Rapid T2*-weighted MRI using multishot EPI with retrospective motion and phase correction in the emergency department

被引:0
|
作者
Li, Zhiqiang [1 ,3 ]
Ooi, Melvyn B. [2 ]
Murchison, James A. [1 ]
Karis, John P. [1 ]
机构
[1] Barrow Neurol Inst, Dept Neuroradiol, Phoenix, AZ USA
[2] Philips Healthcare, Houston, TX USA
[3] St Josephs Hosp, Barrow Neurol Inst, Neurosci Publicat, 350 W Thomas Rd, Phoenix, AZ 85013 USA
关键词
emergency department; motion correction; multishot EPI; phase correction; T*(2)-weighted MRI; CEREBRAL MICROBLEEDS; ECHO; HEMORRHAGE; BRAIN; SENSE;
D O I
10.1002/mrm.29809
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Brain MRI is increasingly used in the emergency department (ED), where T*(2)-weighted MRI is an essential tool for detecting hemorrhage and stroke. The goal of this study was to develop a rapid T*(2)-weighted MRI technique capable of correcting motion-induced artifacts, thereby simultaneously improving scan time and motion robustness for ED applications.Methods: A 2D gradient-echo (GRE)-based multishot EPI (msEPI) technique was implemented using a navigator echo for estimating motion-induced errors. Bulk rigid head motion and phase errors were retrospectively corrected using an iterative conjugate gradient approach in the reconstruction pipeline. Three volunteers and select patients were imaged at 3 T and/or 1.5 T with an approximately 1-min full-brain protocol using the proposed msEPI technique and compared to an approximately 3-min standard-of-care GRE protocol to examine its performance.Results: Data from volunteers demonstrated that in-plane motion artifacts could be effectively corrected with the proposed msEPI technique, and through-plane motion artifacts could be mitigated. Patient images were qualitatively reviewed by one radiologist without a formal statistical analysis. These results suggested the proposed technique could correct motion-induced artifacts in the clinical setting. In addition, the conspicuity of susceptibility-related lesions using the proposed msEPI technique was comparable, or improved, compared to GRE.Conclusion: A 1-min full-brain T*(2)-weighted MRI technique was developed using msEPI with a navigator echo to correct motion-induced errors. Preliminary clinical results suggest faster scans and improved motion robustness and lesion conspicuity make msEPI a competitive alternative to traditional T*(2)-weighted MRI techniques for brain studies in the ED.
引用
收藏
页码:2500 / 2509
页数:10
相关论文
共 50 条
  • [11] Free-breathing myocardial T2*mapping using GRE-EPI and automatic Non-rigid motion correction
    Jin, Ning
    da Silveira, Juliana Serafim
    Jolly, Marie-Pierre
    Firmin, David N.
    Mathew, George
    Lamba, Nathan
    Subramanian, Sharath
    Pennell, Dudley J.
    Raman, Subha V.
    Simonetti, Orlando P.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2015, 17
  • [12] Free-breathing myocardial T2* mapping using GRE-EPI and automatic Non-rigid motion correction
    Ning Jin
    Juliana Serafim da Silveira
    Marie-Pierre Jolly
    David N. Firmin
    George Mathew
    Nathan Lamba
    Sharath Subramanian
    Dudley J. Pennell
    Subha V. Raman
    Orlando P. Simonetti
    Journal of Cardiovascular Magnetic Resonance, 17
  • [13] Free-breathing myocardial T2* mapping using GRE-EPI and automatic non-rigid motion correction
    Ning Jin
    Marie-Pierre Jolly
    Subha V Raman
    Orlando P Simonetti
    Journal of Cardiovascular Magnetic Resonance, 17 (Suppl 1)
  • [14] Artifact free T2*-weighted imaging at high spatial resolution using segmented EPI sequences
    Heiler, Patrick Michael
    Schmitter, Sebastian
    Schad, Lothar Rudi
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2010, 20 (03): : 166 - 174
  • [15] Rapid simultaneous acquisition of T1 and T2 mapping images using multishot double spin-echo EPI and automated variations of TR and TE (ms-DSEPI-T12)
    Liu, Xin
    Feng, Yi
    Lu, Zheng-Rong
    Morrell, Glen
    Jeong, Eun-Kee
    NMR IN BIOMEDICINE, 2010, 23 (01) : 97 - 104
  • [16] Model for the correction of motion-induced phase errors in multishot diffusion-weighted-MRI of the head: Are cardiac-motion-induced phase errors reproducible from beat-to-beat?
    O'Halloran, R. L.
    Holdsworth, S.
    Aksoy, M.
    Bammer, R.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (02) : 430 - 440
  • [17] Navigator-Guided Motion and B0 Correction of T2*-weighted MRI Improves Multiple Sclerosis Cortical Lesion Detection
    Beck, E. S.
    Liu, J.
    Filippini, S.
    van Gelderen, P.
    de Zwart, J. A.
    Norato, G.
    Sati, P.
    Al-Louzi, O.
    Kolb, H.
    Donadieu, M.
    Morrison, M.
    Duyn, J. H.
    Reich, D. S.
    MULTIPLE SCLEROSIS JOURNAL, 2021, 27 (1_SUPPL) : 70 - 71
  • [18] Cerebral blood volume measurements by rapid contrast infusion and T2*-weighted echo planar MRI
    Tudorica, A
    Li, HF
    Hospod, F
    Delucia-Deranja, E
    Huang, W
    Patlak, CS
    Newman, GC
    MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) : 1145 - 1157
  • [19] Whole-brain T2 mapping with radial sampling and retrospective motion correction at 3T
    Corbin, Nadege
    Trotier, Aurelien J.
    Anandra, Serge
    Kadalie, Emile
    Dallet, Laurence
    Miraux, Sylvain
    Ribot, Emeline J.
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (03) : 1026 - 1042
  • [20] Accelerated Simultaneous T2 and T2* Mapping of Multiple Sclerosis Lesions Using Compressed Sensing Reconstruction of Radial RARE-EPI MRI
    Herrmann, Carl J. J.
    Starke, Ludger
    Millward, Jason M.
    Kuchling, Joseph
    Paul, Friedemann
    Niendorf, Thoralf
    TOMOGRAPHY, 2023, 9 (01) : 299 - 314