Development of CMOS dosimetry in proton minibeams for enhanced QA and primary standard absorbed dose calorimetry

被引:2
|
作者
Flynn, Samuel [1 ,2 ]
Allport, Philip [2 ]
De Marzi, Ludovic [3 ]
Green, Stuart [4 ]
Homer, Michael [1 ]
Lee, Nigel [1 ]
Ortiz, Ramon [3 ,5 ]
Patriarca, Annalisa [3 ]
Prezado, Yolanda [3 ,5 ]
Thomas, Russell [1 ,6 ]
Price, Tony [1 ,2 ]
机构
[1] Natl Phys Lab, Med Radiat Sci Grp, Hampton Rd, Teddington TW11 0LW, England
[2] Univ Birmingham, Sch Phys & Astron, Edgbaston Campus, Birmingham B15 2TT, England
[3] Univ PSL, Inst Curie, CNRS UMR3347, Inserm U1021,Signalisat Radiobiol & Canc, F-91898 Orsay, France
[4] Univ Hosp Birmingham NHS Trust, Med Phys, Mindelsohn Way, Birmingham B15 2TH, England
[5] Univ Paris Saclay, CNRS UMR3347, Inserm U1021, Signalisat Radiobiol & Canc, F-91400 Orsay, France
[6] Univ Surrey, Fac Engn & Phys Sci, Stag Hill, Guildford GU2 7XH, England
基金
欧洲研究理事会;
关键词
Dosimetry concepts and apparatus; Instrumentation for hadron therapy; MICROBEAM; RADIOTHERAPY;
D O I
10.1088/1748-0221/18/03/P03014
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Performing accurate and reliable dosimetry in spatially fractionated beams remains a significant challenge due to the steep dose gradients and microscopic scale of features. This results in many conventional detectors and instrumentation being unsuitable for online dosimetry, necessitating frequent offline validation using radiochromic film.In this study, the use of a Complementary Metal-Oxide-Semiconductor (CMOS) detector for evaluation of relative real-time dosimetry of proton minibeam radiation therapy (pMBRT) was investigated. The linearity of the CMOS detector was investigated by varying the proton beam current, with a comparison to a PTW 34001 Roos ionisation chamber used to carry out an independent check. It was found that the relative peaks and valleys of the pMBRT beam could be measured, with results comparable to EBT3XD film. The high sensitivity of the CMOS detector meant it was able to measure dose profiles from peak to valley regions, something not possible with the EBT3XD. The CMOS detector was compared to the treatment delivery log files, with correlation in beam positionseen as the beam is scanned along each slit, but not across; and agreement in beam intensity, with the CMOS detector able to observe beam interruptions.Lastly, the CMOS detector was used in conjunction with the NPL primary-standard proton calorimeter (NPL PSPC) for a preliminary study on combining the NPL PSPC with high resolution temporal information about the incident pMBRT beam. The ultimate aim of this approach is to facilitate detailed thermal modelling to reduce the overall uncertainty in the absolute dose measured from the calorimeter. In these experiments, saturation in the CMOS pixels prevented further thermal modelling of the radiation induced heat flow, however the instantaneous dose rate was observed to be comparable with the predicted NPL PSPC response obtained by masking the CMOS detector.
引用
收藏
页数:24
相关论文
共 50 条
  • [1] DEVELOPMENT OF A PRIMARY STANDARD OF ABSORBED DOSE TO WATER FOR PROTON BEAMS
    Cojocaru, Claudiu
    Renaud, James
    Hoehr, Cornelia
    Belanger-Champagne, Camille
    Trinczek, Michael
    McEwen, Malcolm
    RADIOTHERAPY AND ONCOLOGY, 2023, 186 : S60 - S60
  • [2] Stability of a water calorimetry system as a primary standard for absorbed dose to water
    Cojocaru, Claudiu
    Muir, Bryan
    McEwen, Malcolm
    Ross, Carl
    Klassen, Norman
    Marchington, David
    MEDICAL PHYSICS, 2016, 43 (08) : 4960 - 4961
  • [3] Development of a hybrid alanine-calorimetry absorbed dose standard for linac electron beams
    Surensoy, Rodi
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (08) : 5639 - 5639
  • [4] Development of An Absorbed Dose to Water Primary Standard for HDR Ir-192 Brachytherapy Based On the Fricke Dosimetry System
    El Gamal, I.
    Cojocaru, C.
    Mainegra-Hing, E.
    Ross, C.
    McEwen, M.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [5] Development of a secondary standard for the absorbed dose to water based on the alanine EPR dosimetry system
    Anton, M
    APPLIED RADIATION AND ISOTOPES, 2005, 62 (05) : 779 - 795
  • [6] Development of a primary standard for absorbed dose from unsealed radionuclide solutions
    Billas, I.
    Shipley, D.
    Galer, S.
    Bass, G.
    Sander, T.
    Fenwick, A.
    Smyth, V.
    METROLOGIA, 2016, 53 (06) : 1259 - 1271
  • [7] Development and application of absorbed dose primary standard for 60Co and high-energy photon beams using water calorimetry
    Wang, Zhipeng
    Wang, Kun
    Jin, Sunjun
    Wang, Feifei
    Zhang, Xi
    Liu, Fubin
    Yao, Jintao
    Huang, Ji
    Qi, Yaping
    Zhang, Jian
    METROLOGIA, 2023, 60 (02)
  • [8] Comment on 'Development of a primary standard for absorbed dose from unsealed radionuclide solutions'
    Ross, C. K.
    METROLOGIA, 2017, 54 (04) : 610 - 614
  • [9] Towards An Absorbed Dose-Based Calibration for Ir-192 Brachytherapy Dosimetry-Development of a Primary Standard Water Calorimeter
    Sarfehnia, A.
    Seuntjens, J.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [10] The new NMi orthovolt x-rays absorbed dose to water primary standard based on water calorimetry
    de Prez, L. A.
    de Pooter, J. A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (13): : 3531 - 3542