PET imaging and quantification of small animals using a clinical SiPM-based camera

被引:2
|
作者
Desmonts, Cedric [1 ,2 ]
Lasnon, Charline [2 ,3 ]
Jaudet, Cyril [4 ]
Aide, Nicolas [2 ]
机构
[1] Univ Hosp Caen, Nucl Med Dept, Ave Cote Nacre, F-14033 Caen 9, France
[2] Normandy Univ, UNICAEN, INSERM 1086 ANTICIPE, Caen, France
[3] Comprehens Canc Ctr F Baclesse, Nucl Med Dept, UNICANC, Caen, France
[4] Comprehens Canc Ctr F Baclesse, Radiophys Dept, UNICANC, Caen, France
关键词
SiPM PET; Preclinical research; Small-animal imaging; PET quantification; EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY; PERFORMANCE EVALUATION; QUANTITATION; IMPACT;
D O I
10.1186/s40658-023-00583-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Small-animal PET imaging is an important tool in preclinical oncology. This study evaluated the ability of a clinical SiPM-PET camera to image several rats simultaneously and to perform quantification data analysis.Methods: Intrinsic spatial resolution was measured using 18F line sources, and image quality was assessed using a NEMA NU 4-2018 phantom. Quantification was evaluated using a fillable micro-hollow sphere phantom containing 4 spheres of different sizes (ranging from 3.95 to 7.86 mm). Recovery coefficients were computed for the maximum (Amax) and the mean (A50) pixel values measured on a 50% isocontour drawn on each sphere. Measurements were performed first with the phantom placed in the centre of the field of view and then in the off-centre position with the presence of three scattering sources to simulate the acquisition of four animals simultaneously. Quantification accuracy was finally validated using four 3D-printed phantoms mimicking rats with four subcutaneous tumours each. All experiments were performed for both 18F and 68Ga radionuclides.Results: Radial spatial resolutions measured using the PSF reconstruction algorithm were 1.80 mm and 1.78 mm for centred and off-centred acquisitions, respectively. Spill-overs in air and water and uniformity computed with the NEMA phantom centred in the FOV were 0.05, 0.1 and 5.55% for 18F and 0.08, 0.12 and 2.81% for 68Ga, respectively. Recovery coefficients calculated with the 18F-filled micro-hollow sphere phantom for each sphere varied from 0.51 to 1.43 for Amax and from 0.40 to 1.01 for A50. These values decreased from 0.28 to 0.92 for Amax and from 0.22 to 0.66 for A50 for 68 Ga acquisition. The results were not significantly different when imaging phantoms in the off-centre position with 3 scattering sources. Measurements performed with the four 3D-printed phantoms showed a good correlation between theoretical and measured activity in simulated tumours, with r(2) values of 0.99 and 0.97 obtained for 18F and 68Ga, respectively.Conclusion: We found that the clinical SiPM-based PET system was close to that obtained with a dedicated small-animal PET device. This study showed the ability of such a system to image four rats simultaneously and to perform quantification analysis for radionuclides commonly used in oncology.
引用
收藏
页数:19
相关论文
共 50 条
  • [31] A fast preamplifier concept for SiPM-based time-of-flight PET detectors
    Huizenga, J.
    Seifert, S.
    Schreuder, F.
    van Dam, H. T.
    Dendooven, P.
    Lohner, H.
    Vinke, R.
    Schaart, D. R.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 695 : 379 - 384
  • [32] Performance Evaluation of the SiPM-based Siemens Biograph Vision PET/CT System
    Reddin, Janet S.
    Scheuermann, Joshua S.
    Bharkhada, Deepak
    Smith, Anne M.
    Casey, Michael E.
    Conti, Maurizio
    Karp, Joel S.
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [33] A study on performance boost of a 17 m class Cherenkov telescope with a SiPM-based camera
    Arcaro, Cornelia
    Doro, Michele
    Sitarek, Julian
    Baack, Dominik
    ASTROPARTICLE PHYSICS, 2024, 155
  • [34] High Quality Imaging and Dosimetry of Yttrium-90 (90Y) Radioembolization Using a SiPM-Based PET/CT scanner
    Duan, Heving
    Khalaf, Mohamed Hatem
    Baratto, Lucia
    Sze, Daniel
    Srinivas, Shyam
    Iagaru, Andrei
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [35] Clinical Follow-Up after Imaging and Dosimetry for Yttrium-90 (90Y) Liver Radioembolization Using a SiPM-Based PET/CT Scanner
    Duan, Heying
    Khalaf, Mohamed
    Baratto, Lucia
    Srinivas, Shyam
    Sze, Daniel
    Iagaru, Andrei
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [36] Clinical Follow-up after Imaging and Dosimetry for Yttrium-90 (90Y) Liver Radioembolization Using a SiPM-based PET/CT Scanner
    Duan, H.
    Khalaf, M. H.
    Baratto, L.
    Srinivas, S.
    Sze, D.
    Iagaru, A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S259 - S259
  • [37] Performance comparison of two signal multiplexing readouts for SiPM-based pet detector
    Yang, Qian
    Kuang, Zhonghua
    Sang, Ziru
    Yang, Yongfeng
    Du, Junwei
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (23):
  • [38] Revisiting physiological FDG uptake throughout SiPM-based PET/CT scanner
    Toriihara, Akira
    Yoshida, Katsuya
    Matsubara, Daigoro
    Harata, Naoki
    Isogai, Jun
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [39] A Statistical DOI Estimation Algorithm for a SiPM-Based Clinical SPECT Insert
    D'Adda, Ilenia
    Morahan, Ashley J.
    Carminati, Marco
    Erlandsson, Kjell
    Ljungberg, Michael
    Hutton, Brian F.
    Fiorini, Carlo
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (07) : 771 - 777
  • [40] Clinical SiPM-Based MRI-Compatible SPECT: Preliminary Characterization
    Carminati, M.
    D'Adda, I
    Morahan, A. J.
    Erlandsson, K.
    Nagy, K.
    Czeller, M.
    Tolgyesi, B.
    Nyitrai, Z.
    Savi, A.
    van Mullekom, P.
    Hutton, B. F.
    Fiorini, C.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (03) : 371 - 377