Improving image quality using the pause function combination to PROPELLER sequence in brain MRI: a phantom study

被引:0
|
作者
Saotome, Kousaku [1 ,2 ]
Matsumoto, Koji [3 ]
Kato, Yoshiaki [4 ]
Ozaki, Yoshihiro [5 ]
Nagai, Motohiro [4 ]
Hasegawa, Tomoyuki [6 ]
Tsuchiya, Hiroki [7 ]
Yamao, Tensho [1 ]
机构
[1] Fukushima Med Univ, Sch Hlth Sci, Dept Radiol Sci, 10-6 Sakaemachi, Fukushima, Fukushima 9608516, Japan
[2] Univ Tsukuba, Ctr Cybern Res, Tsukuba, Ibaraki 3058577, Japan
[3] Chiba Univ Hosp, Natl Univ Corp, Dept Radiol, Chuo Ku, Chiba, Chiba 2608677, Japan
[4] Kameda Gen Hosp, Med Technol Dept, Diagnost Imaging Room, Kamogawa, Chiba 2968602, Japan
[5] Meiwa Hosp, Dept Radiol, Nishinomiya, Hyogo 6638186, Japan
[6] Hitachi Ltd, Hitachinaka Gen Hosp, Dept Radiol Technol, Hitachinaka, Ibaraki 3120057, Japan
[7] QST Hosp, Natl Inst Quantum Sci & Technol, Dept Med Technol, Radiol Technol Sect, Anagawa, Chiba 2638555, Japan
关键词
PROPELLER; Magnetic resonance imaging; Pause; Radial scan; Phantom; Brain; OVERLAPPING PARALLEL LINES; ENHANCED RECONSTRUCTION PROPELLER; MOTION CORRECTION; BLADE; CONTRAST;
D O I
10.1007/s12194-024-00784-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
While some MRI systems offer a "pause" function, combining it with the PROPELLER method for image quality improvement remains underexplored. This study investigated whether repositioning the head after pausing during PROPELLER imaging enhances image quality. All brain phantom images in this study were obtained using a 3.0 T MRI and acquired using the fast spin-echo T2WI-based PROPELLER with motion correction. By combining the angle of rotational motion of the head phantom and the number of repositioning after a pause, two studies including seven trials were performed. Increasing the rotation angle decreased the image quality; however, pausing the image and repositioning the head phantom to the original angle improved the image quality. A similar result was obtained by repositioning the angle closer to its original angle. Experiments with multiple head movements showed that pausing the scan and repositioning the phantom with each movement improved image quality.
引用
收藏
页码:518 / 526
页数:9
相关论文
共 50 条
  • [1] Improving Image Quality in Brain MRI Using Blade Sequences
    Mavroidis, P.
    Lavdas, E.
    Kostopoulos, S.
    Glotsos, D.
    Roka, V.
    Topalzikis, T.
    Bakas, A.
    Oikonomou, G.
    Stathakis, S.
    Papanikolaou, N.
    Batsikas, G.
    Kaffes, I.
    Kechagias, D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [2] A brain phantom for motion-corrected PROPELLER showing image contrast and construction similar to those of in vivo MRI
    Saotome, Kousaku
    Matsushita, Akira
    Matsumoto, Koji
    Kato, Yoshiaki
    Nakai, Kei
    Murata, Koichi
    Yamamoto, Tetsuya
    Sankai, Yoshiyuki
    Matsumura, Akira
    MAGNETIC RESONANCE IMAGING, 2017, 36 : 32 - 39
  • [3] Image quality using synthetic brain MRI: an age-stratified study
    Zou, Mengsha
    Zhou, Qin
    Li, Ruocheng
    Hu, Manshi
    Qian, Long
    Yang, Zhiyun
    Zhao, Jing
    ACTA RADIOLOGICA, 2023, 64 (05) : 2010 - 2023
  • [4] Compressed sense accelerated MRI simulation: image quality assessment using an ACR MRI phantom
    Wong, O. L.
    Yuan, J.
    Zhou, Y.
    Cheung, K. Y.
    Yu, S. K.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S993 - S994
  • [5] Improving EIT image quality by using multiple meshes combination
    Tang, MX
    Dong, XZ
    You, FS
    Shi, ST
    Qing, MX
    Fu, F
    PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 1055 - 1057
  • [6] Efficacy of PROPELLER in reducing ocular motion artefacts and improving image quality of orbital MRI at 3 T using an eye surface coil
    Jiang, H.
    Wang, S.
    Xian, J.
    Chen, Q.
    Wei, W.
    CLINICAL RADIOLOGY, 2019, 74 (09) : 734.e7 - 734.e12
  • [7] Improving the efficiency of brain MRI image analysis using feature selection
    Konevsky, V. V.
    Blagov, A., V
    Gaidel, A., V
    Kapishnikov, A., V
    Kupriyanov, A., V
    Surovtsev, E. N.
    Asatryan, D. G.
    COMPUTER OPTICS, 2022, 46 (04) : 621 - +
  • [8] Brain image quality according to beam collimation width and image reconstruction algorithm: A phantom study
    Greffier, Joel
    Viry, Anais
    Durand, Quentin
    Hajdu, Steven David
    Frandon, Julien
    Beregi, Jean Paul
    Dabli, Djamel
    Racine, Damien
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 108
  • [9] Image Quality Verification for Off-Center Variation Using ACR Phantom in Open MRI
    Hong, C. P.
    Lee, D. H.
    Han, B.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [10] Image quality testing using an oil-filled ACR MRI phantom at 3 T
    Wey, H. Y.
    Huang, J. C.
    Hsu, Y. Y.
    Lim, K. E.
    Kuan, W. C.
    Chen, C. C.
    Liu, H. L.
    MEDICAL PHYSICS, 2006, 33 (06) : 2009 - 2009