Improving adult pulsatile minimal invasive extracorporeal circulation in a mock circulation

被引:1
|
作者
Duerr, Anke [1 ]
Weber, Elena [1 ]
Eisenmann, Lisa [1 ]
Albrecht, Guenter [1 ]
Liebold, Andreas [1 ]
Hoenicka, Markus [1 ]
机构
[1] Ulm Univ, Dept Cardiothorac & Vasc Surg, Med Ctr, Ulm, Germany
关键词
cannula; diagonal pump; minimal invasive extracorporeal circulation; oxygenator; pulsatile perfusion; surplus hemodynamic energy; PERFUSION; FLOW; PUMP; METAANALYSIS;
D O I
10.1111/aor.14439
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background Pulsatile extracorporeal circulation (ECC) may improve perfusion of critical organs during cardiac surgery. This study analyzed the influence of the components of a minimal invasive ECC (MiECC) on the transfer of pulsatile energy into the pseudo-patient of a mock circulation. Methods An aortic model with human-like geometry and compliance was perfused by a diagonal pump. Surplus hemodynamic energy (SHE) was determined from flow and pressure data. Five adult-size oxygenator models and three sizes of cannulas were compared. Pulsatile pump settings were optimized, and parallel dual-pump configurations were evaluated. Results Oxygenator models showed up to twofold differences in pressure gradients and influenced SHE at flow rates up to 2.0 L min(-1). Adjustments of frequency, systole duration, and rotational speed gain significantly improved SHE compared with empirical settings, with SHE above 21% of mean arterial pressure at flow rates of 1.0 L min(-1) to 1.5 L min(-1) and SHE above 5% at 3.5 L min(-1). Small diameter cannula (15 Fr) limited SHE compared with larger cannula (21 Fr and 23 Fr). Two diagonal pumps did not provide higher SHE than a single pump, but permitted additional control over pulse pressure and SHE by varying the total fraction of pulsatile flow and the fraction of flow bypassing the oxygenator. Conclusions Proper selection of components and optimizations of pump settings significantly improved pulse pressure and SHE of pulsatile MiECC. Surplus hemodynamic energy depended on flow rate with a maximum at 1.0 L min(-1)-1.5 L min(-1). Pulsatile MiECC may specifically assist organ perfusion during phases of low flow.
引用
收藏
页码:828 / 839
页数:12
相关论文
共 50 条
  • [1] From less invasive to minimal invasive extracorporeal circulation
    Anastasiadis, Kyriakos
    Antonitsis, Polychronis
    Deliopoulos, Apostolos
    Argiriadou, Helena
    JOURNAL OF THORACIC DISEASE, 2021, 13 (03) : 1909 - 1921
  • [2] Effects of pulsatile minimal invasive extracorporeal circulation on fibrinolysis and organ protection in adult cardiac surgery -a prospective randomized trial
    Grassler, Angelika
    Bauernschmitt, Robert
    Guthoff, Irene
    Kunert, Andreas
    Hoenicka, Markus
    Albrecht, Guenter
    Liebold, Andreas
    JOURNAL OF THORACIC DISEASE, 2019, 11 : S1453 - S1463
  • [3] PULSATILE EXTRACORPOREAL-CIRCULATION
    KAGE, H
    MERTIG, D
    METZ, H
    KEMSMANN, U
    HELLBERG, K
    HASPER, B
    ARTIFICIAL ORGANS, 1977, 1 (02) : 124 - 124
  • [4] A Novel Minimal Invasive Mouse Model of Extracorporeal Circulation
    Luo, Shuhua
    Tang, Menglin
    Du, Lei
    Gong, Lina
    Xu, Jin
    Chen, Youwen
    Wang, Yabo
    Lin, Ke
    An, Qi
    MEDIATORS OF INFLAMMATION, 2015, 2015
  • [5] Quantification of Operational Learning in Minimal Invasive Extracorporeal Circulation
    Anastasiadis, Kyriakos
    Antonitsis, Polychronis
    Asteriou, Christos
    Argiriadou, Helena
    Deliopoulos, Apostolos
    Konstantinou, Dimitrios
    Grosomanidis, Vassilios
    Tossios, Paschalis
    ARTIFICIAL ORGANS, 2017, 41 (07) : 628 - 636
  • [6] Minimal invasive extracorporeal circulation preserves coagulation integrity
    Argiriadou, Helena
    Antonitsis, Polychronis
    Gkiouliava, Anna
    Papapostolou, Evangelia
    Deliopoulos, Apostolos
    Anastasiadis, Kyriakos
    PERFUSION-UK, 2022, 37 (03): : 257 - 265
  • [7] A Novel Minimal Invasive Mouse Model Of Extracorporeal Circulation
    Luo, S.
    Du, L.
    Chen, Y.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [8] PULSATILE BLOOD FLOW + EXTRACORPOREAL CIRCULATION
    SHEPARD, RB
    LYONS, C
    EDWARDS, WS
    REEVES, TJ
    STINECIP.PA
    CIRCULATION, 1964, 30 (4S3) : 158 - &
  • [9] Pulsatile augmentation device for extracorporeal circulation
    Doty, DB
    Wang, ND
    Chiang, BY
    Mohammad, SF
    ARTIFICIAL ORGANS, 1996, 20 (12) : 1304 - 1309
  • [10] The impact of minimal invasive extracorporeal circulation on postoperative kidney function
    Media, Ara Shwan
    Juhl-Olsen, Peter
    Magnusson, Nils Erik
    Modrau, Ivy Susanne
    PERFUSION-UK, 2021, 36 (07): : 745 - 750