Improving adult pulsatile minimal invasive extracorporeal circulation in a mock circulation

被引:1
|
作者
Duerr, Anke [1 ]
Weber, Elena [1 ]
Eisenmann, Lisa [1 ]
Albrecht, Guenter [1 ]
Liebold, Andreas [1 ]
Hoenicka, Markus [1 ]
机构
[1] Ulm Univ, Dept Cardiothorac & Vasc Surg, Med Ctr, Ulm, Germany
关键词
cannula; diagonal pump; minimal invasive extracorporeal circulation; oxygenator; pulsatile perfusion; surplus hemodynamic energy; PERFUSION; FLOW; PUMP; METAANALYSIS;
D O I
10.1111/aor.14439
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background Pulsatile extracorporeal circulation (ECC) may improve perfusion of critical organs during cardiac surgery. This study analyzed the influence of the components of a minimal invasive ECC (MiECC) on the transfer of pulsatile energy into the pseudo-patient of a mock circulation. Methods An aortic model with human-like geometry and compliance was perfused by a diagonal pump. Surplus hemodynamic energy (SHE) was determined from flow and pressure data. Five adult-size oxygenator models and three sizes of cannulas were compared. Pulsatile pump settings were optimized, and parallel dual-pump configurations were evaluated. Results Oxygenator models showed up to twofold differences in pressure gradients and influenced SHE at flow rates up to 2.0 L min(-1). Adjustments of frequency, systole duration, and rotational speed gain significantly improved SHE compared with empirical settings, with SHE above 21% of mean arterial pressure at flow rates of 1.0 L min(-1) to 1.5 L min(-1) and SHE above 5% at 3.5 L min(-1). Small diameter cannula (15 Fr) limited SHE compared with larger cannula (21 Fr and 23 Fr). Two diagonal pumps did not provide higher SHE than a single pump, but permitted additional control over pulse pressure and SHE by varying the total fraction of pulsatile flow and the fraction of flow bypassing the oxygenator. Conclusions Proper selection of components and optimizations of pump settings significantly improved pulse pressure and SHE of pulsatile MiECC. Surplus hemodynamic energy depended on flow rate with a maximum at 1.0 L min(-1)-1.5 L min(-1). Pulsatile MiECC may specifically assist organ perfusion during phases of low flow.
引用
收藏
页码:828 / 839
页数:12
相关论文
共 50 条
  • [31] A Novel Valveless Pulsatile Flow Pump for Extracorporeal Blood Circulation
    Anatol, Joaquin
    Vignali, Emanuele
    Gasparotti, Emanuele
    Castro-Ruiz, Francisco
    Rubio, Manuel
    Barrios-Collado, Cesar
    Sierra-Pallares, Jose
    Celi, Simona
    ANNALS OF BIOMEDICAL ENGINEERING, 2025, : 966 - 979
  • [32] Minimally invasive extracorporeal circulation Fact or fiction?
    El-Essawi, Aschraf
    Bauer, Adrian
    Groeger, Steffen
    Hausmann, Harald
    Gehron, Johannes
    Boening, Andreas
    Harringer, Wolfgang
    ZEITSCHRIFT FUR HERZ THORAX UND GEFASSCHIRURGIE, 2020, 34 (02): : 108 - 114
  • [33] A pulsatile pneumatically driven neonatal extracorporeal membrane oxygenation system using neck vessel cannulas tested with neonatal mock circulation
    Trittenwein, G
    Burda, G
    Trittenwein, H
    Golej, J
    Hermon, M
    Pollak, A
    ARTIFICIAL ORGANS, 2001, 25 (01) : 29 - 35
  • [34] CEREBRAL PERFUSION IN DOGS DURING PULSATILE AND NON-PULSATILE EXTRACORPOREAL-CIRCULATION
    DERNEVIK, L
    ARVIDSSON, S
    WILLIAMOLSSON, G
    JOURNAL OF CARDIOVASCULAR SURGERY, 1985, 26 (01): : 32 - 35
  • [35] Effect of minimal extracorporeal circulation versus conventional extracorporeal circulation on postoperative outcome of coronary artery bypass grafting
    Srivastava P.K.
    Rajan S.D.
    Agrawal S.
    Radakrishnan S.
    Varghese T.
    Cherian K.M.
    Indian Journal of Thoracic and Cardiovascular Surgery, 2010, 26 (1) : 1 - 4
  • [36] Coronary artery bypass graft with minimal extracorporeal circulation
    Folliguet, TA
    Villa, E
    Vandeneyden, F
    Laborde, F
    HEART SURGERY FORUM, 2003, 6 (05): : 297 - 301
  • [37] Minimal Extracorporeal Circulation and Minimally Invasive Valve Operations: Should They Be the Right Combination in the Future? Reply
    Rustenbach, Christian Joerg
    Wachter, Kristina
    Baumbach, Hardy
    ANNALS OF THORACIC SURGERY, 2017, 103 (03): : 1038 - 1039
  • [38] From minimal invasive extracorporeal circulation type I to type III: the perfusionist's view
    Jenni, Hansjoerg
    Heinisch, Paul Philipp
    Eberle, Balthasar
    Gisler, Fabian
    Kadner, Alexander
    Carrel, Thierry
    JOURNAL OF THORACIC DISEASE, 2019, 11 : S1515 - S1518
  • [39] Importance of minimal extracorporeal Circulation in the modern Coronary Surgery
    Liebold, A.
    ZEITSCHRIFT FUR HERZ THORAX UND GEFASSCHIRURGIE, 2011, 25 (01): : 19 - 20
  • [40] Minimal Extracorporeal Circulation: The Real Impact on Postoperative Outcome
    Ranucci, Marco