Evaluation of Diasorin Liaison® calprotectin fecal assay adapted for pleural effusion

被引:0
|
作者
Poves, Cristina de Paz [5 ]
Barneo-Caragol, Clara [1 ]
Sanchez, Ana Isabel Cillero [2 ]
Mendiguchia, Lucia Jimenez [3 ]
Caso, Covadonga Quiros [1 ]
Rodriguez, Maria Moreno [3 ]
Gonzalez, Francisco J. Lopez [4 ]
Garcia, Ma Belen Prieto [1 ]
机构
[1] Hosp Univ Cent Asturias, Dept Clin Biochem, Lab Med, Oviedo, Asturias, Spain
[2] Hosp Alvarez Buylla, Dept Clin Biochem, Mieres, Asturias, Spain
[3] Hosp Univ San Agustin, Dept Clin Biochem, Aviles, Asturias, Spain
[4] Hosp Univ Cent Asturias, Dept Pneumol, Oviedo, Asturias, Spain
[5] Hosp Univ Cent Asturias, Dept Clin Biochem, Lab Med, Ave Roma S-N, Oviedo 33011, Asturias, Spain
关键词
automation; calprotectin; chemiluminescence; pleural fluid; validation; MARKER;
D O I
10.1515/almed-2023-0106
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Calprotectin (CP) is a calcium and zinc binding protein that is widely measured on faecal samples but its determination in other biological fluids might be of interest. The aim of this work was to validate the measurement of CP in pleural fluid by chemiluminescence.Methods: LIAISON (R) XL, a fully automated chemiluminescence analyzer, was used for CP quantification on pleural fluid. A validation protocol was designed using both quality control materials provided by the manufacturer and pools of pleural fluid samples. Stability, imprecision, bias, linearity, detection capability and carry over effect were evaluated.Results: CP was stable on pleural fluid at least one week, under refrigerated conditions, and four weeks at -80 80 degrees C. The observed intra- and inter-day imprecision was between 2.2 and 6.49 %, with a negative bias under 5.51 %. The linearity of the method was verified up to 2,000 ng/mL. The LoQ for the assay was 48.52 ng/mL. A statistically significant carry-over effect was observed after measuring CP concentrations above the upper limit of linearity, but given the observed magnitude, a clinically relevant impact should not be expected.Conclusions: Diasorin Liaison (R) calprotectin assay allows reliable measurement of CP in pleural fluid.
引用
收藏
页码:419 / 424
页数:6
相关论文
共 50 条
  • [41] Aldosterone testing: Evaluation of serum and urine applications on the DiaSorin LIAISON automated chemiluminescent analyzer
    Raizman, Joshua
    Yip, Paul
    Kulasingam, Vathany
    CLINICAL BIOCHEMISTRY, 2014, 47 (12) : 1155 - 1155
  • [42] Evaluation of the Diasorin Liaison® dsDNA method for the detection of anti double standard DNA antibodies
    Rouquette, A. -M.
    Desgruelles, C.
    Le Pendeven, C.
    Abuaf, N.
    IMMUNO-ANALYSE & BIOLOGIE SPECIALISEE, 2007, 22 (02): : 120 - 124
  • [43] The Pleural Effusion And Symptom Evaluation (PLEASE) study of breathlessness in patients with a symptomatic pleural effusion
    Muruganandan, Sanjeevan
    Azzopardi, Maree
    Thomas, Rajesh
    Fitzgerald, Deirdre B.
    Kuok, Yi Jin
    Cheah, Hui Min
    Read, Catherine A.
    Budgeon, Charley A.
    Eastwood, Peter R.
    Jenkins, Susan
    Singh, Bhajan
    Murray, Kevin
    Gary, Y. C. Lee
    EUROPEAN RESPIRATORY JOURNAL, 2020, 55 (05)
  • [44] Malignant Pleural Effusion Evaluation and Management
    Gorden, Jed A.
    Fathi, Joelle Thirsk
    CURRENT RESPIRATORY MEDICINE REVIEWS, 2012, 8 (04) : 314 - 321
  • [45] Performance Evaluation of Open Channel Buhlmann Fecal Calprotectin Turbo Assay on Abbott Alinity C Analyzer
    Sataranatarajan, Kavithalakshmi
    Adhikari, Shishir
    Nguyen, Ngoc
    Narasimhan, Madhusudhanan
    Balani, Jyoti
    Muthukumar, Alagarraju
    DIAGNOSTICS, 2024, 14 (16)
  • [46] Evaluation and management of a child with a pleural effusion
    Boyer, DM
    PEDIATRIC EMERGENCY CARE, 2005, 21 (01) : 63 - 68
  • [47] Lung ultrasound in the evaluation of pleural effusion
    Prina, Elena
    Torres, Antoni
    Ribeiro Carvalho, Carlos Roberto
    JORNAL BRASILEIRO DE PNEUMOLOGIA, 2014, 40 (01) : 1 - 5
  • [48] DIAGNOSTIC EVALUATION OF BLOODY PLEURAL EFFUSION
    Mamatha, S.
    Singh, Manish Kumar
    Dutta, Pravati
    Manjhi, Rekha
    Pothal, Sudarsan
    Jha, Amit Kumar
    Nigam, Nilam
    Nigam, Sanjay Kumar
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2013, 2 (02): : 149 - 155
  • [49] The Evaluation of Fatigue in Malignant Pleural Effusion
    Aujayeb, A.
    Wakefield, D.
    JOURNAL OF THORACIC ONCOLOGY, 2021, 16 (03) : S397 - S397
  • [50] EVALUATION OF TALC PLEURAL SYMPHYSIS IN MANAGEMENT OF MALIGNANT PLEURAL EFFUSION
    SHEDBALKAR, AR
    HEAD, JM
    HEAD, LR
    MURPHY, DJ
    MASON, JH
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1971, 61 (03): : 492 - +