Vaginal Applicators in High-Dose Rate Brachytherapy for Endometrial Carcinoma: A Dosimetric Study

被引:0
|
作者
Khan, Aijaz Ahmad [1 ,2 ]
Vijay, Anuj [1 ]
Wani, Shaqul Qamar [3 ]
Haq, Malik Mohib-ul [2 ]
机构
[1] GLA Univ, Inst Appl Sci & Humanities, Dept Phys, Chaumuhan, Uttar Pradesh, India
[2] Sherikashmir Inst Med Sci, Dept Radiol Phys & Bioengn, Srinagar, India
[3] Sherikashmir Inst Med Sci, Dept Radiat Oncol, Srinagar, India
来源
关键词
Brachytherapy; Elasto-Gel; overestimation of dose; treatment planning system; vaginal applicators; EXTERNAL-BEAM RADIOTHERAPY; QUALITY-OF-LIFE; RADIATION-THERAPY; STAGE-I; CANCER; PORTEC-2; TRIAL; AAPM; IR-192;
D O I
10.5505/tjo.2022.3765
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVE Intracavitary brachytherapy is one of the most important methods of gynecological cancer treatment. The effect of attenuation by the applicators is not considered in the dose calculation method released by the AAPM TG-43. In this study, the attenuation due to various segmented vaginal cylinders made up of polyetheretherketone was measured. METHODS A plan created by the treatment planning system (TPS) for the patient was executed using Gafchromic films that were taped on the surface of the applicator, and the dose recorded on the films was measured. The same plan was also executed using a well-type ionization chamber. First, a water equivalent material, Elasto-Gel was used in a well-type ionization chamber for charge collection, and then a segmented vaginal cylinder was used for charge collection measurement; thus, the attenuation due to the various segmented vaginal cylinders was obtained. RESULTS The doses recorded on the Gafchromic films were measured and showed that the TPS overestimated the dose for the segmented vaginal applicators and that the maximum overestimation of the dose was 4.75% for 35 mm diameter vaginal cylinders. This was further confirmed by calculating the percentage attenuation due to vaginal cylinders using a well-type ionization chamber. CONCLUSION Due to the attenuation by the various applicators used in vaginal vault brachytherapy, it can be concluded that due to overestimation of dose by the TPS, the dose actually received by the target and the organs at risk in vaginal vault brachytherapy are less and needs to be considered for actual dose estimation.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 50 条
  • [1] High-dose rate vaginal brachytherapy in early stage endometrial cancer
    Perrucci, E.
    Lancellotta, V.
    Bini, V.
    Mariucci, C.
    Di Benedetto, M.
    Trinari, S.
    Verzini, S.
    Aristei, C.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S549 - S550
  • [2] Dosimetric impact of cylinder size in high-dose rate vaginal cuff brachytherapy (VCBT) for primary endometrial cancer
    Zhang, Hualin
    Gopalakrishnan, Mahesh
    Lee, Plato
    Kang, Zhuang
    Sathiaseelan, Vythialingam
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 262 - 272
  • [3] Dosimetric Evaluation of Gammamed High Dose Rate Intraluminal Brachytherapy Applicators
    Chandraraj, V.
    Supe, S.
    Manickam, R.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [4] Dosimetric evaluation of Gammamed High Dose Rate intraluminal brachytherapy applicators
    Supe, Sanjay Sudhakar
    Varatharaj, Chandraraj
    Bijina, Themantavida Krishnan
    Bondel, Shwetha
    Sathiyan, Saminathan
    Ganesh, Kadirampatti Mani
    Arunkumar, Thangaraj
    Ravikumar, Manickam
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2007, 12 (06) : 313 - 317
  • [5] DOSIMETRIC EFFECTS OF AIR POCKETS AROUND HIGH-DOSE RATE BRACHYTHERAPY VAGINAL CYLINDERS
    Richardson, Susan
    Palaniswaamy, Geethpriya
    Grigsby, Perry W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (01): : 276 - 279
  • [6] Treatment of Vaginal Recurrences in Endometrial Carcinoma by High-dose-rate Brachytherapy
    Sorbe, Bengt
    Soderstrom, Karin
    ANTICANCER RESEARCH, 2013, 33 (01) : 241 - 247
  • [7] Dosimetric comparison between different applicators for high dose rate gynecologic brachytherapy
    Portocarrero-Bonifaz, Andres
    Syed, Salman
    Kassel, Maxwell
    McKenzie, Grant W.
    Shah, Vishwa M.
    Forry, Bryce
    Masters, Adrianna H.
    Gaskins, Jeremy T.
    Sowards, Keith T.
    Silva, Scott R.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S260 - S262
  • [8] High-dose rate brachytherapy for selected patients with stage I endometrial carcinoma
    Niazi, MT
    Souhami, L
    Stanimir, GW
    Portelance, L
    Bahoric, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S239 - S240
  • [9] Prophylactic vaginal high dose rate brachytherapy after pelvic irradiation in endometrial carcinoma
    Carrascosa, M
    Tortajada, M
    Guinot, J
    Aguayo, M
    Beato, I
    Escolar, P
    Mengual, J
    Casana, M
    Pesudo, C
    Arribas, L
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S22 - S22
  • [10] High dose rate vaginal brachytherapy in early stage endometrial carcinoma: Preliminary analysis
    Hong, JJ
    Yacoub, SF
    Blevins, JC
    Kuban, DA
    ElMahdi, AM
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 1997, 20 (03): : 254 - 258