Case report: Extravascular catheter migration in a child: A rare complication of the totally implantable venous access device

被引:2
|
作者
Lu, Xuning [1 ]
Ren, Yueyi [2 ]
Wan, Hao [2 ]
Xu, Qiteng [2 ]
Duan, Shuhua [2 ,3 ]
机构
[1] Dalian Municipal Women & Childrens Med Ctr Grp, Heart Ctr, Dalian, Liaoning, Peoples R China
[2] Qingdao Univ, Qingdao Women & Childrens Hosp, Heart Ctr, Qingdao, Shandong, Peoples R China
[3] Qingdao Univ, Qingdao Women & Childrens Hosp, Heart Ctr, Qingdao 266034, Shandong, Peoples R China
关键词
child; complication; extravascular migration; pleural effusion; totally implantable venous access device; PORT CATHETER; JUGULAR-VEIN; ONCOLOGY; SYSTEMS; CUTDOWN;
D O I
10.1097/MD.0000000000032710
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background:Totally implantable venous access devices (TIVADs) are widely used to gain intermittent central venous access, such as in patients who need long-term chemotherapy, total parenteral nutrition, and long-term antibiotic treatment. At present, there are many complications associated with the use of these devices. Complete extravascular migration of TIVADs via the internal jugular vein is a very rare and potentially serious condition, especially in children. Case presentation:A 1-year-old girl needed palliative chemotherapy because of hepatoblastoma complicated by inferior vena cava thrombosis. A TIVAD was implanted through the right internal jugular vein with a routine heparin flushing tube. On the second day after the operation, a pale bloody liquid was drawn out from the device and the chest X-ray was checked to confirm that the position of the catheter was normal. On the third day after the operation, however, the patient's right respiratory sound was weakened on physical examination and auscultation. Fluoroscopy showed that the tip of the catheter was located in the right thoracic cavity, and there was a large amount of effusion in the right thoracic cavity. The pleural effusion was removed, the TIVAD was replaced again, and the child was discharged 2 days later. Conclusions:Following TIVAD implantation, if abnormalities are found, in addition to chest X-ray, saline flush and echocardiography should be performed to determine the position of the catheter and rule out extravascular migration of the catheter to avoid irreparable consequences.
引用
收藏
页数:3
相关论文
共 50 条
  • [21] Femoral placement of a totally implantable venous access port with spontaneous catheter fracture: case report
    Tomohiro Kondo
    Shigemi Matsumoto
    Keitaro Doi
    Motoo Nomura
    Manabu Muto
    CVIR Endovascular, 3
  • [22] Hepatic migration of a catheter fragment followed by disconnection of a totally implantable venous access port
    Wang, C.-S.
    Yang, C.-J.
    Chen, S.-C.
    Chen, H.-C.
    Huang, Ming-Shyan
    International Journal of Artificial Organs, 2008, 31 (12): : 1059 - 1061
  • [23] Hepatic migration of a catheter fragment followed by disconnection of a totally implantable venous access port
    Wang, C. -S.
    Yang, C. -J.
    Chen, S. -C.
    Chen, H. -C.
    Huang, M. -S.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2008, 31 (12): : 1059 - 1061
  • [24] TECHNIQUE FOR PLACEMENT OF A TOTALLY IMPLANTABLE VENOUS ACCESS DEVICE
    KONDI, ES
    PIETRAFITTA, JJ
    BARRIOLA, JA
    JOURNAL OF SURGICAL ONCOLOGY, 1988, 37 (04) : 272 - 277
  • [25] Intrapulmonal dislocation of a totally implantable venous access device
    Hackert T.
    Tjaden C.
    Kraft A.
    Sido B.
    Dienemann H.
    Buchler M.W.
    World Journal of Surgical Oncology, 3 (1)
  • [26] Spontaneous catheter tip migration in Totally Implantable Venous Access Port (TIVAP) to right atrium
    Wetan, Ni Gusti A. A. Manik Yuniawaty
    Irawan, Hendry
    Adiputra, Putu Anda Tusta
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2024, 123
  • [27] Malposition of a catheter in the persistent left superior vena cava - A rare complication of totally implantable venous devices
    Kao, CL
    Chang, JP
    JOURNAL OF CARDIOVASCULAR SURGERY, 2003, 44 (01): : 145 - 147
  • [28] Intramuscular Migration of Venous Catheter as a Rare Complication of Ventriculoatrial Shunt: Case Report and Literature Review
    Hung, Cheng-Che
    Chuang, Hao-Yu
    Lin, Hung-Lin
    Chu, Yen-Tse
    Cheng, Cheng Hsin
    JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2017, 78 (04) : 412 - 416
  • [29] Complete Catheter Disconnection and Migration of an Implantable Venous Access Device: The Disconnected Cap Sign
    Faraj, Walid
    Zaghal, Ahmad
    El-Beyrouthy, Ornella
    Kutoubi, Aghiad
    ANNALS OF VASCULAR SURGERY, 2010, 24 (05) : 692.e11 - 692.e15
  • [30] Totally implantable venous access device infections in patients with AIDS
    Hellard, M
    Fuller, A
    Spelman, D
    Spicer, WJ
    Pickles, R
    Jenney, A
    AIDS, 1997, 11 (05) : 697 - 698