Fretting corrosion testing of acetabular modular tapers for total hip replacements: A comparison of two designs

被引:2
|
作者
Mace, Annsley [1 ]
Goodwin, Charley M. [1 ]
Gilbert, Jeremy L. [1 ]
机构
[1] Clemson Univ, MUSC Bioengn Program, Charleston, SC 29634 USA
关键词
acetabular implant testing; acetabular taper; fretting corrosion; modular orthopedic implant; total hip arthroplasty; RETRIEVAL; BACKSIDE; LINER; ALLOY;
D O I
10.1002/jor.25512
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Acetabular components (DePuy Pinnacle (A) and Stryker Trident (B), Ti-6Al-4V shells and CoCrMo liners) with varying geometries were assembled under a 4 kN seating load. Liner-displacement was recorded. Cyclic compression to 4 kN, R = 0.01, 9 Hz was applied for three million cycles to evaluate fretting corrosion currents (n = 5). Fretting currents, load-displacement, ion dissolution, and disassembly loads were used to compare device performance. Data were analyzed using ANOVA with Tukey post hoc comparisons (p < 0.05). Liner seating displacements were not significantly different between groups. Fretting currents averaged over the initial 10 h and over three million cycles were 0.17 mu A (A) and 0.55 mu A (B) and 0.05 mu A (A) and 0.17 mu A (B), respectively (p = 0.19). No variation in ion averages between A and B (0.23 and 0.45 ppm for Ti [p = 0.21], 0.63 and 0.85 ppm for Co [p = 0.47]) existed. Average push-out forces, -2.41 (A) and -2.42 kN (B), were not significantly different (p = 0.97). SEM and EDS showed some titanium and metal oxide transfer from the shell to the liner in both designs. Overall, both implant designs exhibited very minor MACC in these experiments. This study demonstrates quantitative measures of in vitro fretting corrosion over the course of three million cycles and the minimal degree of acetabular taper damage. Clinical Significance: Retrieval studies show dual mobility acetabular shell-liner tapers with metal-on-metal contacts are susceptible to fretting corrosion in vivo.
引用
收藏
页码:1738 / 1745
页数:8
相关论文
共 50 条
  • [21] Enhanced Wear and Corrosion in Modular Tapers in Total Hip Replacement Is Associated With the Contact Area and Surface Topography
    Panagiotidou, Anna
    Meswania, Jay
    Hua, Jia
    Muirhead-Allwood, Sarah
    Hart, Alister
    Blunn, Gordon
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (12) : 2032 - 2039
  • [22] Corrosion of Modular Tapers in Total Joint Replacements: A Critical Assessment of Design, Materials, Surface Structure, Mechanics, Electrochemistry, and Biology
    Gilbert, Jeremy L.
    Mali, Sachin A.
    Sivan, Shiril
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 192 - 223
  • [23] Early catastrophic acetabular failure in Furlong total hip replacements
    Knox D.
    Hamilton S.W.
    Wardlaw D.
    Gibson P.H.
    Journal of Orthopaedics and Traumatology, 2009, 10 (1) : 39 - 42
  • [24] Compliant interfaces and fretting corrosion of modular taper junctions in total hip implants: The micromechanics of contact
    Smith, Stephanie M.
    Gilbert, Jeremy L.
    TRIBOLOGY INTERNATIONAL, 2020, 151
  • [25] In Vitro Evidence for Cell-Accelerated Corrosion Within Modular Junctions of Total Hip Replacements
    Bijukumar, Divya Rani
    Salunkhe, Shruti
    Morris, Dalton
    Segu, Abhijith
    Hal, Deborah J.
    Pourzal, Robin
    Mathew, Mathew T.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2020, 38 (02) : 393 - 404
  • [26] Corrosion testing of modular hip implants
    Goldberg, JR
    Buckley, CA
    Jacobs, JJ
    Gilbert, JL
    MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 : 157 - 176
  • [27] THE RESULTS WITH 2 TYPES OF UNCEMENTED ACETABULAR CUPS IN TOTAL HIP REPLACEMENTS
    SCHMITT, D
    BIGARD, O
    BRESLER, F
    MOLE, D
    BERNARD, P
    INTERNATIONAL ORTHOPAEDICS, 1993, 17 (06) : 357 - 364
  • [28] Fretting and Corrosion at the Backside of Modular Cobalt Chromium Acetabular Inserts: A Retrieval Analysis
    Tarity, T. David
    Koch, Chelsea N.
    Burket, Jayme C.
    Wright, Timothy M.
    Westrich, Geoffrey H.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (03): : 1033 - 1039
  • [29] Severe Corrosion of Modular Dual Mobility Acetabular Components Identified During Revision Total Hip Arthroplasty
    Abdelaal, Mohammad S.
    Zachwieja, Eric
    Sharkey, Peter F.
    ARTHROPLASTY TODAY, 2021, 8 : 78 - 83
  • [30] In vitro comparison of the two hard-hard articulations for total hip replacements
    Rieker, C
    Konrad, R
    Schön, R
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H2) : 153 - 160