Impact of deep learning-based image reconstruction on image quality and lesion visibility in renal computed tomography at different doses

被引:3
|
作者
Bie, Yifan [1 ]
Yang, Shuo [1 ]
Li, Xingchao [1 ]
Zhao, Kun [1 ]
Zhang, Changlei [1 ]
Zhong, Hai [1 ,2 ]
机构
[1] Shandong Univ, Hosp 2, Dept Radiol, Jinan, Peoples R China
[2] Shandong Univ, Hosp 2, Dept Radiol, 247 Beiyuan Rd, Jinan 250033, Peoples R China
关键词
Deep learning image reconstruction; adaptive statistical iterative reconstruction-Veo (ASiR-V); image quality; tomography; kidney; ITERATIVE RECONSTRUCTION; ABDOMINAL CT; VERSION;
D O I
10.21037/qims-22-852
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Numerous computed tomography (CT) image reconstruction algorithms have been developed to improve image quality, and high-quality renal CT images are crucial to clinical diagnosis. This study evaluated the image quality and lesion visibility of deep learning-based image reconstruction (DLIR) compared with adaptive statistical iterative reconstruction-Veo (ASiR-V) in contrast-enhanced renal CT at different reconstruction strengths and doses. Methods: From January 2020 to May 2021, we prospectively included 101 patients who underwent renal contrast-enhanced CT scanning (69 at 120 kV; 32 at 80 kV). All image data were reconstructed with ASiR-V (30% and 70%) and DLIR at low, medium, and high reconstruction strengths (DLIR-L, DLIR-M, and DLIR-H, respectively). The CT number, noise, noise reduction rate (NRR), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), overall image quality, and the proportion of acceptable images were compared. Lesions of DLIR groups were evaluated, and the conspicuity-to-noise ratio (C/N) was calculated.Results: Quantitative values (noise, SNR, CNR, and NRR) significantly differed between all reconstructions at 120 and 80 kV (P<0.00 1) and between each reconstruction, except ASiR-V 70% vs. DLIR-M. At 120 kV, the overall image quality and the proportion of acceptable images significantly differed between all reconstructions (P<0.001) and between each reconstruction, except ASiR-V 30% vs. DLIR-L and ASiR-V 70% vs. DLIR-M. At 80 kV, the overall image quality significantly differed between all reconstructions (P<0.001) and between each reconstruction, except between ASiR-V 30%, ASiR-V 70%, and DLIR-L. Quantitative and qualitative values were highest in DLIR-H, while the values were close in DLIR-H (80 kV) vs. ASiR-V 70% (120 kV) and DLIR-M (80 kV) vs. ASiR-V 30% (120 kV). The lesion conspicuity and noise significantly differed in DLIR at 120 kV and 80 kV (P<0.001). C/N significantly differed in DLIR at 120 kV (P<0.001) but not at 80 kV. DLIR-L and DLIR-M exhibited much-improved lesion display (C/N >1), and DLIR-H exhibited much-improved noise (C/N <1) at 120 kV. Conclusions: DLIR significantly improved the image quality and lesion visibility of renal CT compared with ASiR-V, even at a low dose.
引用
收藏
页码:2197 / 2207
页数:11
相关论文
共 50 条
  • [41] Optimizing computed tomography image reconstruction for focal hepatic lesions: Deep learning image reconstruction vs iterative reconstruction
    Jaruvongvanich, Varin
    Muangsomboon, Kobkun
    Teerasamit, Wanwarang
    Suvannarerg, Voraparee
    Komoltri, Chulaluk
    Thammakittiphan, Sastrawut
    Lornimitdee, Wimonrat
    Ritsamrej, Witchuda
    Chaisue, Parinya
    Pongnapang, Napapong
    Apisarnthanarak, Piyaporn
    HELIYON, 2024, 10 (15)
  • [42] Deep learning-based extended field of view computed tomography image reconstruction: influence of network design on image estimation outside the scan field of view
    Khural, Bhupinder Singh
    Baer-Beck, Matthias
    Fournie, Eric
    Stierstorfer, Karl
    Huang, Yixing
    Maier, Andreas
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (02)
  • [43] Comparison of Noise Analysis and Deep Learning-Based Image Quality Assessment (IQA) Methods for Thoracic Computed Tomography (CT)
    Grant, B.
    Lee, J.
    Chung, I.
    Reiser, I.
    Lan, L.
    Papaioannou, J.
    Giger, M.
    MEDICAL PHYSICS, 2018, 45 (06) : E493 - E493
  • [44] Deep Learning-Based Computed Tomography Image Features in the Detection and Diagnosis of Perianal Abscess Tissue
    Han, Song
    Yang, Jun
    Xu, Jihua
    JOURNAL OF HEALTHCARE ENGINEERING, 2021, 2021 (2021)
  • [45] Physical characteristics of deep learning-based image processing software in computed tomography: a phantom study
    Sato, Seiya
    Urikura, Atsushi
    Mimatsu, Makoto
    Miyamae, Yuta
    Jibiki, Yuji
    Yamashita, Mami
    Ishihara, Toshihiro
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1713 - 1721
  • [46] Physical characteristics of deep learning-based image processing software in computed tomography: a phantom study
    Seiya Sato
    Atsushi Urikura
    Makoto Mimatsu
    Yuta Miyamae
    Yuji Jibiki
    Mami Yamashita
    Toshihiro Ishihara
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1713 - 1721
  • [47] Improved image quality in abdominal computed tomography reconstructed with a novel Deep Learning Image Reconstruction technique - initial clinical experience
    Njolstad, Tormund
    Schulz, Anselm
    Godt, Johannes C.
    Brogger, Helga M.
    Johansen, Cathrine K.
    Andersen, Hilde K.
    Martinsen, Anne Catrine T.
    ACTA RADIOLOGICA OPEN, 2021, 10 (04)
  • [48] Computed Tomography Effective Dose and Image Quality in Deep Learning Image Reconstruction in Intensive Care Patients Compared to Iterative Algorithms
    Quaia, Emilio
    de Cristoforis, Elena Kiyomi Lanza
    Agostini, Elena
    Zanon, Chiara
    TOMOGRAPHY, 2024, 10 (06) : 912 - 921
  • [49] Deep learning based image reconstruction algorithm for limited-angle translational computed tomography
    Wang, Jiaxi
    Liang, Jun
    Cheng, Jingye
    Guo, Yumeng
    Zeng, Li
    PLOS ONE, 2020, 15 (01):
  • [50] Deep Learning-Based Versus Iterative Image Reconstruction for Unenhanced Brain CT: A Quantitative Comparison of Image Quality
    Cozzi, Andrea
    Ce, Maurizio
    De Padova, Giuseppe
    Libri, Dario
    Caldarelli, Nazarena
    Zucconi, Fabio
    Oliva, Giancarlo
    Cellina, Michaela
    TOMOGRAPHY, 2023, 9 (05) : 1629 - 1637