Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy

被引:0
|
作者
Chirdchid, Thamonwan [1 ]
Ruangchan, Sirinya [2 ,3 ]
Sanghangthum, Taweap [1 ,2 ,4 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Fac Med, Bangkok, Thailand
[2] Her Royal Highness Princess Maha Chakri Sirindhor, Bangkok, Thailand
[3] Thai Red Cross Soc, King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, Bangkok, Thailand
[4] Chulalongkorn Univ, Div Radiat Oncol, Dept Radiol, Fac Med, Bangkok, Thailand
关键词
Dual-energy computed tomography; proton therapy; single-energy computed tomography; stoichiometric method; DOSE CALCULATION; CT; CALIBRATION;
D O I
10.4103/jmp.jmp_27_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. Materials and Methods: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose-volume histograms (DVHs) comparison. Results: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% +/- 2.1% and 2.5% +/- 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. Conclusion: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
引用
收藏
页码:292 / 297
页数:6
相关论文
共 50 条
  • [41] Dual-Energy Computed Tomography Applications in Neurointervention
    Wolman, Dylan N.
    Patel, Bhavik P.
    Wintermark, Max
    Heit, Jeremy J.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 831 - 839
  • [42] Cardiac Applications of Dual-Energy Computed Tomography
    Lempel, Matthew
    Frishman, William H.
    CARDIOLOGY IN REVIEW, 2019, 27 (04) : 208 - 210
  • [43] Dual-energy computed tomography: pediatric considerations
    Gallo-Bernal, Sebastian
    Pena-Trujillo, Valeria
    Gee, Michael S.
    PEDIATRIC RADIOLOGY, 2024, 54 (13) : 2112 - 2126
  • [44] In Vivo Measurement of Proton Stopping Power Ratios in Patients Using Dual Energy Computed Tomography
    Yang, M.
    Virshup, G.
    Clayton, J.
    Zhu, X.
    Mohan, R.
    Dong, L.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [45] Evaluation of Stopping-Power Prediction by Dual- and Single-Energy Computed Tomography in an Anthropomorphic Ground-Truth Phantom
    Wohlfahrt, Patrick
    Moehler, Christian
    Richter, Christian
    Greilich, Steffen
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 100 (01): : 244 - 253
  • [46] SINGLE-ENERGY VERSUS DUAL-ENERGY QUANTITATIVE COMPUTED-TOMOGRAPHY FOR SPINAL DENSITOMETRY IN PATIENTS WITH RHEUMATOID-ARTHRITIS
    LAAN, RFJM
    VANERNING, LJTO
    LEMMENS, JAM
    VANDEPUTTE, LBA
    RUIJS, SHJ
    VANRIEL, PLCM
    BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (778): : 901 - 904
  • [47] Effect of extended field-of-view approaches on the accuracy of stopping power ratio estimation for single-energy computed tomography simulators
    Lin, Chang-Shiun
    Tsai, Yi-Chun
    Chen, Liang-Hsin
    Wang, Chun-Wei
    Wu, Chia-Jung
    Chen, Wan-Yu
    Liang, Hsiang-Kung
    Kuo, Sung-Hsin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [48] Deriving the mean excitation energy map from dual-energy and proton computed tomography
    Vilches-Freixas, Gloria
    Quinones, Catherine Therese
    Letang, Jean Michel
    Rit, Simon
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 6 : 20 - 24
  • [49] Optimal Contrast of Computed Tomography Portal Venography Using Dual-Energy Computed Tomography
    Wang, Qi
    Shi, Gaofeng
    Liu, Xiang
    Wu, Runze
    Wang, Shijie
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (02) : 142 - 148
  • [50] Dual-Energy Computed Tomography to Photon Counting Computed Tomography: Emerging Technological Innovations
    Srinivas-Rao, Shravya
    Cao, Jinjin
    Marin, Daniele
    Kambadakone, Avinash
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 933 - 944