Mass-independent fractionation during Mo isotope measurement by MC-ICP-MS: implications for application of the double spike technique

被引:3
|
作者
Zhao, Yan [1 ]
Yang, Wubin [2 ,3 ]
Leng, Chengbiao [1 ]
机构
[1] East China Univ Technol, State Key Lab Nucl Resources & Environm, Nanchang 330013, Peoples R China
[2] Chinese Acad Sci, Guangzhou Inst Geochem, Key Lab Mineral & Metallogeny, Guangdong Prov Key Lab Mineral Phys & Mat, Guangzhou 510640, Peoples R China
[3] CAS Ctr Excellence Deep Earth Sci, Guangzhou, Peoples R China
基金
中国国家自然科学基金;
关键词
GEOLOGICAL SAMPLES; ERROR PROPAGATION; AMOUNT RATIOS; MOLYBDENUM; EQUATIONS; PRECISE;
D O I
10.1039/d2ja00427e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study presents our new observation on mass-independent fractionation (MIF) during Mo isotope measurement by MC-ICP-MS and investigates its implication for the application of the double spike (DS) technique. A mixture of NIST SRM 3134 Mo solution with a calibrated Mo-97-Mo-100 DS (1 : 1) was prepared for measurements over similar to 15 h. The Mo isotopic data in this study show a good linear relationship in the scatter plots of ln(Mo-97/95) vs. ln(Mo-98/95), ln(Mo-97/95) vs. ln(Mo-100/95), and ln(Mo-98/95) vs. ln(Mo-100/95). Considering measurement errors on both vertical and horizontal axes, York linear regression was applied to linear fit. The slopes derived from the York linear regression are 1.33 +/- 0.12 (2SE), 2.30 +/- 0.20 (2SE), and 1.73 +/- 0.14 (2SE), respectively. Given the error, the slopes are not exactly consistent with their theoretically expected values (1.49, 2.46, and 1.64) based on the assumption of mass-dependent fractionation (MDF). This study, for the first time, identifies MIF during Mo isotope measurement by MC-ICP-MS. DS calculation using theoretical mass fractionation factors shows that the analyzed NIST SRM 3134 Mo solution yielded delta Mo-98/95 values of 0.00 +/- 0.04 parts per thousand (2SD, n = 17) relative to the originally calibrated NIST SRM 3134 Mo. However, using experimental mass fractionation factors calculated from the slopes of line relationships, there is a deviation of about 0.46 parts per thousand for delta Mo-98/95. The results indicate that the calibrated isotope ratios of the standard and DS could not be the true values but approximate values for better matching with the MDF-based equations. After DS calculation using experimental and theoretical mass fractionation factors, respectively, an additional standard-sample bracketing (SSB) method was conducted. Both calculations gave identical and accurate delta Mo-98/95 values (0.00 +/- 0.06 parts per thousand), indicating that the SSB method is able to correct deviation from MIF during DS calculation. Accordingly, we proposed that an additional SSB method is recommended to acquire accurate isotope ratios during the application of the DS technique. This study highlights that researchers should pay attention to the possible existence of instrument-related MIF during isotope measurement.
引用
收藏
页码:785 / 791
页数:7
相关论文
共 50 条
  • [1] Instrumental mass-independent fractionation accounting for the sensitivity of the double spike proportion effect by MC-ICP-MS: a case study of Zn isotope measurements
    Wang, Chunhong
    Zhu, Xiang-kun
    Li, Jin
    Zhu, Zhiyong
    Ma, Jianxiong
    Chen, Yuelong
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (06) : 1529 - 1539
  • [2] Correction of mass fractionation for isotope dilution analysis by MC-ICP-MS
    Zhang, Le
    Ren, Zhong-Yuan
    Ma, Jin-Long
    Xia, Xiao-Ping
    GEOCHEMICAL JOURNAL, 2017, 51 (02) : 157 - 165
  • [3] Vanadium isotope measurement by MC-ICP-MS
    Wu, Fei
    Qi, Yuhan
    Yu, Huimin
    Tian, Shengyu
    Hou, Zhenhui
    Huang, Fang
    CHEMICAL GEOLOGY, 2016, 421 : 17 - 25
  • [4] New methods for determination of the mass-independent and mass-dependent platinum isotope compositions of iron meteorites by MC-ICP-MS
    Poole, Graeme M.
    Stumpf, Roland
    Rehkamper, Mark
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (04) : 783 - 794
  • [5] Isotope fractionation during microbial metal uptake measured by MC-ICP-MS
    Wasylenki, L. E.
    Anbar, A. D.
    Liermann, L. J.
    Mathur, R.
    Gordon, G. W.
    Brantley, S. L.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2007, 22 (08) : 905 - 910
  • [6] Isotope fractionation during microbial metal uptake measured by MC-ICP-MS
    Wasylenki, L.E.
    Anbar, A.D.
    Liermann, L.J.
    Mathur, R.
    Gordon, G.W.
    Brantley, S.L.
    Journal of Analytical Atomic Spectrometry, 2007, 22 (08): : 905 - 910
  • [7] Molybdenum isotope compositions of uranium ore concentrates by double spike MC-ICP-MS
    Rolison, John M.
    Druce, Matthew
    Shollenberger, Quinn R.
    Kayzar-Boggs, Theresa M.
    Lindvall, Rachel E.
    Wimpenny, Josh
    APPLIED GEOCHEMISTRY, 2019, 103 : 97 - 105
  • [8] Precise determination of Pb isotope ratios by simple double spike MC-ICP-MS technique without Tl addition
    Makishima, Akio
    Nath, B. Nagender
    Nakamura, Eizo
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2007, 22 (04) : 407 - 410
  • [9] Plutonium and uranium isotope ratio measurement by MC-ICP-MS
    Taylor, RN
    Warneke, T
    Croudace, IW
    Milton, JA
    Warwick, PE
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2002, 66 (15A) : A766 - A766
  • [10] Measurement of isotope ratios on transient signals by MC-ICP-MS
    Günther-Leopold, I
    Wernli, B
    Kopajtic, Z
    Günther, D
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) : 241 - 249