Vaginal Applicators in High-Dose Rate Brachytherapy for Endometrial Carcinoma: A Dosimetric Study

被引:0
|
作者
Khan, Aijaz Ahmad [1 ,2 ]
Vijay, Anuj [1 ]
Wani, Shaqul Qamar [3 ]
Haq, Malik Mohib-ul [2 ]
机构
[1] GLA Univ, Inst Appl Sci & Humanities, Dept Phys, Chaumuhan, Uttar Pradesh, India
[2] Sherikashmir Inst Med Sci, Dept Radiol Phys & Bioengn, Srinagar, India
[3] Sherikashmir Inst Med Sci, Dept Radiat Oncol, Srinagar, India
来源
关键词
Brachytherapy; Elasto-Gel; overestimation of dose; treatment planning system; vaginal applicators; EXTERNAL-BEAM RADIOTHERAPY; QUALITY-OF-LIFE; RADIATION-THERAPY; STAGE-I; CANCER; PORTEC-2; TRIAL; AAPM; IR-192;
D O I
10.5505/tjo.2022.3765
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVE Intracavitary brachytherapy is one of the most important methods of gynecological cancer treatment. The effect of attenuation by the applicators is not considered in the dose calculation method released by the AAPM TG-43. In this study, the attenuation due to various segmented vaginal cylinders made up of polyetheretherketone was measured. METHODS A plan created by the treatment planning system (TPS) for the patient was executed using Gafchromic films that were taped on the surface of the applicator, and the dose recorded on the films was measured. The same plan was also executed using a well-type ionization chamber. First, a water equivalent material, Elasto-Gel was used in a well-type ionization chamber for charge collection, and then a segmented vaginal cylinder was used for charge collection measurement; thus, the attenuation due to the various segmented vaginal cylinders was obtained. RESULTS The doses recorded on the Gafchromic films were measured and showed that the TPS overestimated the dose for the segmented vaginal applicators and that the maximum overestimation of the dose was 4.75% for 35 mm diameter vaginal cylinders. This was further confirmed by calculating the percentage attenuation due to vaginal cylinders using a well-type ionization chamber. CONCLUSION Due to the attenuation by the various applicators used in vaginal vault brachytherapy, it can be concluded that due to overestimation of dose by the TPS, the dose actually received by the target and the organs at risk in vaginal vault brachytherapy are less and needs to be considered for actual dose estimation.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 50 条
  • [41] The postoperative vaginal high dose rate brachytherapy for endometrial cancer -: Outcome and quality of life
    Hänggi, M
    Ulmer, U
    Kubli, M
    Lütolf, U
    Huguenin, P
    STRAHLENTHERAPIE UND ONKOLOGIE, 2005, 181 : 141 - 141
  • [42] 3D-printed applicators for high dose rate brachytherapy: Dosimetric assessment at different infill percentage
    Ricotti, Rosalinda
    Vavassori, Andrea
    Bazani, Alessia
    Ciardo, Delia
    Pansini, Floriana
    Spoto, Ruggero
    Sammarco, Vittorio
    Cattani, Federica
    Baroni, Guido
    Orecchia, Roberto
    Jereczek-Fossa, Barbara Alicja
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (12): : 1698 - 1706
  • [43] The importance of immobilization of gynecological applicators in high dose rate brachytherapy
    Pella, S.
    Mikko, H.
    Nicolae, D.
    Casey, C.
    Shereen, C.
    Janeil, P.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S936 - S936
  • [44] Prognostic factors on adjuvant radiotherapy of endometrial carcinoma: experience with external beam radiotherapy and high-dose rate brachytherapy
    Novaes, P
    Chen, M
    Nishimoto, I
    Pellizzon, A
    Ferrigno, R
    Fogaroli, R
    Maia, M
    Salvajoli, J
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S23 - S23
  • [45] Adaptive High-Dose Rate Brachytherapy in the Setting of Recurrent Nasal Carcinoma
    Pursley, J.
    Chan, A.
    Nguyen, T.
    Ruggieri, T.
    Mauceri, T.
    MEDICAL PHYSICS, 2019, 46 (06) : E333 - E333
  • [46] High-dose rate brachytherapy in the treatment of early stages of penile carcinoma
    Pohankova, D.
    Sirak, I.
    Kasaova, L.
    Grepl, J.
    Paluska, P.
    Louda, M.
    Holub, L.
    Spacek, J.
    Prosvic, P.
    Petera, J.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S636 - S637
  • [47] Dosimetric comparison of perineal and intra-vaginal interstitial template in image guided high dose rate brachytherapy for carcinoma cervix
    Kaliyaperumal, V.
    Banerjee, S.
    Kataria, T.
    Abraham, S. K.
    Tamilselvan, S.
    Gupta, D.
    Dayanithi, K.
    Manigandan, D.
    Mishra, S. R.
    Bisht, S. S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (03): : 593 - 600
  • [48] Comparison of high-dose-rate and low-dose-rate brachytherapy in the treatment of endometrial carcinoma
    Fayed, Alaa
    Mutch, David G.
    Rader, Janet S.
    Gibb, Randall K.
    Powell, Matthew A.
    Wright, Jason D.
    ElNaqa, Issam
    Zoberi, Imran
    Grigsby, Perry W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (02): : 480 - 484
  • [49] Endometrial cancer recurrence: Adjuvant high-dose rate vaginal brachytherapy (HDR) versus HDR with whole pelvic radiation.
    Villella, Jeannine A.
    Palaia, Thomas
    Chalas, Eva
    Patrick, Patricia
    Haas, Jonathan A.
    JOURNAL OF CLINICAL ONCOLOGY, 2012, 30 (15)
  • [50] Inter-fraction movements of brachytherapy cylinder applicators during high dose rate Ir-192 vaginal vault treatments for endometrial cancer
    French, CA
    Al-Booz, H
    Boiangiu, I
    Cornes, PG
    RADIOTHERAPY AND ONCOLOGY, 2005, 75 : S44 - S44