa
    Hashim, Hashim Talib
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (16) : S28 - S30
  • [42] Impact of minimal invasive extracorporeal circulation on atrial fibrillation after coronary artery bypass surgery
    Ellam, Sten
    Hartikainen, Juha
    Korvenoja, Pekka
    Pitkanen, Otto
    Tyrvainen, Esko
    Valtola, Antti
    Halonen, Jari
    ARTIFICIAL ORGANS, 2020, 44 (11) : 1176 - 1183
  • [43] Ascending aorta outflow graft location and pulsatile ventricular assist provide optimal hemodynamic support in an adult mock circulation
    Litwak, KN
    Koenig, SC
    Cheng, RC
    Giridharan, GA
    Gillars, KJ
    Pantalos, GM
    ARTIFICIAL ORGANS, 2005, 29 (08) : 629 - 635
  • [44] COMPARATIVE EXPERIMENTAL-STUDY OF PULSATILE PUMPS FOR EXTRACORPOREAL-CIRCULATION
    BELENGER, J
    BRUGGER, JP
    KNIGHT, CJ
    SPROVIERI, L
    EUROPEAN SURGICAL RESEARCH, 1979, 11 : 54 - 54
  • [45] EFFECT OF PULSATILE FLOW ON RENAL BLOOD FLOW DURING EXTRACORPOREAL CIRCULATION
    DALTON, ML
    MOSLEY, EC
    WOODWARD, KE
    BARILA, TG
    JOURNAL OF SURGICAL RESEARCH, 1965, 5 (03) : 127 - &
  • [46] TRANSCRANIAL DOPPLER SONOGRAPHY DURING PULSATILE AND NONPULSATILE EXTRACORPOREAL-CIRCULATION
    THIEL, A
    RUSS, W
    KAPS, M
    STERTMANN, WA
    HEMPELMANN, G
    ANAESTHESIST, 1990, 39 (04): : 226 - 230
  • [47] Characterization of an adult mock circulation for testing cardiac support devices
    Pantalos, GM
    Koenig, SC
    Gillars, KJ
    Giridharan, GA
    Ewert, DL
    ASAIO JOURNAL, 2004, 50 (01) : 37 - 46
  • [48] CLINICAL EXPERIENCES OF PULSATILE FLOW IN EXTRACORPOREAL-CIRCULATION BY USE OF PULSATILE ASSIST DEVICE (PAD)
    YADA, I
    KUSAGAWA, M
    YUASA, H
    CHIKUSA, H
    KINOSHITA, T
    IKOMA, S
    SAKAI, T
    KUBO, M
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1978, 42 (06): : 787 - 787
  • [49] EXTRACORPOREAL CIRCULATION
    不详
    BRITISH MEDICAL JOURNAL, 1959, 1 (JAN3): : 35 - 36
  • [50] EXTRACORPOREAL CIRCULATION
    DABREU, AL
    LANCET, 1958, 1 (MAY24): : 1130 - 1130