Improved Resting-State Functional MRI Using Multi-Echo Echo-Planar Imaging on a Compact 3T MRI Scanner with High-Performance Gradients

被引:3
|
作者
Kang, Daehun [1 ]
In, Myung-Ho [1 ]
Jo, Hang Joon [1 ,2 ]
Halverson, Maria A. [1 ]
Meyer, Nolan K. [1 ,3 ]
Ahmed, Zaki [1 ]
Gray, Erin M. [1 ]
Madhavan, Radhika [4 ]
Foo, Thomas K. [4 ]
Fernandez, Brice [5 ]
Black, David F. [1 ]
Welker, Kirk M. [1 ]
Trzasko, Joshua D. [1 ]
Huston III, John [1 ]
Bernstein, Matt A. [1 ]
Shu, Yunhong [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Hanyang Univ, Dept Physiol, Seoul 04763, South Korea
[3] Mayo Clin, Mayo Clin Grad Sch Biomed Sci, Rochester, MN 55905 USA
[4] GE Global Res, Niskayuna, NY 12309 USA
[5] GE Healthcare, F-78530 Buc, France
基金
美国国家卫生研究院;
关键词
fMRI; BOLD; resting state; multi-echo EPI; compact; 3T; gradient system; slew rate; BOLD-CONTRAST SENSITIVITY; TASK-BASED FMRI; BRAIN; NOISE; EPI; COMPENSATION; NONLINEARITY; ENHANCEMENT; SIGNALS;
D O I
10.3390/s23094329
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In blood-oxygen-level-dependent (BOLD)-based resting-state functional (RS-fMRI) studies, usage of multi-echo echo-planar-imaging (ME-EPI) is limited due to unacceptable late echo times when high spatial resolution is used. Equipped with high-performance gradients, the compact 3T MRI system (C3T) enables a three-echo whole-brain ME-EPI protocol with smaller than 2.5 mm isotropic voxel and shorter than 1 s repetition time, as required in landmark fMRI studies. The performance of the ME-EPI was comprehensively evaluated with signal variance reduction and region-of-interest-, seed- and independent-component-analysis-based functional connectivity analyses and compared with a counterpart of single-echo EPI with the shortest TR possible. Through the multi-echo combination, the thermal noise level is reduced. Functional connectivity, as well as signal intensity, are recovered in the medial orbital sulcus and anterior transverse collateral sulcus in ME-EPI. It is demonstrated that ME-EPI provides superior sensitivity and accuracy for detecting functional connectivity and/or brain networks in comparison with single-echo EPI. In conclusion, the high-performance gradient enabled high-spatial-temporal resolution ME-EPI would be the method of choice for RS-fMRI study on the C3T.
引用
收藏
页数:19
相关论文
共 50 条
  • [31] Factors Affecting Detection Power of Blood Oxygen-Level Dependent Signal in Resting-State Functional Magnetic Resonance Imaging Using High-Resolution Echo-Planar Imaging
    Caparelli, Elisabeth C.
    Ross, Thomas J.
    Gu, Hong
    Yang, Yihong
    BRAIN CONNECTIVITY, 2019, 9 (08) : 638 - 648
  • [32] Spatial resolution enhancement using sensitivity-encoded echo-planar imaging at 3T in a typical motor paradigm
    Schmidt, Conny R.
    Pruessmann, Klaas. P.
    Boesiger, Peter
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2007, 31 (08) : 704 - 714
  • [33] Zoomed echo-planar imaging using parallel transmission: impact on image quality of diffusion-weighted imaging of the prostate at 3T
    Andrew B. Rosenkrantz
    Hersh Chandarana
    Josef Pfeuffer
    Michael J. Triolo
    Mohammed Bilal Shaikh
    David J. Mossa
    Christian Geppert
    Abdominal Imaging, 2015, 40 : 120 - 126
  • [34] High Performance Gradient On a Compact 3T MR Scanner for Functional MRI Geometrical Distortion Reduction: Preliminary Experience and Comparison with Conventional Whole-Body 3T Scanner
    Kang, D.
    In, M.
    Jo, H.
    Tan, E. K.
    Yarach, U.
    Meyer, N.
    Trzasko, J.
    Huston, J.
    Foo, T.
    Bernstein, M.
    Shu, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E387 - E387
  • [35] Zoomed echo-planar imaging using parallel transmission: impact on image quality of diffusion-weighted imaging of the prostate at 3T
    Rosenkrantz, Andrew B.
    Chandarana, Hersh
    Pfeuffer, Josef
    Triolo, Michael J.
    Shaikh, Mohammed Bilal
    Mossa, David J.
    Geppert, Christian
    ABDOMINAL IMAGING, 2015, 40 (01): : 120 - 126
  • [36] Robust Myelin Quantitative Imaging from Multi-echo T2 MRI Using Edge Preserving Spatial Priors
    Shen, Xiaobo
    Nguyen, Thanh D.
    Gauthier, Susan A.
    Raj, Ashish
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION (MICCAI 2013), PT I, 2013, 8149 : 622 - 630
  • [37] Measurement of changes in cerebrospinal fluid pulsation after traumatic brain injury using echo-planar imaging-based functional MRI
    Kim, Jun-Hee
    Im, Jae-Geun
    Park, Sung-Hong
    NMR IN BIOMEDICINE, 2024, 37 (03)
  • [38] The Usefulness of Readout-Segmented Echo-Planar Imaging (RESOLVE) for Bio-phantom Imaging Using 3-Tesla Clinical MRI
    Yoshimura, Yuuki
    Kuroda, Masahiro
    Sugianto, Irfan
    Bamgbose, Babatunde O.
    Miyahara, Kanae
    Ohmura, Yuichi
    Kurozumi, Akira
    Matsushita, Toshi
    Ohno, Seiichiro
    Kanazawa, Susumu
    Asaumi, Junichi
    ACTA MEDICA OKAYAMA, 2018, 72 (01) : 53 - 59
  • [39] Fast Distortion-Free T2 Imaging Using DIADEM Combined with Tilted-CAIPI On a Compact 3T Scanner with High-Performance Gradients
    In, M. H.
    Kang, D.
    Yarach, U.
    Shu, Y.
    Trzasko, J. D.
    Huston, J.
    Bernstein, M. A.
    MEDICAL PHYSICS, 2019, 46 (06) : E387 - E387
  • [40] MRI of the knee: value of short echo time fast spin-echo using high performance gradients versus conventional spin-echo imaging for the detection of meniscal tears
    R. M. Kowalchuk
    J. B. Kneeland
    M. K. Dalinka
    E. S. Siegelman
    W. D. Dockery
    Skeletal Radiology, 2000, 29 : 520 - 524