Gridless adult cervical spine radiography and its' effect on image quality and radiation dose: A phantom study

被引:0
|
作者
Mekis, N. [1 ,6 ]
Bianchi, T. [2 ]
Doyle, C. [3 ]
Gauchat, M. [2 ]
Geerling, I. [4 ]
Linneman, J. [4 ]
Staats, S. [5 ]
Campeanu, C. [2 ]
机构
[1] Univ Ljubljana, Med Imaging & Radiotherapy Dept, Zdravstvena Pot 5, Ljubljana 100, Slovenia
[2] Univ Appl Sci & Arts Western Switzerland HES SO, Sch Hlth Sci HESAV, Dept Tech Med Radiol, Ave Beaumont 21, CH-1011 Lausanne, Switzerland
[3] Univ Coll Dublin, Sch Med, Radiog & Diagnost Imaging, Dublin 4, Ireland
[4] Hanze Univ Appl Sci, Dept Med Imaging & Radiat Therapy, Petrus Driessenstr 3, NL-9714 CA Groningen, Netherlands
[5] Cent Univ Technol, Dept Clin Sci, Radiog Diagnost, Bloemfontein, South Africa
[6] Zdravstvena Pot 5, Ljubljana 1000, Slovenia
关键词
Antero-posterior C -spine; Lateral C -spine; Anti-scatter grid; Digital radiography; Image quality; Dose reduction;
D O I
10.1016/j.radi.2023.12.009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Imaging of the cervical spine in general radiography is most frequently performed using an anti-scatter grid. The purpose of this study was to investigate the effects of a gridless setting on image quality and radiation dose during digital radiography of the anteroposterior (AP) and lateral (LAT) cervical spine.Methods: A phantom study was performed with a variety of tube voltages (63-75 kV) with and without an anti-scatter grid. The tube current time product (mAs) and dose area product (DAP) were recorded and used to calculate effective dose (ED) and individual organ dose using PCXMC 2.0 software, as well as entrance surface dose (ESD) and objective image quality: signal-to-noise ratio (SNR) and contrast-tonoise ratio (CNR). Subjective visual image quality grading characteristics (VGC) was performed by five qualified radiographers.Results: In a gridless setting, the AP and LAT positions showed significantly lower DAP (1.6 mu Gym2; 61.3 % and 1.6 mu Gym2; 51.2%), ESD (27.6 mu Gy; 57.3% and 77.2 mu Gy; 47.2%) and ED (4.2 mu Sv; 61.3% and 2.3 mu Sv; 48.9%). In a gridless setting in the AP position, there is a slight significant deterioration in image quality. In the lateral projection, on the other hand, the image quality without the use of grid was only significantly reduced in three of six criteria and there was no difference in the objective image quality between the two settings examined.Conclusion: The results of this study show that gridless setting significantly decreases radiation dose and image quality, but the quality in the lateral projection is still acceptable for diagnostic purpose. Implications for practice: The protocol without the use of the anti-scatter grid in cervical spine radiography leads to a reduction in the radiation dose in both projections, but the image quality in the AP is significantly reduced for all criteria examined, with a slight deterioration in image quality in the lateral projection.(c) 2023 The Authors. Published by Elsevier Ltd on behalf of The College of Radiographers. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:359 / 366
页数:8
相关论文
共 50 条
  • [41] Image Quality Assessment of the Digital Radiography Units in Tabriz, Iran: A Phantom Study
    Gharehaghaji, Nahideh
    Khezerloo, Davood
    Abbasiazar, Tohid
    JOURNAL OF MEDICAL SIGNALS & SENSORS, 2019, 9 (02): : 137 - 142
  • [42] Reducing the Breast Cancer Risk and Radiation Dose of Radiography for Scoliosis in Children: A Phantom Study
    Nemoto, Manami
    Chida, Koichi
    DIAGNOSTICS, 2020, 10 (10)
  • [43] Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study
    Kim, Daehong
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    Chung, Myung-Ae
    SYMMETRY-BASEL, 2023, 15 (02):
  • [44] Imaging of Acetabular Fractures: A Phantom Study Comparing Radiation Dose by Radiography and Computed Tomography
    Favinger, Jennifer L.
    Zamora, David A.
    Kanal, Kalpana M.
    Gross, Joel A.
    Gunn, Martin L.
    SEMINARS IN ROENTGENOLOGY, 2019, 54 (02) : 86 - 91
  • [45] Variation in Tube Voltage for Adult Neck MDCT: Effect on Radiation Dose and Image Quality
    Hoang, Jenny K.
    Yoshizumi, Terry T.
    Giao Nguyen
    Toncheva, Greta
    Choudhury, Kingshuk Roy
    Gafton, Andreia R.
    Eastwood, James D.
    Lowry, Carolyn
    Hurwitz, Lynne M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (03) : 621 - 627
  • [46] REDUCING RADIATION-DOSE WHILE MAINTAINING IMAGE QUALITY IN CHEST RADIOGRAPHY
    GILLAN, G
    MILNE, ENC
    ROECK, W
    PEARCE, J
    INVESTIGATIVE RADIOLOGY, 1979, 14 (05) : 379 - 379
  • [47] AMBER AND CONVENTIONAL CHEST RADIOGRAPHY - COMPARISON OF RADIATION-DOSE AND IMAGE QUALITY
    GELEIJNS, J
    BROERSE, JJ
    JULIUS, HW
    VROOMAN, HA
    ZOETELIEF, J
    ZWEERS, D
    KOOL, LJS
    RADIOLOGY, 1992, 185 (03) : 719 - 723
  • [48] Effects of Projections on Radiation Dose for and Image Quality of Chest Digital Radiography for Children
    Xu, Hongrong
    Liu, Bo
    Cai, Jinhua
    Zheng, Huan
    Zheng, Helin
    Yang, Qiurui
    Yao, Changhong
    CURRENT MEDICAL IMAGING, 2021, 17 (12) : 1496 - 1502
  • [49] Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations
    Brindhaban A.
    Al Khalifah K.
    Al Wathiqi G.
    Al Ostath H.
    Australasian Physics & Engineering Sciences in Medicine, 2005, 28 (4): : 216 - 222
  • [50] Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom study
    Geijer, H
    Beckman, KW
    Andersson, T
    Persliden, J
    EUROPEAN RADIOLOGY, 2001, 11 (09) : 1704 - 1709