Noninvasive blood pressure measurement in patients with continuous-flow left ventricular assist devices

被引:1
|
作者
Tanaka, Aiko [1 ]
Kiriyama, Yuki [2 ]
Kubo, Naoko [1 ]
Sakaguchi, Ryota [1 ]
Uchiyama, Akinori [1 ]
Fujino, Yuji [1 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Anesthesiol & Intens Care Med, 2-15 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Natl Org Osaka Natl Hosp, Div Anesthesiol, Chuo Ku, 2-1-14 Hoenzaka, Osaka 5400006, Japan
关键词
Blood pressure; Noninvasive measurement; Left ventricular assist device; Doppler ultrasound;
D O I
10.1007/s10047-022-01349-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillatory blood pressure (OBP) with a slow cuff-deflation system has been proposed as noninvasive measurement of mean arterial pressure (MAP) in patients with continuous-flow left ventricular assist devices (LVADs). However, the challenge is that the measurement is not obtainable in certain patients. We hypothesized that the combined use of color Doppler imaging during OBP measurement (CDBP) could derive MAP accurately. We conducted a prospective observational study in critically ill patients (30 patients with continuous-flow LVADs and 30 control patients without LVADs). Triplicate OBP and CDBP measurements were performed and invasive blood pressure (IBP) was recorded. The overall success rate of OBP was 63.3% in the LVAD group and 98.9% in the control group. The CDBP was successfully obtained in 100% of all study patients. The CDBP in the LVAD group was closest to the MAP of measured IBP, while that in the control group was closest to the systolic IBP. The mean absolute differences in OBP and CDBP from the closest IBP were similar in both the control and LVAD groups. In nonpulsatile LVAD patients with a pulse pressure IBP < 10 mmHg, the success rate of OBP measurement was only 10.0%, and CDBP showed significantly reduced error in MAP measurement (mean absolute difference: OBP 23.2 +/- 8.7 vs CDBP 5.2 +/- 3.6 mmHg, p < 0.001). The validity of OBP measurement with a slow cuff-deflation system limited particularly in nonpulsatile LVAD patients. The concurrent use of color Doppler imaging is encouraged for more accurate measurement of MAP in patients with continuous-flow LVADs.
引用
收藏
页码:160 / 164
页数:5
相关论文
共 50 条
  • [21] Gastrointestinal bleeding with continuous-flow left ventricular assist devices
    Amer S.
    Shah P.
    Hassan S.
    Clinical Journal of Gastroenterology, 2015, 8 (2) : 63 - 67
  • [22] Perils of bridge to transplantation in patients on continuous-flow left ventricular assist devices
    Moazami, Nader
    Reyentovich, Alex
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04): : 481 - 482
  • [23] Allograft Rejection in Patients Supported With Continuous-Flow Left Ventricular Assist Devices
    Healy, Aaron H.
    Mason, Nathan O.
    Hammond, M. Elizabeth
    Reid, Bruce B.
    Clayson, Stephen E.
    Drakos, Stavros G.
    Kfoury, Abdallah G.
    Patel, Amit N.
    Bull, David A.
    Budge, Deborah
    Alharethi, Rami A.
    Bader, Feras M.
    Gilbert, Edward M.
    Stehlik, Josef
    Selzman, Craig H.
    ANNALS OF THORACIC SURGERY, 2011, 92 (05): : 1601 - 1607
  • [24] Prevalence of Cerebral Microbleeds in Patients With Continuous-Flow Left Ventricular Assist Devices
    Yoshioka, Daisuke
    Okazaki, Shuhei
    Toda, Koichi
    Murase, Sho
    Saito, Shunsuke
    Domae, Keitaro
    Miyagawa, Shigeru
    Yoshikawa, Yasushi
    Daimon, Takashi
    Sakaguchi, Manabu
    Sawa, Yoshiki
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2017, 6 (09):
  • [25] Factor Xa inhibitors in patients with continuous-flow left ventricular assist devices
    Parikh, Vishal Y.
    Parikh, Umang M.
    Moctezuma-Ramirez, Angel
    Lamba, Harveen K.
    George, Joggy K.
    Fedson, Savitri
    Nair, Ajith P.
    Frazier, O. H.
    Delgado, Reynolds M.
    GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2020, 68 (11) : 1278 - 1284
  • [26] Neurologic Events in Continuous-Flow Left Ventricular Assist Devices
    Kadakkal, Ajay
    Najjar, Samer S.
    CARDIOLOGY CLINICS, 2018, 36 (04) : 531 - +
  • [27] Physiologic effects of continuous-flow left ventricular assist devices
    Healy, Aaron H.
    McKellar, Stephen H.
    Drakos, Stavros G.
    Koliopoulou, Antigoni
    Stehlik, Josef
    Selzman, Craig H.
    JOURNAL OF SURGICAL RESEARCH, 2016, 202 (02) : 363 - 371
  • [28] Measurement of Blood Pressure during Support with a Continuous-Flow Left Ventricular Assist Device in the Outpatient Setting
    Coyle, L. A.
    Gallagher, C.
    Maier, J.
    Bhat, G.
    Pappas, P.
    Tatooles, A.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S235 - S235
  • [29] A Decade of Experience With Continuous-Flow Left Ventricular Assist Devices
    John, Ranjit
    Holley, Christopher T.
    Eckman, Peter
    Roy, Samit S.
    Cogswell, Rebecca
    Harvey, Laura
    Shumway, Sara
    Liao, Kenneth
    SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2016, 28 (02) : 363 - 374
  • [30] Metabolic Dysfunction in Continuous-Flow Left Ventricular Assist Devices Patients and Outcomes
    Nguyen, Ann B.
    Imamura, Teruhiko
    Besser, Stephanie
    Rodgers, Daniel
    Chung, Ben
    Raikhelkar, Jayant
    Kalantari, Sara
    Smith, Bryan
    Sarswat, Nitasha
    LaBuhn, Colleen
    Jeevanandam, Valluvan
    Kim, Gene
    Sayer, Gabriel
    Uriel, Nir
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2019, 8 (22):