Comparison of electrical impedance tomography and spirometry-based measures of airflow in healthy adult horses

被引:1
|
作者
Byrne, David P. [1 ]
Keeshan, Ben [2 ]
Hosgood, Giselle [1 ]
Adler, Andy [2 ]
Mosing, Martina [3 ]
机构
[1] Murdoch Univ, Sch Vet Med, Perth, WA, Australia
[2] Carleton Univ, Dept Syst & Comp Engn, Ottawa, ON, Canada
[3] Univ Vet Med, Dept Compan Anim & Horses, Anaesthesiol & Perioperat Intens Care, Vienna, Austria
关键词
asthma; equine; expiratory; inspiratory; peak flow; plethysmography; REGIONAL LUNG-FUNCTION; RESPIRATORY-FUNCTION; ANESTHETIZED HORSES; VENTILATION; OBSTRUCTION; ANESTHESIA; FREQUENCY; PRESSURE; HEAVES; MASK;
D O I
10.3389/fphys.2023.1164646
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Electrical impedance tomography (EIT) is a non-invasive diagnostic tool for evaluating lung function. The objective of this study was to compare respiratory flow variables calculated from thoracic EIT measurements with corresponding spirometry variables. Ten healthy research horses were sedated and instrumented with spirometry via facemask and a single-plane EIT electrode belt around the thorax. Horses were exposed to sequentially increasing volumes of apparatus dead space between 1,000 and 8,500 mL, in 5-7 steps, to induce carbon dioxide rebreathing, until clinical hyperpnea or a tidal volume of 150% baseline was reached. A 2-min stabilization period followed by 2 minutes of data collection occurred at each timepoint. Peak inspiratory and expiratory flow, inspiratory and expiratory time, and expiratory nadir flow, defined as the lowest expiratory flow between the deceleration of flow of the first passive phase of expiration and the acceleration of flow of the second active phase of expiration were evaluated with EIT and spirometry. Breathing pattern was assessed based on the total impedance curve. Bland-Altman analysis was used to evaluate the agreement where perfect agreement was indicated by a ratio of EIT:spirometry of 1.0. The mean ratio (bias; expressed as a percentage difference from perfect agreement) and the 95% confidence interval of the bias are reported. There was good agreement between EIT-derived and spirometry-derived peak inspiratory [-15% (-46-32)] and expiratory [10% (-32-20)] flows and inspiratory [-6% (-25-18)] and expiratory [5% (-9-20)] times. Agreement for nadir flows was poor [-22% (-87-369)]. Sedated horses intermittently exhibited Cheyne-Stokes variant respiration, and a breath pattern with incomplete expiration in between breaths (crown-like breaths). Electrical impedance tomography can quantify airflow changes over increasing tidal volumes and changing breathing pattern when compared with spirometry in standing sedated horses.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Exercise-induced airflow changes in horses with asthma measured by electrical impedance tomography
    Herteman, Nicolas
    Mosing, Martina
    Waldmann, Andreas D.
    Gerber, Vinzenz
    Schoster, Angelika
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2021, 35 (05) : 2500 - 2510
  • [2] Spirometry-Based Measures of Dysanapsis Among Adults with or at Risk of Obstructive Lung Diseases
    Kingdom, L.
    Kirby, M.
    Jensen, D.
    Benedettie, A.
    Vameghestahbanati, M.
    Aaron, S. D.
    Chapman, K. R.
    Doiron, D.
    Hernandez, P.
    Marciniuk, D. D.
    Maltais, F.
    O'Donnell, D. E.
    Walker, B.
    Tan, W.
    Bourbeau, J.
    Smith, B. M.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2022, 205
  • [3] Evaluation of the spatial spirometry indicators using electrical impedance tomography
    Kwok, Wang Chun
    Zouari, Fedi
    Cheung, Pak To
    Touboul, Adrien
    Sin, Venice
    Wong, Eddie C.
    Zhou, Iris Y.
    Tam, Terence Chi Chun
    Chan, Russell W.
    EUROPEAN RESPIRATORY JOURNAL, 2023, 62
  • [4] The optimal PEEP after alveolar recruitment maneuver assessed by electrical impedance tomography in healthy horses
    Rego Monteiro Andrade, Felipe Silveira
    Ambrosio, Aline Magalhaes
    Rodrigues, Renata Ramos
    Facco, Lara Lopes
    Goncalves, Lucas Alaiao
    Garcia Filho, Sergio Grandisoli
    dos Santos, Rosana Thurler
    Rossetto, Thais Colombo
    Amador Pereira, Marco Aurelio
    Fantoni, Denise Tabacchi
    FRONTIERS IN VETERINARY SCIENCE, 2022, 9
  • [5] Electrical impedance tomography to measure lung ventilation distribution in healthy horses and horses with left-sided cardiac volume overload
    Sacks, Muriel
    Byrne, David P.
    Herteman, Nicolas
    Secombe, Cristy J. Cristy
    Adler, Andy
    Hosgood, Giselle
    Raisis, Anthea L.
    Mosing, Martina
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2021, 35 (05) : 2511 - 2523
  • [6] Monitoring of tidal ventilation by electrical impedance tomography in anaesthetised horses
    Mosing, M.
    Waldmann, A. D.
    Raisis, A.
    Boehm, S. H.
    Drynan, E.
    Wilson, K.
    EQUINE VETERINARY JOURNAL, 2019, 51 (02) : 222 - 226
  • [7] Assessment of distribution of ventilation by electrical impedance tomography in standing horses
    Ambrisko, T. D.
    Schramel, J. P.
    Adler, A.
    Kutasi, O.
    Makra, Z.
    Moens, Y. P. S.
    PHYSIOLOGICAL MEASUREMENT, 2016, 37 (02) : 175 - 186
  • [8] Electrical Impedance Tomography: Tissue Properties to Image Measures
    Adler, Andy
    Boyle, Alistair
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (11) : 2494 - 2504
  • [9] Can bronchoconstriction and bronchodilatation in horses be detected using electrical impedance tomography?
    Secombe, Cristy
    Adler, Andy
    Hosgood, Giselle
    Raisis, Anthea
    Mosing, Martina
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2021, 35 (04) : 2035 - 2044
  • [10] Respiratory assessment of chronic lung disease patients using electrical impedance tomography, spirometry, and computational tomography
    Kwok, Wang Chun
    Zouari, Fedi
    Touboul, Adrien
    Cheung, Pak T.
    Sin, Venice
    Wong, Eddie C.
    Zhou, Iris Y.
    Tam, Terence C. C.
    Chan, Russell W.
    RESPIROLOGY, 2023, 28 : 96 - 96