Development and application of absorbed dose primary standard for 60Co and high-energy photon beams using water calorimetry

被引:1
|
作者
Wang, Zhipeng [1 ]
Wang, Kun [1 ]
Jin, Sunjun [1 ]
Wang, Feifei [2 ]
Zhang, Xi [1 ]
Liu, Fubin [1 ]
Yao, Jintao [1 ,3 ]
Huang, Ji [1 ]
Qi, Yaping [1 ]
Zhang, Jian [1 ]
机构
[1] Natl Inst Metrol NIM, Div Ionizing Radiat Metrol, Beijing, Peoples R China
[2] China Inst Atom Energy CIAE, Beijing, Peoples R China
[3] Anhui Univ Sci & Technol, Huainan, Peoples R China
关键词
primary standard; absorbed-dose-to-water; water calorimeter; ionization chamber; beam quality conversion factor; IONIZATION CHAMBERS; K(Q) FACTORS; EFFECTIVE POINT; MV PHOTON; DOSIMETRY; HEAT;
D O I
10.1088/1681-7575/acc01b
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Several new glass vessels, thermistor probes, and water phantoms have been designed and built at the National Institute of Metrology (China) to upgrade and develop the existing water calorimeter. The increased plane-parallel vessels have shorter thermal stability times and shallower positioning depths (similar to 1.3 cm) than the previous cylindrical vessels, which makes them suitable for electron beams. The sensitivity of the new probes is 10% greater than the previous ones. In this study, detailed experimental and theoretical investigations of various factors affecting the new water calorimeter are performed. The system uncertainty of the water calorimeter is reduced and the robustness of the determination of the absorbed-dose-to-water D (w) is improved using a variety of geometric detector vessels and two kinds of high-purity water systems saturated with high-purity gases. This new calorimeter is employed as a primary standard for determining the D (w), has achieved a combined standard uncertainty of 0.24% for a Co-60 beam, 0.27% for 6 MV and 10 MV photon beams and 0.30% for a 25 MV photon beam. The beam quality conversion factors k(Q) of ten cylindrical and three plane-parallel ionization chambers are measured using the new calorimeter to improve the reference dosimetry accuracy of high-energy clinical photon beams.
引用
收藏
页数:17
相关论文
共 50 条
  • [31] Comparison of Absorbed-Dose to Water in High-Energy Photon Beams Due to Differences in Beam Quality Conversion Factor
    Kinoshita, N.
    Takemura, A.
    Nishimoto, Y.
    Kita, A.
    Toi, A.
    Murai, E.
    Adachi, T.
    Shioura, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [32] ABSORBED DOSE AND CAVITY IONIZATION FOR HIGH-ENERGY ELECTRON BEAMS
    KESSARIS, ND
    RADIATION RESEARCH, 1970, 43 (02) : 288 - +
  • [33] Development of a hybrid alanine-calorimetry absorbed dose standard for linac electron beams
    Surensoy, Rodi
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (08) : 5639 - 5639
  • [34] The PTB water calorimeter for the absolute determination of absorbed dose to water in 60Co radiation
    Krauss, A
    METROLOGIA, 2006, 43 (03) : 259 - 272
  • [35] Absorbed dose to water based dosimetry versus air kerma based dosimetry for high-energy photon beams: an experimental study
    Palmans, H
    Nafaa, L
    De Jans, J
    Gillis, S
    Hoornaert, MT
    Martens, C
    Piessens, M
    Thierens, H
    Van der Plaetsen, A
    Vynckier, S
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (03): : 421 - 440
  • [36] CALORIMETRIC DETERMINATION OF THE ABSORBED DOSE-TO-WATER BEAM QUALITY CORRECTION FACTOR K(Q) FOR HIGH-ENERGY PHOTON BEAMS
    VATNITSKY, SM
    SIEBERS, JV
    MILLER, DW
    MEDICAL PHYSICS, 1995, 22 (11) : 1749 - 1752
  • [37] DOSE CALCULATIONS FOR HIGH-ENERGY ELECTRON AND PHOTON BEAMS
    CUNNINGHAM, JR
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1989, 13 (03) : 241 - 250
  • [38] Comparison of Absorbed Dose to Water of High-Energy Photon Beams Based On the IAEA IRS-277, TRS-398 and Direct Calibration in Megavoltage Photon Beams
    Wang, K.
    Jin, S.
    Fan, Y.
    Chang, X.
    MEDICAL PHYSICS, 2017, 44 (06) : 2932 - 2932
  • [39] HIGH-ENERGY PHOTON PRIMARY AND SCATTER DOSE CALCULATION IN WATER USING PRIMARY COLLISION KERMA AS A PARAMETER
    IWASAKI, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (08): : 1013 - 1019
  • [40] Changes in absorbed dose to water for high-energy electron beams by beam quality correction factor
    Kinoshita, N.
    Takemura, A.
    Toi, A.
    Kita, A.
    Nishimoto, Y.
    Adachi, T.
    Shioura, H.
    Kimura, H.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S56 - S56