Carbon Paste Electrode Modified by Dibenzo-18-crown-6 for the Determination of Paracetamol Using Differential Pulse Voltammetry Technique

被引:3
|
作者
Irdhawati, Irdhawati [1 ]
Methaninditya, Ni Ketut Shinta Mas [1 ]
Putra, Anak Agung Bawa [1 ]
机构
[1] Udayana Univ, Fac Math & Nat Sci, Dept Chem, Kampus Bukit Jimbaran, Bali 80361, Indonesia
关键词
carbon paste electrode; dibenzo-18-crown-6; differential pulse voltammetry; paracetamol;
D O I
10.22146/ijc.74393
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The fabrication, optimization and validation of measurement using a modified carbon paste electrode (CPE) with dibenzo-18-crown-6 were carried out for the determination of paracetamol in commercial products. The pH, dibenzo-18-crown-6 concentration in carbon paste, and scan rate parameters were optimized. Validation of the measurement was observed from linear concentration range, LoD, LoQ, precision, and percentage of the recovery. The result showed that the optimum pH was at 2, the optimum concentration of dibenzo-18-crown-6 in carbon paste was 0.6%, optimum scan rates were 35 mV/s for CPE, and increased to 50 mV/s using modified CPE. The linear concentration range for CPE was obtained at 10-300 mu M with LoD and LoQ of 28.88 and 96.28 mu M, respectively. Meanwhile, CPE modified with dibenzo-18-crown-6 gave wider linear concentration range at 1-700 mu M with LoD and LoQ of 52.36 and 174.53 mu M, respectively. The CPE and modified CPE had good precision, with Horwitz ratio values of less than two. The percentage of recovery for two samples with three replicates measurements was obtained (89.81 +/- 1.38)% and (108.02 +/- 0.42)% for samples A and B, respectively. Dibenzo-18-crown-6 modified CPE was used for the determination of paracetamol in both samples yielding 97-98% compared with the paracetamol composition on its labels.
引用
收藏
页码:53 / 61
页数:9
相关论文
共 50 条
  • [11] Nanoalumina Modified Carbon Paste Electrode for Sensitive and Rapid Determination of Hypochlorite Ion by Differential Pulse Voltammetry
    Malmir, Maryam
    Afkhami, Abbas
    Madrakian, Tayyebeh
    Soltani-Felehgari, Farzaneh
    Madrakian, Elaheh
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY RESEARCH, 2020, 7 (01): : 99 - 109
  • [12] DETERMINATION OF TETRAMETHRIN (NEO-PYNAMIN) BY DIFFERENTIAL PULSE VOLTAMMETRY WITH A CARBON PASTE ELECTRODE MODIFIED WITH SEPIOLITE
    HERNANDEZ, P
    VICENTE, J
    HERNANDEZ, L
    FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 334 (06): : 550 - 553
  • [13] Determination of Trace Tl(I) by Differential Pulse Anodic Stripping Voltammetry Using a Novel Modified Carbon Paste Electrode
    Ghanei-Motlagh, Masoud
    Baghayeri, Mehdi
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2020, 167 (06)
  • [14] Differential pulse voltammetry determination of L-cysteine with ferrocene-modified carbon paste electrode
    Raoof, JB
    Ojani, R
    Kolbadinezhad, M
    BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 2005, 78 (05) : 818 - 826
  • [15] Direct determination of boscalid in grape samples by differential pulse voltammetry using a carbon paste electrode
    Liberato, Priscila Azevedo
    Okumura, Leonardo Luiz
    de Souza Silva, Astrea Filomena
    Aleixo, Herbert
    Silva, Junio Goncalves
    Diniz, Juliana Aparecida
    Oliveira, Andre Fernando
    ANALYTICAL METHODS, 2021, 13 (43) : 5195 - 5203
  • [16] Determination of buprenorphine by differential pulse voltammetry on carbon paste electrode using SDS as an enhancement factor
    Behpour, Mohsen
    Valipour, Akram
    Keshavarz, Mahin
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2014, 42 : 500 - 505
  • [17] Differential pulse voltammetry for determination of benorilate in pharmaceutical formulations at carbon paste electrode
    Wang, CY
    Hu, XY
    Chen, Q
    Rong, G
    ANALYTICAL LETTERS, 2005, 38 (05) : 893 - 905
  • [18] ELECTRODE CHARACTERISTICS OF MEMBRANES BASED ON DIBENZO-18-CROWN-6 DERIVATIVES
    NOROV, SK
    GULAMOVA, MT
    ZHUKOV, AF
    NOROVA, GM
    MAMEDOVA, YG
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1988, 43 (06): : 777 - 783
  • [19] Determination of methylparaben by differential pulse voltammetry using a glassy carbon electrode modified with polypyrrole
    Seyfullah Madakbaş
    Senem Kamiloǧlu
    Ece Kök Yetimoǧlu
    Journal of Analytical Chemistry, 2015, 70 : 725 - 730
  • [20] Determination of methylparaben by differential pulse voltammetry using a glassy carbon electrode modified with polypyrrole
    Madakbas, Seyfullah
    Kamiloglu, Senem
    Yetimoglu, Ece Kok
    JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 70 (06) : 725 - 730