Evaluation of PA990pro, an Analyzer for C-Reactive Protein Using Whole Blood: Hematocrit Correction is Required

被引:0
|
作者
Dong, Feng [1 ]
Chen, Yanyan [1 ]
Yao, Yong [1 ]
Wu, Jun [1 ,2 ,3 ]
机构
[1] Beijing Jishuitan Hosp, Dept Clin Lab, Beijing, Peoples R China
[2] Peking Univ, Sch Clin Med 4, Beijing, Peoples R China
[3] Peking Univ, Beijing Jishuitan Hosp, Sch Clin Med 4, Beijing 100035, Peoples R China
关键词
C-reactive protein; performance evaluation; hematocrit; correction;
D O I
10.7754/Clin.Lab.2022.221005)
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The rapid detection of C-reactive protein (CRP) in whole blood can provide a basis for reducing the use of antibiotics, especially for infants in whom blood collection is difficult. Whether the performance of PA990pro for CRP detection can meet clinical needs has not been studied.Methods: Between May and June 2022, 230 blood samples were collected to explore the analytical performance of the PA990pro for CRP detection. The blank check, carryover, repeatability, intermediate precision, linearity, sample stability, the influence of hematocrit (HCT)/triglyceride/bilirubin, and the trueness of the PA990pro were evaluated. Whole blood CRP test results analyzed using the PA990pro were compared to plasma CRP test results from a Hitachi 7180 biochemical analyzer, using the same samples.Results: The blank check (& LE; 0.3 mg/L), carryovers (& LE; 0.05%), repeatability (& LE; 7.23%), and intermediate precision (& LE; 7.36%) can meet clinical needs. The linear correlation coefficients of different ranges of CRP were good (r >0.975), and the slopes were all within 0.950 - 1.050. The stability of samples was good within 72 hours whether stored at 18 - 25 & DEG;C or 2 - 8 & DEG;C (CV < 10%). With interference from triglycerides at & LE; 7 mmol/L, the deviation of CRP was < 10% and with bilirubin at & LE; 216 & mu;mol/L, the deviation of CRP was < 10%. PA990pro has no HCT quantification function; abnormal HCT will significantly interfere with whole blood CRP results (the relative deviation in the basic experiment was up to 73.71%). We suggest that the HCT results of the patient during the same period are provided through the laboratory information system (LIS) so that a CRP correction formula (CRPcorrected = CRPmeasured*(1 - 40%)/(1 - HCTmeasured)) can be used on the LIS. After using the formula to perform the HCT correction, the results of PA990pro were well correlated with the 7180 analyzer plasma CRP detection (r > 0.975). PA990pro could pass the external quality assessment of the National Center for Clinical Laboratories.Conclusions: The CRP detection performance of PA990pro can meet clinical needs, but it is suggested that HCT should be corrected by the formula set in LIS. It is a simple, rapid, and cost-free approach that can be used to obtain a modified whole blood CRP test result that meets clinical needs.
引用
收藏
页码:1257 / 1268
页数:12
相关论文
共 50 条
  • [41] Quantification of C-Reactive protein in human blood plasma using near-infrared Raman spectroscopy
    Bergholt, M. S.
    Hassing, S.
    ANALYST, 2009, 134 (10) : 2123 - 2127
  • [42] Combined detection of C-reactive protein and PBMC quantification from whole blood in an integrated lab-on-a-disc microfluidic platform
    Uddin, Rokon
    Donolato, Marco
    Hwu, En-Te
    Hansen, Mikkel Fougt
    Boisen, Anja
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 272 : 634 - 642
  • [43] C-reactive protein (CRP) evaluation in human urine using optical sensor supported by machine learning
    Cierpiak, Kacper
    Wityk, Pawel
    Kosowska, Monika
    Sokolowski, Patryk
    Talaska, Tomasz
    Gierowski, Jakub
    Markuszewski, Michal J.
    Szczerska, Malgorzata
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [44] THE ROLE OF PRO-ADRENOMEDULLIN IN THE DIAGNOSIS OF URINARY TRACT INFECTION IN CHILDREN AND COMPARISON WITH C-REACTIVE PROTEIN, PROCALCITONIN, BLOOD LEUKOCYTE COUNT
    Atis, Seyma Koksal
    Kiral, Asuman
    Goknar, Nilufer
    PEDIATRIC NEPHROLOGY, 2021, 36 (10) : 3370 - 3370
  • [45] p38 Inhibition attenuates the pro-inflammatory response to C-reactive protein by human peripheral blood mononuclear cells
    Lim, MY
    Wang, H
    Kapoun, AM
    O'Connell, M
    O'Young, G
    Brauer, HA
    Luedtke, GR
    Chakravarty, S
    Dugar, S
    Schreiner, GS
    Protter, AA
    Higgins, LS
    JOURNAL OF MOLECULAR AND CELLULAR CARDIOLOGY, 2004, 37 (06) : 1111 - 1114
  • [46] Complete blood count and C-reactive protein to predict positive blood culture among neonates using machine learning algorithms
    Matsushita, Felipe Yu
    Krebs, Vera Lucia Jornada
    de Carvalho, Werther Brunow
    CLINICS, 2023, 78
  • [47] Assessing the association between cigarette smoking and blood C-reactive protein levels using restructured cohort data
    Koh, Dong-Hee
    Choi, Sangjun
    Park, Ju-Hyun
    Lee, Sang-Gil
    Kim, Hwan-Cheol
    Kim, Inah
    Han, Soon-Sil
    Park, Dong-Uk
    PREVENTIVE MEDICINE, 2024, 189
  • [48] Development of a canine blood C-reactive protein-measuring device using a flow-type immunosensor
    Tomoko Kubota
    Norio Tateishi
    Hideki Toita
    Nobutoshi Kanaki
    Akihisa Hata
    Noboru Fujitani
    Analytical Sciences, 2022, 38 : 1269 - 1276
  • [49] Development of a canine blood C-reactive protein-measuring device using a flow-type immunosensor
    Kubota, Tomoko
    Tateishi, Norio
    Toita, Hideki
    Kanaki, Nobutoshi
    Hata, Akihisa
    Fujitani, Noboru
    ANALYTICAL SCIENCES, 2022, 38 (10) : 1269 - 1276
  • [50] Opposing effects of C-reactive protein isoforms on shear-induced neutrophil-platelet adhesion and neutrophil aggregation in whole blood
    Khreiss, T
    József, L
    Potempa, LA
    Filep, JG
    CIRCULATION, 2004, 110 (17) : 2713 - 2720