Failure modes and effects analysis study for accelerator-based Boron Neutron Capture Therapy

被引:4
|
作者
Takemori, Mihiro [1 ,2 ,3 ]
Nakamura, Satoshi [1 ,3 ,4 ,9 ]
Sofue, Toshimitsu [5 ]
Ito, Mikiko [6 ]
Goka, Tomonori [5 ]
Miura, Yuki [7 ]
Iijima, Kotaro [1 ]
Chiba, Takahito [1 ,2 ]
Nakayama, Hiroki [1 ,2 ]
Nakaichi, Tetsu [1 ]
Mikasa, Shohei [1 ]
Takano, Yuki [1 ]
Kon, Mitsuhiro [1 ,5 ]
Shuto, Yasunori [1 ,5 ]
Urago, Yuka [1 ,2 ]
Nishitani, Masato [1 ,2 ]
Kashihara, Tairo [8 ]
Takahashi, Kana [8 ]
Murakami, Naoya [8 ]
Nishio, Teiji
Okamoto, Hiroyuki [1 ]
Chang, Weishan [2 ]
Igaki, Hiroshi [3 ,8 ]
机构
[1] Natl Canc Ctr, Div Radiat Safety & Qual Assurance, Chuo Ku, Tokyo, Japan
[2] Tokyo Metropolitan Univ, Grad Sch Human Hlth Sci, Dept Radiol Sci, Arakawa Ku, Tokyo, Japan
[3] Natl Canc Ctr, Div Boron Neutron Capture Therapy, Exploratory Oncol Res & Clin Trial Ctr, Chuo Ku, Tokyo, Japan
[4] Osaka Univ, Grad Sch Med, Div Hlth Sci, Med Phys Lab, Suita, Osaka, Japan
[5] Natl Canc Ctr, Dept Radiol Technol, Chuo Ku, Tokyo, Japan
[6] Natl Canc Ctr, Dept Nursing, Chuo Ku, Tokyo, Japan
[7] Natl Canc Ctr Hosp East, Dept Radiol Technol, Kashiwa, Chiba, Japan
[8] Natl Canc Ctr, Dept Radiat Oncol, Chuo Ku, Tokyo, Japan
[9] 5-1-1 Tsukiji, Chuo Ku, Tokyo 1040045, Japan
关键词
boron neutron capture therapy (BNCT); failure mode and effects analysis (FMEA); specific FM; RADIATION-THERAPY; RECURRENT; HEAD;
D O I
10.1002/mp.16104
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundBoron Neutron Capture Therapy (BNCT) has recently been used in clinical oncology thanks to recent developments of accelerator-based BNCT systems. Although there are some specific processes for BNCT, they have not yet been discussed in detail. PurposeThe aim of this study is to provide comprehensive data on the risk of accelerator-based BNCT system to institutions planning to implement an accelerator-based BNCT system. MethodsIn this study, failure mode and effects analysis (FMEA) was performed based on a treatment process map prepared for the accelerator-based BNCT system. A multidisciplinary team consisting of a medical doctor (MD), a registered nurse (RN), two medical physicists (MP), and three radiologic technologists (RT) identified the failure modes (FMs). Occurrence (O), severity (S), and detectability (D) were scored on a scale of 10, respectively. For each failure mode (FM), risk priority number (RPN) was calculated by multiplying the values of O, S, and D, and it was then categorized as high risk, very high risk, and other. Additionally, FMs were statistically compared in terms of countermeasures, associated occupations, and whether or not they were the patient-derived. ResultsThe identified FMs for BNCT were 165 in which 30 and 17 FMs were classified as high risk and very high risk, respectively. Additionally, 71 FMs were accelerator-based BNCT-specific FMs in which 18 and 5 FMs were classified as high risk and very high risk, respectively. The FMs for which countermeasures were "Education" or "Confirmation" were statistically significantly higher for S than the others (p = 0.019). As the number of BNCT facilities is expected to increase, staff education is even more important. Comparing patient-derived and other FMs, O tended to be higher in patient-derived FMs. This could be because the non-patient-derived FMs included events that could be controlled by software, whereas the patient-derived FMs were impossible to prevent and might also depend on the patient's condition. Alternatively, there were non-patient-derived FMs with higher D, which were difficult to detect mechanically and were classified as more than high risk. In O, significantly higher values (p = 0.096) were found for FMs from MD and RN associated with much patient intervention compared to FMs from MP and RT less patient intervention. Comparing conventional radiotherapy and accelerator-based BNCT, although there were events with comparable risk in same FMs, there were also events with different risk in same FMs. They could be related to differences in the physical characteristics of the two modalities. ConclusionsThis study is the first report for conducting a risk analysis for BNCT using FMEA. Thus, this study provides comprehensive data needed for quality assurance/quality control (QA/QC) in the treatment process for facilities considering the implementation of accelerator-based BNCT in the future. Because many BNCT-specific risks were discussed, it is important to understand the characteristics of BNCT and to take adequate measures in advance. If the effects of all FMs and countermeasures are discussed by multidisciplinary team, it will be possible to take countermeasures against individual FMs from many perspectives and provide BNCT more safely and effectively.
