Monte Carlo Simulation of Liver Dosimetry with Yttrium-90 Radionuclide Using Gate: 3D Phantom

被引:1
|
作者
Karadeniz-Yildirim, Ayse [1 ]
Ozkorucuklu, Suat [2 ]
Tanyildizi-Kokkulunk, Handan [3 ]
Demir, Mustafa [4 ]
Yasar, Dogan [5 ]
机构
[1] Istanbul Aydin Univ, Opticianry Program, Istanbul, Turkiye
[2] Istanbul Univ, Phys Dept, Istanbul, Turkiye
[3] Altinbas Univ, Radiotherapy Program, Istanbul, Turkiye
[4] Istanbul Cerrahpasa Univ, Nucl Med Dept, Istanbul, Turkiye
[5] Ahi Evren Univ, Phys Dept, Merkez, Kirsehir, Turkiye
关键词
monte carlo method; gate; dosimetry; yttrium; technetium; phantom; MICROSPHERE BRACHYTHERAPY;
D O I
10.3103/S1068335623600201
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
This study is aimed to validate Y-90 dosimetry with MC simulation by comparing the TLD measurements results obtained in 3D liqid phantom (LP) and its monte carlo (MC) simulation. In the first step, tumor imitation was filled with 1900 mu Ci (70.3 MBq) amount of Tc-99m activity in LP. In addition, 24 thermoluminescence dosimetry chips (TLD-100) were placed in 8 different locations in the LP. The same measurement system was created in GEANT4 Tomography Emission Application (GATE version 8.1). DoseActor library was defined for 24 points instead of TLDs and doses were taken. In the second step, Y-90 with the amount of 2 mCi (74 MBq) activity and 27 DoseActor libraries were placed in tumor mimic and different locations in LP to calculate the dose distribution with GATE. GATE and TLD results for Tc-99m were compared for validation. The strong correlation coefficient was calculated between the instant absorbed dose rates obtained via TLD measurements and GATE calculations as R-2 = 0.994. At the Y-90 dosimetry, the highest, lowest and mean cumulative absorbed doses calculated by GATE for a half-life were found to be 1.166 x 10(2) +/- 3.907 x 10(-1), 5.440 x 10(1) +/- 2.597 x 10(-1), and 75.145 +/- 0.303 Gy in the tumor imitation, and 1.167 x 10(1) +/- 1.777 x 10(-1), 9.929 x 10(-5) +/- 6.905 x 10(-5), 1.169 +/- 0.019 Gy in the liver parenchymal tissue imitation, respectively. GATE program is considered to be a promising application in internal radionuclide dosimetry, based on linear correlation between TLD and GATE measurements in our study.
引用
收藏
页码:30 / 37
页数:8
相关论文
共 50 条
  • [11] Dosimetry of Rhenium-188, Praseodymium-142, and Yttrium-90 Using Monte Carlo Simulation for Application in High-Dose-Rate Eye Plaque Brachytherapy
    Ferreira, M.
    Jung, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S749 - S750
  • [12] Geant4/GATE Monte Carlo Code for Internal Dosimetry Using Voxelized Phantom
    Kaddouch, S.
    El Khayati, N.
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2017, 72 (06) : 658 - 662
  • [13] Geant4/GATE Monte Carlo Code for Internal Dosimetry Using Voxelized Phantom
    S. Kaddouch
    N. El Khayati
    Moscow University Physics Bulletin, 2017, 72 : 658 - 662
  • [14] Modeling of a Single Blood Vessel Embolized with Yttrium-90 Or Praseodymium-142 Glass Microspheres Using Monte Carlo Simulation
    Ferreira, M.
    Podder, T.
    Jung, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [15] Image-based 3D Dosimetry Techniques For Yttrium-90 SIRT Of HCC: Quantitative Comparison Of Tumor And Healthy Liver Absorbed Doses
    Brosch, J.
    Gosewisch, A.
    von Zimmermann, H.
    Kaiser, L.
    Bartenstein, P.
    Ilhan, H.
    Todica, A.
    Boening, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S481 - S481
  • [16] 3D Monte Carlo Simulation of Tri-Gate MOSFETs Using Tetrahedral Finite Elements
    Aldegunde, Manuel
    Garcia-Loureiro, Antonio J.
    Martinez, Antonio
    Kalna, Karol
    SISPAD: 2008 INTERNATIONAL CONFERENCE ON SIMULATION OF SEMICONDUCTOR PROCESSES AND DEVICES, 2008, : 153 - +
  • [17] 3D Monte Carlo Simulation of Gate-All-Around Germanium nMOSFET
    Zhu, Shufang
    Wei, Kangliang
    Du, Gang
    Liu, Xiaoyan
    2011 INTERNATIONAL CONFERENCE OF ELECTRON DEVICES AND SOLID-STATE CIRCUITS (EDSSC), 2011,
  • [18] Comparison of Y-90 and Ho-166 Dosimetry Using Liver Phantom: A Monte Carlo Study7
    Yildirim, Ayse Karadeniz
    Kokkulunk, Handan Tanyildizi
    ANTI-CANCER AGENTS IN MEDICINAL CHEMISTRY, 2022, 22 (07) : 1348 - 1353
  • [19] A preliminary study of Monte Carlo simulation for the 3D printing small animal phantom design
    Song, H.
    Jung, H.
    Kim, K.
    Ji, Y.
    Park, S.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S738 - S738
  • [20] RADIOTHERAPY MONTE CARLO DOSIMETRY USING GATE/GEANT4: VALIDATION STUDY IN ANTHROPOMORPHIC PHANTOM
    Perrot, Y.
    Breton, V.
    Donnarieix, D.
    Maigne, L.