Incomplete Screw Thread Engagement of Proximal Fragment: A Possible Failure Risk After Internal Fixation for Femoral Neck Fractures

被引:1
|
作者
Tanaka, Atsuki [1 ,2 ]
Hiranaka, Takafumi [1 ,2 ]
Fujishiro, Takaaki [1 ,2 ]
Koide, Motoki [1 ,2 ]
Suda, Yoshihito [1 ,2 ]
Saito, Akira [1 ,2 ]
Arimoto, Akihiko [1 ,2 ]
Okamoto, Koji [1 ,2 ]
机构
[1] Takatsuki Gen Hosp, Dept Orthopaed Surg, Takatsuki, Japan
[2] Takatsuki Gen Hosp, Joint Surg Ctr, Takatsuki, Japan
关键词
incomplete thread engagement; fixation failure; sliding mechanism; thread length; internal fixation; femoral neck fracture; OUTCOMES;
D O I
10.7759/cureus.41349
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background For successful internal fixation for femoral neck fracture, the sliding mechanism of the screw is important because it can induce inter-fragmental compression. The thread should penetrate the fracture line and be located within the proximal fragment. If screw thread engagement is incomplete and a part of the thread remains within the distal fragment, the screw sliding can be disturbed, potentially leading to fixation failure. We hypothesized that screw thread in the fracture is a risk of fixation failure. Methods We studied 133 hips that underwent internal fixation for femoral neck fracture using dual sliding and compression screws (DSCS) with 20 mm threads. The existence of incomplete thread engagement and fixation failure (cut out, perforation, pseudoarthrosis, or femoral neck shortening) were evaluated on anteroposterior hip radiography postoperatively. The distances from the thread end to the fracture line, screw head to the femoral head cortex, and femoral head diameter were measured to analyze their relationships with any incomplete thread engagement and fixation failure. Differences in evaluation data were assessed using Fisher's exact test, Student's t -test, and receiver operating characteristic (ROC) analysis. Results Forty-six cases had at least one screw with incomplete thread engagement, and the other 87 hips had a complete engagement. The failure rate in the group of hips with incomplete thread engagement was significantly higher (7/46, 15.2%) than that in the group of hips with complete thread engagement (3/87, 3.4%) (P = 0.032). Incomplete thread engagement was found in 59 out of 266 screws (22.2%), and a femoral head <= 43.9 mm in diameter was associated with an increased risk of incomplete thread engagement. Most incomplete thread engagement screws (81.4%) had < 5 mm thread length within the distal fragment. Conclusion A partially threaded screw is a significant risk of fixation failure after internal fixation for a femoral neck fracture. The smaller femoral head diameter increases the possibility of incomplete thread engagement. Shortening the thread length by 5 mm may help to avoid incomplete thread engagement.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Subtrochanteric fracture : A rare but severe complication after screw fixation of femoral neck fractures in the elderly
    Jansen, Hendrik
    Frey, Soenke P.
    Meffert, Rainer H.
    ACTA ORTHOPAEDICA BELGICA, 2010, 76 (06): : 778 - 784
  • [42] High incidence of “in–out–in” posterosuperior screws after cannulated screw fixation of femoral neck fractures
    Brandon J. Yuan
    Mohamad Tayseer Shamaa
    William R. Aibinder
    Joshua A. Parry
    William W. Cross
    Jonathan D. Barlow
    Stephen A. Sems
    European Journal of Orthopaedic Surgery & Traumatology, 2020, 30 : 1417 - 1420
  • [43] Subtrochanteric fracture after cannulated screw fixation of femoral neck fractures: A report of four cases
    Kloen, P
    Rubel, IF
    Lyden, JP
    Helfet, DL
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2003, 17 (03) : 225 - 229
  • [44] Patients With Femoral Neck Fractures Are at Risk for Conversion to Arthroplasty After Internal Fixation: A Machine-learning Algorithm
    van de Kuit, Anouk
    Oosterhoff, Jacobien H. F.
    Dijkstra, Hidde
    Sprague, Sheila
    Bzovsky, Sofia
    Bhandari, Mohit
    Swiontkowski, Marc
    Schemitsch, Emil H. H.
    IJpma, Frank F. A.
    Poolman, Rudolf W. W.
    Doornberg, Job N. N.
    Hendrickx, Laurent A. M.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2022, 480 (12) : 2350 - 2360
  • [45] Femoral neck fractures after internal fixation of trochanteric fractures with implants in situ in adults: A systematic review
    Barquet, Antonio
    Giannoudis, Peter V.
    Gelink, Andres
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2018, 49 (12): : 2121 - 2131
  • [46] Computer-assisted cannulated screw internal fixation versus conventional cannulated screw internal fixation for femoral neck fractures: a systematic review and meta-analysis
    Cheng, Qing-Hao
    Li, Peng-Biao
    Lu, Ting-Ting
    Guo, Shi-Fang
    Di, Wen-Fei
    Yang, Ke-Hu
    Qian, Yao-Wen
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
  • [47] Computer-assisted cannulated screw internal fixation versus conventional cannulated screw internal fixation for femoral neck fractures: a systematic review and meta-analysis
    Qing-hao Cheng
    Peng-biao Li
    Ting-ting Lu
    Shi-fang Guo
    Wen-fei Di
    Ke-hu Yang
    Yao-wen Qian
    Journal of Orthopaedic Surgery and Research, 16
  • [48] Evaluation of the Posterior Tilt Angle in Predicting Failure of Nondisplaced Femoral Neck Fractures After Internal Fixation: A Systematic Review
    Papadelis, Efstratios
    Chaudhry, Yash P.
    Hayes, Hunter
    Talone, Christopher
    Shah, Mitesh P.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2023, 37 (02) : E89 - E94
  • [49] Salvage procedures for trochanteric femoral fractures after internal fixation failure: Biomechanical comparison of a plate fixator and the dynamic condylar screw
    Konstantinidis, L.
    Papaioannou, C.
    Mehlhorn, A.
    Hirschmueller, A.
    Suedkamp, N. P.
    Helwig, P.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H7) : 710 - 717
  • [50] Displaced femoral neck fractures at the bone-screw interface after in situ fixation of slipped capital femoral epiphysis
    Canale, ST
    Casillas, M
    Banta, JV
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 1997, 17 (02) : 212 - 215