Innovative experimental animal models for real-time comparison of antithrombogenicity between two oxygenators using dual extracorporeal circulation circuits and indocyanine green fluorescence imaging
被引:0
|作者:
Sakurai, Hironobu
[1
]
Fujiwara, Tatsuki
[1
]
Ohuchi, Katsuhiro
[2
]
Hijikata, Wataru
[3
]
Inoue, Yusuke
[4
]
Maruyama, Osamu
[5
]
Tahara, Tomoki
[1
]
Yokota, Sachie
[6
]
Tanaka, Yui
[3
]
Takewa, Yoshiaki
[4
]
Mizuno, Tomohiro
[1
]
Arai, Hirokuni
[1
]
机构:
[1] Tokyo Med & Dent Univ, Grad Sch Med & Dent Sci, Dept Cardiovasc Surg, Tokyo, Japan
[2] Tokyo Med & Dent Univ, Ctr Expt Anim, Tokyo, Japan
[3] Tokyo Inst Technol, Sch Engn, Tokyo, Japan
[4] Asahikawa Med Univ, Adv Med Engn Res Ctr, Asahikawa, Hokkaido, Japan
[5] Natl Inst Adv Ind Sci & Technol, Hlth & Med Res Inst, Tsukuba, Ibaraki, Japan
[6] Tokyo Med & Dent Univ, Fac Med, Tokyo, Japan
基金:
日本学术振兴会;
关键词:
antithrombotic comparison;
COVID-19;
dual extracorporeal circulation;
experimental animal model;
extracorporeal membrane oxygenation;
indocyanine green fluorescence imaging;
oxygenator;
pulsatile flow;
CENTRIFUGAL BLOOD PUMP;
MEDTECH MAG-LEV;
IN-VITRO;
MEMBRANE-OXYGENATION;
SINGLE-USE;
COVID-19;
SUPPORT;
PORCINE;
BYPASS;
D O I:
10.1111/aor.14380
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
Background Antithrombogenicity of extracorporeal membrane oxygenation (ECMO) devices, particularly oxygenators, is a current problem, with numerous studies and developments underway. However, there has been limited progress in developing methods to accurately compare the antithrombogenicity of oxygenators. Animal experiments are commonly conducted to evaluate the antithrombogenicity of devices; however, it is challenging to maintain a steady experimental environment. We propose an innovative experimental animal model to evaluate different devices in a constant experimental environment in real-time. Methods This model uses two venous-arterial ECMO circuits attached to one animal (one by jugular vein and carotid artery, one by femoral vein and artery) and real-time assessment of thrombus formation in the oxygenator by indocyanine green (ICG) fluorescence imaging. Comparison studies were conducted using three pigs: one to compare different oxygenators (MERA vs. CAPIOX) (Case 1), and two to compare antithrombotic properties of the oxygenator (QUADROX) when used under different hydrodynamic conditions (continuous flow vs. pulsatile flow) (Cases 2 and 3). Results Thrombi, visualized using ICG imaging, appeared as black dots on a white background in each oxygenator. In Case 1, differences in the site of thrombus formation and rate of thrombus growth were observed in real-time in two oxygenators. In Case 2 and 3, the thrombus region was smaller in pulsatile than in continuous conditions. Conclusions We devised an innovative experimental animal model for comparison of antithrombogenicity in ECMO circuits. This model enabled simultaneous evaluation of two different ECMO circuits under the same biological conditions and reduced the number of sacrificed experimental animals.