Noninvasive blood pressure measurement in patients with continuous-flow left ventricular assist devices

被引:1
|
作者
Tanaka, Aiko [1 ]
Kiriyama, Yuki [2 ]
Kubo, Naoko [1 ]
Sakaguchi, Ryota [1 ]
Uchiyama, Akinori [1 ]
Fujino, Yuji [1 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Anesthesiol & Intens Care Med, 2-15 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Natl Org Osaka Natl Hosp, Div Anesthesiol, Chuo Ku, 2-1-14 Hoenzaka, Osaka 5400006, Japan
关键词
Blood pressure; Noninvasive measurement; Left ventricular assist device; Doppler ultrasound;
D O I
10.1007/s10047-022-01349-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillatory blood pressure (OBP) with a slow cuff-deflation system has been proposed as noninvasive measurement of mean arterial pressure (MAP) in patients with continuous-flow left ventricular assist devices (LVADs). However, the challenge is that the measurement is not obtainable in certain patients. We hypothesized that the combined use of color Doppler imaging during OBP measurement (CDBP) could derive MAP accurately. We conducted a prospective observational study in critically ill patients (30 patients with continuous-flow LVADs and 30 control patients without LVADs). Triplicate OBP and CDBP measurements were performed and invasive blood pressure (IBP) was recorded. The overall success rate of OBP was 63.3% in the LVAD group and 98.9% in the control group. The CDBP was successfully obtained in 100% of all study patients. The CDBP in the LVAD group was closest to the MAP of measured IBP, while that in the control group was closest to the systolic IBP. The mean absolute differences in OBP and CDBP from the closest IBP were similar in both the control and LVAD groups. In nonpulsatile LVAD patients with a pulse pressure IBP < 10 mmHg, the success rate of OBP measurement was only 10.0%, and CDBP showed significantly reduced error in MAP measurement (mean absolute difference: OBP 23.2 +/- 8.7 vs CDBP 5.2 +/- 3.6 mmHg, p < 0.001). The validity of OBP measurement with a slow cuff-deflation system limited particularly in nonpulsatile LVAD patients. The concurrent use of color Doppler imaging is encouraged for more accurate measurement of MAP in patients with continuous-flow LVADs.
引用
收藏
页码:160 / 164
页数:5
相关论文
共 50 条
  • [1] Noninvasive blood pressure measurement in patients with continuous-flow left ventricular assist devices
    Aiko Tanaka
    Yuki Kiriyama
    Naoko Kubo
    Ryota Sakaguchi
    Akinori Uchiyama
    Yuji Fujino
    Journal of Artificial Organs, 2023, 26 : 160 - 164
  • [2] Blood Pressure Measurement in Patients with Continuous-Flow Left Ventricular Assist Devices
    Woldendorp, K. D.
    Gupta, S.
    Markey, P.
    Robson, D.
    Dhital, K.
    Hayward, C. S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S91 - S92
  • [3] Noninvasive measurement of arterial blood pressure in patients with continuous-flow left ventricular assist devices: a systematic review
    Paulino A. Alvarez
    Rakesh Ponnapureddy
    Dinesh Voruganti
    Ernesto Ruiz Duque
    Alexandros Briasoulis
    Heart Failure Reviews, 2021, 26 : 47 - 55
  • [4] Noninvasive measurement of arterial blood pressure in patients with continuous-flow left ventricular assist devices: a systematic review
    Alvarez, Paulino A.
    Ponnapureddy, Rakesh
    Voruganti, Dinesh
    Duque, Ernesto Ruiz
    Briasoulis, Alexandros
    HEART FAILURE REVIEWS, 2021, 26 (01) : 47 - 55
  • [5] Noninvasive Arterial Blood Pressure Waveforms in Patients with Continuous-Flow Left Ventricular Assist Devices
    Martina, Jerson R.
    Westerhof, Berend E.
    de Jonge, Nicolaas
    van Goudoever, Jeroen
    Westers, Paul
    Chamuleau, Steven
    van Dijk, Diederik
    Rodermans, Ben F. M.
    de Mol, Bas A. J. M.
    Lahpor, Jaap R.
    ASAIO JOURNAL, 2014, 60 (02) : 154 - 161
  • [6] A novel method of blood pressure measurement in patients with continuous-flow left ventricular assist devices
    Woldendorp, Kei
    Gupta, Sunil
    Lai, Jacqueline
    Dhital, Kumud
    Hayward, Christopher S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (11): : 1183 - 1186
  • [7] Measuring Blood Pressure in Patients with Continuous-Flow Left Ventricular Assist Devices
    Kerk, K. L.
    Truong, T. H.
    Lim, C. H.
    Tan, T. E.
    Sim, D.
    Lim, C. P.
    Sivathasan, C.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04): : S340 - S340
  • [8] Noninvasive Blood Pressure Monitor Designed for Patients With Heart Failure Supported with Continuous-Flow Left Ventricular Assist Devices
    Sajgalik, Pavol
    Kremen, Vaclav
    Fabian, Vratislav
    Maltais, Simon
    Stulak, John M.
    Kushwaha, Sudhir S.
    Joyce, Lyle D.
    Schirger, John A.
    Johnson, Bruce D.
    ASAIO JOURNAL, 2019, 65 (02) : 127 - 133
  • [9] Accuracy of Doppler blood pressure measurement in continuous-flow left ventricular assist device patients
    Li, Song
    Beckman, Jennifer A.
    Welch, Nathan G.
    Bjelkengren, Jason
    Masri, Sofia Carolina
    Minami, Elina
    Stempien-Otero, April
    Levy, Wayne C.
    O'Brien, Kevin D.
    Lin, Shin
    Farris, Stephen D.
    Chen, Richard K.
    Wood, Gregory
    Koomalsingh, Kevin
    Kirkpatrick, James
    McCabe, James
    Leary, Peter J.
    Chassagne, Fanette
    Chivukula, Venkat Keshav
    Aliseda, Alberto
    Mahr, Claudius
    ESC HEART FAILURE, 2019, 6 (04): : 793 - 798
  • [10] Ventricular Arrhythmias in Patients With Continuous-flow Left Ventricular Assist Devices
    Kuten, David
    McLean, Rhondalyn
    Russell, Stuart
    CIRCULATION, 2013, 128 (22)