Robotic pullback technique of a precurved cochlear-implant electrode array using real-time impedance sensing feedback

被引:1
|
作者
Riojas, Katherine E. [1 ]
Bruns, Trevor L. [1 ]
Granna, Josephine [1 ]
Webster, Robert J., III [1 ]
Labadie, Robert F. [2 ]
机构
[1] Vanderbilt Univ, Dept Mech Engn, Nashville, TN 37235 USA
[2] Med Univ South Carolina, Dept Otolaryngol, Charleston, SC 29425 USA
基金
美国国家科学基金会;
关键词
Cochlear implants; Impedance sensing; Robotic surgery; Otology; INSERTION; PROXIMITY;
D O I
10.1007/s11548-022-02772-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose During traditional insertion of cochlear implant (CI) electrode arrays (EAs), surgeons rely on limited tactile feedback and visualization of the EA entering the cochlea to control the insertion. One insertion approach for precurved EAs involves slightly overinserting the EA and then retracting it slightly to achieve closer hugging of the modiolus. In this work, we investigate whether electrical impedance sensing could be a valuable real-time feedback tool to advise this pullback technique. Methods Using a to-scale 3D-printed scala tympani model, a robotic insertion tool, and a custom impedance sensing system, we performed experiments to assess the bipolar insertion impedance profiles for a cochlear CI532/632 precurved EA. Four pairs of contacts from the 22 electrode contacts were chosen based on preliminary testing and monitored in real time to halt the robotic insertion once the closest modiolar position had been achieved but prior to when the angular insertion depth (AID) would be reduced. Results In this setting, the open-loop robotic insertion impedance profiles were very consistent between trials. The exit of each contact from the external stylet of this EA was clearly discernible on the impedance profile. In closed-loop experiments using the pullback technique, the average distance from the electrode contacts to the modiolus was reduced without greatly affecting the AID by using impedance feedback in real time to determine when to stop EA retraction. Conclusion Impedance sensing, and specifically the access resistance component of impedance, could be a valuable real-time feedback tool in the operating room during CI EA insertion. Future work should more thoroughly analyze the effects of more realistic operating room conditions and inter-patient variability on this technique.
引用
收藏
页码:413 / 421
页数:9
相关论文
共 50 条
  • [41] Magnetic nanobead chain-assisted real-time impedance monitoring using PCB interdigitated electrode for Salmonella detection
    Jiang, Fan
    Wang, Lei
    Jin, Nana
    Yuan, Jing
    Li, Yanbin
    Lin, Jianhan
    ISCIENCE, 2023, 26 (11)
  • [42] Monitoring Cochlear Nerve Integrity During Vestibular Schwannoma Microsurgery in Real-Time Using Cochlear Implant Evoked Auditory Brainstem Response and Streaming Neural Response Imaging
    Patel, Neil S.
    Saoji, Aniket A.
    Olund, Amy P.
    Carlson, Matthew L.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (02) : E201 - E207
  • [43] Real-Time Feature Extraction From Electrocochleography With Impedance Measurements During Cochlear Implantation Using Linear State-Space Models
    Andonie, Raphael R.
    Wimmer, Wilhelm
    Wildhaber, Reto A.
    Caversaccio, Marco
    Weder, Stefan
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2023, 70 (11) : 3137 - 3146
  • [44] Mirror Langmuir probe: A technique for real-time measurement of magnetized plasma conditions using a single Langmuir electrode
    LaBombard, B.
    Lyons, L.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (07):
  • [45] Real-time in situ sensing of multiple water quality related parameters using micro-electrode array (MEA) fabricated by inkjet-printing technology (IPT)
    Xu, Zhiheng
    Dong, Qiuchen
    Otieno, Brunah
    Liu, Yucheng
    Williams, Isaiah
    Cai, Dingyi
    Li, Yan
    Lei, Yu
    Li, Baikun
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 237 : 1108 - 1119
  • [46] Label-free electrochemical impedance spectroscopy using a micro interdigitated electrode inside a PCR chip for real-time monitoring
    Jeong, Sangdo
    Lim, Juhun
    Kim, Jinhwan
    Kim, Mi-Young
    Lee, Jong-Hyun
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2019, 25 (09): : 3503 - 3510
  • [47] Label-free electrochemical impedance spectroscopy using a micro interdigitated electrode inside a PCR chip for real-time monitoring
    Sangdo Jeong
    Juhun Lim
    Jinhwan Kim
    Mi-Young Kim
    Jong-Hyun Lee
    Microsystem Technologies, 2019, 25 : 3503 - 3510
  • [48] A 3D porous electrode for real-time monitoring of microalgal growth and exopolysaccharides yields using Electrochemical Impedance Spectroscopy
    Cotta, Francisco C.
    Amaral, Raquel
    Bacellar, Felipe L.
    Correia, Diogo
    Asadi, Kamal
    Rocha, Paulo R. F.
    BIOSENSORS & BIOELECTRONICS, 2025, 277
  • [49] Design, development, and testing of real-time feedback controllers for semiconductor etching processes using in situ spectroscopic ellipsometry sensing
    Rosen, IG
    Parent, T
    Fidan, B
    Wang, CM
    Madhukar, A
    IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 2002, 10 (01) : 64 - 75
  • [50] Real-Time Monitoring In Vitro Cellular Cytotoxicity of Silica Nanotubes Using Electric Cell-Substrate Impedance Sensing (ECIS)
    Trong Binh Tran
    Phuong Diem Nguyen
    Um, Soong Ho
    Son, Sang Jun
    Min, Junhong
    JOURNAL OF BIOMEDICAL NANOTECHNOLOGY, 2013, 9 (02) : 286 - 290