引用
收藏
页码:424 / 439
页数:16
相关论文
共 50 条
  • [31] Boron neutron capture therapy in the new age of accelerator-based neutron production and preliminary progress in Canada
    Dziura, Dominik
    Tabbassum, Sana
    MacNeil, Amanda
    Maharaj, Dalini D.
    Laxdal, Robert
    Kester, Oliver
    Pan, Ming
    Kumada, Hiroaki
    Marquardt, Drew
    CANADIAN JOURNAL OF PHYSICS, 2023, 101 (08) : 363 - 372
  • [32] Development of an online neutron beam monitoring system for accelerator-based boron neutron capture therapy in a hospital
    Takada, Masashi
    Yagi, Natsumi
    Nakamura, Satoshi
    Shimada, Kenzi
    Itami, Jyun
    Igaki, Hiroshi
    Nakamura, Masaru
    Nunomiya, Tomoya
    Endo, Satoru
    Kajimoto, Tsuyoshi
    Tanaka, Kenichi
    Aoyama, Kei
    Narita, Masakuni
    Nakamura, Takashi
    MEDICAL PHYSICS, 2024,
  • [33] An accelerator-based neutron source design with a thermal neutron port and an epithermal neutron port for boron neutron capture therapy
    Shuai, Ha
    Dian, Eszter
    Mezei, Ferenc
    Sipos, Peter
    Czifrus, Szabolcs
    APPLIED RADIATION AND ISOTOPES, 2025, 217
  • [34] Optimization of an accelerator-based epithermal neutron source for neutron capture therapy
    Kononov, OE
    Kononov, VN
    Bokhovko, MV
    Korobeynikov, VV
    Soloviev, A
    Sysoev, AS
    Gulidov, IA
    Chub, WT
    Nigg, DW
    APPLIED RADIATION AND ISOTOPES, 2004, 61 (05) : 1009 - 1013
  • [35] Accelerator tube construction and characterization for a tandem-electrostatic-quadrupole for accelerator-based boron neutron capture therapy
    Cartelli, D.
    Thatar Vento, V.
    Castell, W.
    Di Paolo, H.
    Kesque, J. M.
    Bergueiro, J.
    Valda, A. A.
    Erhardt, J.
    Kreiner, A. J.
    APPLIED RADIATION AND ISOTOPES, 2011, 69 (12) : 1680 - 1683
  • [36] Design of Beam Shaping Assembly for Accelerator-based Boron Neutron Capture Therapy and Study on Its Clinical Parameter
    Wang Y.
    Wang Z.
    Li N.
    Guan X.
    Gu L.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2022, 56 (07): : 1440 - 1447
  • [37] A feasibility study on a small electron linear accelerator-based beam shaping assembly for boron neutron capture therapy
    Hiraga, F.
    Prateepkaew, J.
    APPLIED RADIATION AND ISOTOPES, 2025, 217
  • [38] Tandem-ESQ for accelerator-based boron neutron capture therapy (AB-BNCT)
    Kreiner, A. J.
    Kwan, J. W.
    Burlon, A. A.
    Di Paolo, H.
    Henestroza, E.
    Minsky, D.
    Valda, A.
    Debray, M.
    Somacal, H. R.
    VI LATIN AMERICAN SYMPOSIUM ON NUCLEAR PHYSICS AND APPLICATIONS, 2007, 884 : 225 - 230
  • [39] Optimization of target, moderator, and collimator in the accelerator-based boron neutron capture therapy system: A Monte Carlo study
    Cheon, Bo-Wi
    Yoo, Dohyeon
    Park, Hyojun
    Lee, Hyun Cheol
    Shin, Wook-Geun
    Choi, Hyun Joon
    Hong, Bong Hwan
    Chung, Heejun
    Min, Chul Hee
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (06) : 1970 - 1978
  • [40] Accelerator-based boron neutron capture therapy and the development of a dedicated Tandem-Electrostatic-Quadrupole
    Kreiner, A. J.
    Di Paolo, H.
    Burlon, A. A.
    Kesque, J. M.
    Valda, A. A.
    Debray, M. E.
    Giboudot, Y.
    Levinas, P.
    Fraiman, M.
    Romeo, V.
    Somacal, H. R.
    Minsky, D. M.
    VII LATIN AMERICAN SYMPOSIUM ON NUCLEAR PHYSICS AND APPLICATIONS, 2007, 947 : 17 - 24