Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy

被引:0
|
作者
Chirdchid, Thamonwan [1 ]
Ruangchan, Sirinya [2 ,3 ]
Sanghangthum, Taweap [1 ,2 ,4 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Fac Med, Bangkok, Thailand
[2] Her Royal Highness Princess Maha Chakri Sirindhor, Bangkok, Thailand
[3] Thai Red Cross Soc, King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, Bangkok, Thailand
[4] Chulalongkorn Univ, Div Radiat Oncol, Dept Radiol, Fac Med, Bangkok, Thailand
关键词
Dual-energy computed tomography; proton therapy; single-energy computed tomography; stoichiometric method; DOSE CALCULATION; CT; CALIBRATION;
D O I
10.4103/jmp.jmp_27_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. Materials and Methods: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose-volume histograms (DVHs) comparison. Results: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% +/- 2.1% and 2.5% +/- 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. Conclusion: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
引用
收藏
页码:292 / 297
页数:6
相关论文
共 50 条
  • [21] Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy
    Murazaki, Hiroo
    Fukunaga, Junichi
    Hirose, Taka-aki
    Funatsu, Naomi
    Matsumoto, Ryoji
    Hidaka, Kyohei
    Nagamine, Shuji
    Nakanishi, Daiki
    Kato, Toyoyuki
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) : 268 - 276
  • [22] The influence of dual-energy computed tomography image noise in proton therapy treatment planning
    Nasmark, Torbjorn
    Andersson, Jonas
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 28
  • [23] Dual-energy computed tomography diagnostics for gout
    Rech, H. J.
    Cavallaro, A.
    ZEITSCHRIFT FUR RHEUMATOLOGIE, 2017, 76 (07): : 580 - 588
  • [24] Imaging of nanoparticles with dual-energy computed tomography
    Ducote, J. L.
    Alivov, Y.
    Molloi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2031 - 2044
  • [25] Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy
    Hiroo Murazaki
    Junichi Fukunaga
    Taka-aki Hirose
    Naomi Funatsu
    Ryoji Matsumoto
    Kyohei Hidaka
    Shuji Nagamine
    Daiki Nakanishi
    Toyoyuki Kato
    Radiological Physics and Technology, 2019, 12 : 268 - 276
  • [26] Dual-energy computed tomography of canine uroliths
    Nykamp, Stephanie G.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2017, 78 (10) : 1150 - 1155
  • [27] Dual-energy Computed Tomography Applications in Uroradiology
    Park, Jong
    Chandarana, Hersh
    Macari, Michael
    Megibow, Alec J.
    CURRENT UROLOGY REPORTS, 2012, 13 (01) : 55 - 62
  • [28] Dual-energy computed tomography: what is it useful for?
    Delgado Sanchez-Gracian, C.
    Martinez Rodriguez, C.
    Trinidad Lopez, C.
    RADIOLOGIA, 2013, 55 (04): : 346 - 352
  • [29] Dual-energy computed tomography in pulmonary embolism
    Lu, G-M
    Wu, S-Y
    Yeh, B. M.
    Zhang, L-J
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (992): : 707 - 718
  • [30] Dual-Energy Computed Tomography in Genitourinary Imaging
    Mileto, Achille
    Marin, Daniele
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2017, 55 (02) : 373 - +