Patient skin dose measurement and risk of deterministic effect during fluoroscopy cardiac procedures

被引:1
|
作者
Nana, Ruth Njantang [1 ,2 ]
Inkoom, Stephen [3 ,4 ]
Moyo, Maurice Ndontchueng [1 ,2 ]
Mvoufo, Fokou [1 ,2 ]
Hasford, Francis [3 ,5 ]
Owusu-Manteaw, Philip [4 ]
Motapon, Ousmanou [1 ,6 ]
机构
[1] Univ Douala, Postgrad Sch Fundamental & Appl Sci, Fundamental Phys Lab, POB 24157, Douala, Cameroon
[2] Natl Radiat Protect Agcy NRPA, POB 5331, Akwa, Douala, Cameroon
[3] Univ Ghana, Sch Nucl & Allied Sci, POB AE 1, Accra, Ghana
[4] Ghana Atom Energy Commiss, Radat Protect Inst, POB LG80, Legon, Accra, Ghana
[5] Ghana Atom Energy Commiss, Radiol & Med Sci Res Inst, POB LG80, Legon, Accra, Ghana
[6] Univ Maroua, Fac Sci, POB 814, Maroua, Cameroon
关键词
BACKSCATTER FACTORS;
D O I
10.1093/rpd/ncad206
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study aimed at assessing patient's peak skin doses (PSD) during fluoroscopy cardiac procedures and proposed a look up table to enhance patient's dose management. Perspex phantom and thermoluminescent dosemeters (TLD) were irradiated for different dose levels with X-ray equipment (Philips Azurion 7). It was found that PSD measures were higher than the kerma at the interventional reference point [K (IRP)] reported with factors 1.55, 1.75 and 2.88 for anterior posterior (AP0(o)), left anterior oblique (LAO45(o)) and right anterior oblique (RAO45(o)), respectively. The equations describing the correlation between the PSD measured kerma area product and cumulative air kerma were found with R-square values of 0.98 and 0.99, respectively. The statistical analysis shows a strong linear correlation between PSD and K (IRP) (P-value = 0.05). It was also found that 27% of the patients population considered in this work, received a skin dose higher than the threshold of deterministic effect of 2 Gy and a look up table with the equation of fitness were proposed to be implemented in the facility for K (IRP) higher than 500 mGy.
引用
收藏
页码:2053 / 2060
页数:8
相关论文
共 50 条
  • [21] Estimating the maximum radiation skin dose during cardiac interventional procedures using the dose- area product, fluoroscopic time, and patient weight
    Chida, K.
    Saito, H.
    Otani, H.
    Zuguchi, M.
    Shimura, H.
    Nakada, M.
    Takahashi, S.
    Shirato, K.
    EUROPEAN HEART JOURNAL, 2005, 26 : 694 - 694
  • [22] Linear Regression for Modeling Common Fluoroscopy Dose Metrics in Cardiac Catheterization Procedures
    Mann, S.
    Wegermann, Z.
    Russ, M.
    Wei, T.
    Jones, W. S.
    MEDICAL PHYSICS, 2020, 47 (06) : E542 - E542
  • [23] Patient and staff radiation dose in fluoroscopy-guided TIPS procedures and dose reduction, using dedicated fluoroscopy exposure settings
    Zweers, D
    Geleijns, J
    Aarts, NJM
    Hardam, LJ
    Laméris, JS
    Scultz, FW
    Kool, LJS
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (846): : 672 - 676
  • [24] Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures
    Morrell, R. E.
    Rogers, A. T.
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (943): : 603 - 607
  • [25] Patient skin dose measurements during coronary interventional procedures using Gafchromic film
    Ying, C. K.
    Kandaiya, S.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2010, 30 (03) : 585 - 596
  • [26] Patient Exposure during Fluoroscopy-guided Pacemaker Implantation Procedures
    Alkhorayef, M.
    Sulieman, A.
    Babikir, E.
    Daar, E.
    Alnaaimi, M.
    Alduaij, M.
    Bradley, D.
    APPLIED RADIATION AND ISOTOPES, 2018, 138 : 14 - 17
  • [27] Estimation of patient skin dose from fluoroscopy time and number of angiographic frames during embolization and cerebral angiograms
    O'Dea, T
    Geise, R
    Ritenour, E
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2477 - 2480
  • [28] Calibration of Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures
    Morrell, RE
    Rogers, A
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (24): : 5559 - 5570
  • [29] Effect of patient support pads on image quality and dose in fluoroscopy
    Department of Radiology, University of Florida, Gamesville, FL 32610-0374, United States
    MED. PHYS., 1600, 3 (377-382):
  • [30] Effect of patient support pads on image quality and dose in fluoroscopy
    Geiser, WR
    Huda, W
    Gkanatsios, NA
    MEDICAL PHYSICS, 1997, 24 (03) : 377 - 382