PET imaging and quantification of small animals using a clinical SiPM-based camera

被引:2
|
作者
Desmonts, Cedric [1 ,2 ]
Lasnon, Charline [2 ,3 ]
Jaudet, Cyril [4 ]
Aide, Nicolas [2 ]
机构
[1] Univ Hosp Caen, Nucl Med Dept, Ave Cote Nacre, F-14033 Caen 9, France
[2] Normandy Univ, UNICAEN, INSERM 1086 ANTICIPE, Caen, France
[3] Comprehens Canc Ctr F Baclesse, Nucl Med Dept, UNICANC, Caen, France
[4] Comprehens Canc Ctr F Baclesse, Radiophys Dept, UNICANC, Caen, France
关键词
SiPM PET; Preclinical research; Small-animal imaging; PET quantification; EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY; PERFORMANCE EVALUATION; QUANTITATION; IMPACT;
D O I
10.1186/s40658-023-00583-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Small-animal PET imaging is an important tool in preclinical oncology. This study evaluated the ability of a clinical SiPM-PET camera to image several rats simultaneously and to perform quantification data analysis.Methods: Intrinsic spatial resolution was measured using 18F line sources, and image quality was assessed using a NEMA NU 4-2018 phantom. Quantification was evaluated using a fillable micro-hollow sphere phantom containing 4 spheres of different sizes (ranging from 3.95 to 7.86 mm). Recovery coefficients were computed for the maximum (Amax) and the mean (A50) pixel values measured on a 50% isocontour drawn on each sphere. Measurements were performed first with the phantom placed in the centre of the field of view and then in the off-centre position with the presence of three scattering sources to simulate the acquisition of four animals simultaneously. Quantification accuracy was finally validated using four 3D-printed phantoms mimicking rats with four subcutaneous tumours each. All experiments were performed for both 18F and 68Ga radionuclides.Results: Radial spatial resolutions measured using the PSF reconstruction algorithm were 1.80 mm and 1.78 mm for centred and off-centred acquisitions, respectively. Spill-overs in air and water and uniformity computed with the NEMA phantom centred in the FOV were 0.05, 0.1 and 5.55% for 18F and 0.08, 0.12 and 2.81% for 68Ga, respectively. Recovery coefficients calculated with the 18F-filled micro-hollow sphere phantom for each sphere varied from 0.51 to 1.43 for Amax and from 0.40 to 1.01 for A50. These values decreased from 0.28 to 0.92 for Amax and from 0.22 to 0.66 for A50 for 68 Ga acquisition. The results were not significantly different when imaging phantoms in the off-centre position with 3 scattering sources. Measurements performed with the four 3D-printed phantoms showed a good correlation between theoretical and measured activity in simulated tumours, with r(2) values of 0.99 and 0.97 obtained for 18F and 68Ga, respectively.Conclusion: We found that the clinical SiPM-based PET system was close to that obtained with a dedicated small-animal PET device. This study showed the ability of such a system to image four rats simultaneously and to perform quantification analysis for radionuclides commonly used in oncology.
引用
收藏
页数:19
相关论文
共 50 条
  • [21] SiPM-based neutron Anger camera with auto-calibration capabilities
    Morozov, A.
    Marcos, J.
    Margato, A. L.
    Roulier, D.
    Solovov, V
    JOURNAL OF INSTRUMENTATION, 2019, 14 (03):
  • [22] Development and validation of a Monte Carlo model for a SiPM-based PET scanner
    Merlet, A.
    Presles, B.
    Su, K. H.
    Cochet, A.
    Vrigneaud, J. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S345 - S345
  • [23] Simultaneous SPECT/MR Imaging with a SiPM-Based Preclinical Insert
    Carminati, M.
    Baratelli, F. M.
    Massara, M.
    Occhipinti, M.
    Nagy, K.
    Nyitrai, Z.
    Czeller, M.
    Kuehne, A.
    Niendorf, T.
    Valtorta, S.
    Belloli, S.
    Moresco, R. M.
    Savi, A.
    Iadanza, A.
    Falini, A.
    Politi, L. S.
    Cadioli, M.
    Fiorini, C.
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [24] Prototype of an array SiPM-based scintillator Compton camera for radioactive materials detection
    Zhang, Ji-peng
    Liang, Xiu-zuo
    Cai, Jia-le
    Liu, Yi
    Huang, Xian-chao
    Li, Dao-wu
    Xiao, Xiong
    Pang, Xiao-yu
    Zhang, Zhi-ming
    Wei, Long
    Shuai, Lei
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2019, 3 (03)
  • [25] Ultra Precise Timing with SiPM-Based TOF PET Scintillation Detectors
    Seifert, Stefan
    Vinke, Ruud
    van Dam, Herman T.
    Loehner, Herbert
    Dendooven, Peter
    Beekman, Freek J.
    Schaart, Dennis R.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2329 - +
  • [26] Prototype of an array SiPM-based scintillator Compton camera for radioactive materials detection
    Ji-Peng Zhang
    Xiu-zuo Liang
    Jia-le Cai
    Yi Liu
    Xian-chao Huang
    Dao-wu Li
    Xiong Xiao
    Xiao-yu Pang
    Zhi-ming Zhang
    Long Wei
    Lei Shuai
    Radiation Detection Technology and Methods, 2019, 3
  • [27] Measurement of the time resolution of small SiPM-based scintillation counters
    Kravchenko, E. A.
    Porosev, V. V.
    Savinov, G. A.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [28] A High Bandwidth Preamplifier for SiPM-Based TOF PET Scintillation Detectors
    Seifert, Stefan
    Schaart, Dennis R.
    van Dam, Herman T.
    Huizenga, Jan
    Vinke, Ruud
    Dendooven, Peter
    Lohner, Herbert
    Beekman, Freek J.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 891 - +
  • [29] An innovative SiPM-based camera for gamma-ray astronomy with the small size telescopes of the Cherenkov Telescope Array
    Schioppa, E. J.
    Heller, M.
    Pujadas, I. Troyano
    della Volpe, D.
    Fevre, Y.
    Montaruli, T.
    Zietara, K.
    Kasperek, J.
    Marszalek, A.
    Rajda, P.
    JOURNAL OF INSTRUMENTATION, 2016, 11
  • [30] A SiPM-Based Clinical MRI-Compatible SPECT Insert
    Carminati, M.
    D'Adda, I.
    Morahan, A. J.
    Erlandsson, K.
    Nagy, K.
    Nyitrai, Z.
    Czeller, M.
    Moresco, R.
    Savi, A.
    Van Mullekom, P.
    Hutton, B. F.
    Fiorini, C.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,