Development and application of absorbed dose primary standard for 60Co and high-energy photon beams using water calorimetry

被引:1
|
作者
Wang, Zhipeng [1 ]
Wang, Kun [1 ]
Jin, Sunjun [1 ]
Wang, Feifei [2 ]
Zhang, Xi [1 ]
Liu, Fubin [1 ]
Yao, Jintao [1 ,3 ]
Huang, Ji [1 ]
Qi, Yaping [1 ]
Zhang, Jian [1 ]
机构
[1] Natl Inst Metrol NIM, Div Ionizing Radiat Metrol, Beijing, Peoples R China
[2] China Inst Atom Energy CIAE, Beijing, Peoples R China
[3] Anhui Univ Sci & Technol, Huainan, Peoples R China
关键词
primary standard; absorbed-dose-to-water; water calorimeter; ionization chamber; beam quality conversion factor; IONIZATION CHAMBERS; K(Q) FACTORS; EFFECTIVE POINT; MV PHOTON; DOSIMETRY; HEAT;
D O I
10.1088/1681-7575/acc01b
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Several new glass vessels, thermistor probes, and water phantoms have been designed and built at the National Institute of Metrology (China) to upgrade and develop the existing water calorimeter. The increased plane-parallel vessels have shorter thermal stability times and shallower positioning depths (similar to 1.3 cm) than the previous cylindrical vessels, which makes them suitable for electron beams. The sensitivity of the new probes is 10% greater than the previous ones. In this study, detailed experimental and theoretical investigations of various factors affecting the new water calorimeter are performed. The system uncertainty of the water calorimeter is reduced and the robustness of the determination of the absorbed-dose-to-water D (w) is improved using a variety of geometric detector vessels and two kinds of high-purity water systems saturated with high-purity gases. This new calorimeter is employed as a primary standard for determining the D (w), has achieved a combined standard uncertainty of 0.24% for a Co-60 beam, 0.27% for 6 MV and 10 MV photon beams and 0.30% for a 25 MV photon beam. The beam quality conversion factors k(Q) of ten cylindrical and three plane-parallel ionization chambers are measured using the new calorimeter to improve the reference dosimetry accuracy of high-energy clinical photon beams.
引用
收藏
页数:17
相关论文
共 50 条
  • [41] On absorbed dose in narrow 60Co gamma-ray beams and dosimetry of the Gamma Knife
    Nizin, PS
    MEDICAL PHYSICS, 1998, 25 (12) : 2347 - 2351
  • [42] Entrance Skin Dose by DAP Meter Calibrated Based on 60Co Absorbed Dose to Water for Diagnostic X-Ray Beams
    Umeno, S.
    Araki, F.
    MEDICAL PHYSICS, 2018, 45 (06) : E596 - E597
  • [43] Re-evaluation of correction factors of a primary standard graphite calorimeter in 60Co gamma ray beams as a basis for the appointment of the BEV absorbed dose rate to water reference value
    Baumgartner, Andreas
    Steurer, Andreas
    Tiefenboeck, Wilhelm
    Gabris, Frantisek
    Maringer, Franz Josef
    Kapsch, Ralf-Peter
    Stucki, Gerhard
    RADIATION PROTECTION DOSIMETRY, 2011, 145 (01) : 3 - 12
  • [44] High Accuracy Measurement of Absorbed Dose to Water Using Optically Stimulated Luminescence Dosimeter in High-Energy Photon Beam Field
    Goto, S.
    Hayashi, H.
    Yamaguchi, H.
    Shimizu, M.
    Akino, R.
    Sekiguchi, H.
    MEDICAL PHYSICS, 2020, 47 (06) : E789 - E789
  • [45] ABSORBED-DOSE CALIBRATIONS IN HIGH-ENERGY PHOTON BEAMS AT THE NATIONAL-PHYSICAL-LABORATORY - CONVERSION PROCEDURE
    BURNS, JE
    PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (10): : 1555 - 1575
  • [46] Comparison of graphite-to-water absorbed-dose transfers for Co-60 photon beams using ionometry and Fricke dosimetry
    Chauvenet, B
    Baltes, D
    Delaunay, F
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11): : 2053 - 2063
  • [47] Comparison of the standards for absorbed dose to water of the VNIIFTRI, Russia and the BIPM in 60Co γ rays
    Allisy-Roberts, P. J.
    Kessler, C.
    Burns, D. T.
    Berlyand, V.
    Berlyand, A.
    METROLOGIA, 2010, 47
  • [48] Comparison of the standards for absorbed dose to water of the VSL, The Netherlands, and the BIPM for 60Co γ rays
    Kessler, C.
    Burns, D.
    Jansen, B. J.
    de Pooter, J. A.
    de Prez, L. A.
    METROLOGIA, 2018, 55
  • [49] Comparison of the standards for absorbed dose to water of the NRC and the BIPM for 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Allisy-Roberts, P. J.
    McCaffrey, J. P.
    McEwen, M. R.
    Ross, C. K.
    METROLOGIA, 2010, 47
  • [50] Comparison of the standards for absorbed dose to water of the ARPANSA and the BIPM for 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Allisy-Roberts, P. J.
    Butler, D.
    Lye, J.
    Webb, D.
    METROLOGIA, 2012, 49