Comparison of Home and Office Blood Pressure Devices in the Clinical Setting

被引:3
|
作者
Juraschek, Stephen P. [1 ,2 ,3 ]
Vyavahare, Medha [2 ]
Cluett, Jennifer L. [1 ,2 ,3 ]
Turkson-Ocran, Ruth-Alma [1 ,2 ,3 ]
Mukamal, Kenneth J. [1 ,2 ]
Ishak, Anthony M. [2 ,3 ]
机构
[1] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Med, Boston, MA 02115 USA
[2] Harvard Med Sch, Dept Med, Boston, MA 02115 USA
[3] Beth Israel Deaconess Med Ctr, Healthcare Associates, Dept Med, Boston, MA 02215 USA
关键词
blood pressure; home blood pressure monitoring; hypertension; pharmacist; self-measured blood pressure; validation; VALIDATION; STANDARD;
D O I
10.1093/ajh/hpad120
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
BACKGROUND: Self-measured blood pressure (SMBP) monitoring is increasingly used for remote hypertension management, but the real-world performance of home blood pressure (BP) devices is unknown. We examined BP measurements from patients' home devices using the American Medical Association's (AMA) SMBP Device Accuracy Test tool.METHODS: Patients at a single internal medicine clinic underwent up to five seated, same-arm BP readings using a home device and an automated BP device (Omron HEM-907XL). Following the AMA's three-step protocol, we used the patient's home device for the first, second, and fourth measurements and the office device for the third and fifth (if needed) measurements. Device agreement failure was defined as an absolute difference in systolic BP >10 mm Hg between the home and office devices in either of two confirmatory steps. Performance was examined by brand (Omron vs. non-Omron). Moreover, we examined patient factors associated with agreement failure via logistic regression models adjusted for demographic characteristics.RESULTS: We evaluated 152 patients (mean age 60 +/- 15 years, 58% women, 31% Black) seen between October 2020 and November 2021. Device agreement failure occurred in 22.4% (95% CI: 16.4%, 29.7%) of devices tested, including 19.1% among Omron devices and 27.6% among non-Omron devices (P = 0.23). No patient characteristics were associated with agreement failure.CONCLUSIONS: Over one-fifth of home devices did not agree based on the AMA SMBP device accuracy protocol. These findings confirm the importance of office-based device comparisons to ensure the accuracy of home BP monitoring.
引用
收藏
页码:342 / 348
页数:7
相关论文
共 50 条
  • [31] ASSOCIATIONS OF OFFICE BRACHIAL BLOOD PRESSURE, OFFICE CENTRAL BLOOD PRESSURE, AND HOME BRACHIAL BLOOD PRESSURE WITH ARTERIAL STIFFNESS
    Mizuno, Hiroyuki
    Hishide, Satoshi
    Nozue, Ryoko
    Shimbo, Daichi
    Kario, Kazuomi
    JOURNAL OF HYPERTENSION, 2023, 41 : E190 - E191
  • [32] Home, ambulatory, and office blood pressure measurements
    Mora, C
    Navarro, JF
    ARCHIVES OF INTERNAL MEDICINE, 2000, 160 (18) : 2865 - 2866
  • [33] Comparison of blood pressure measurements obtained in the home setting: analysis of the Health Measures at Home Study
    Nwankwo, Tatiana
    Gindi, Renee
    Chen, Te-Ching
    Galinsky, Adena
    Miller, Ivey
    Terry, Ana
    BLOOD PRESSURE MONITORING, 2016, 21 (06) : 327 - 334
  • [34] Clinical Impact of after-consult clinic blood pressure: comparison with automated office blood pressure
    Lee, Cheol Ho
    Ahn, Ji Hun
    Ryu, Joon Ha
    Choi, Woong Gil
    CLINICAL HYPERTENSION, 2021, 27 (01)
  • [35] Clinical Impact of after-consult clinic blood pressure: comparison with automated office blood pressure
    Cheol Ho Lee
    Ji Hun Ahn
    Joon Ha Ryu
    Woong Gil Choi
    Clinical Hypertension, 27
  • [36] Home blood pressure measurement with an automatic device (OMRON IC): acceptance and comparison with office blood pressure measurement.
    Gomes, MAM
    Plerin, A
    Alavarce, DC
    Mion, D
    AMERICAN JOURNAL OF HYPERTENSION, 1999, 12 (04) : 165A - 165A
  • [37] MATHEMATICAL COMPARISON OF THE SEQUENTIAL VARIABLE RATIOS AMONG THE BLOOD PRESSURE VARIABILITIES OF AMBULATORY, HOME, AND OFFICE BLOOD PRESSURE MONITORING
    Ohno, Yoichi
    JOURNAL OF HYPERTENSION, 2023, 41 : E219 - E219
  • [38] Comparison of ambulatory blood pressure measurement with home, office and pharmacy measurements: is arterial blood pressure measured at pharmacy reliable?
    Mutlu, Sinan
    Sari, Oktay
    Arslan, Erol
    Aydogan, Umit
    Doganer, Yusuf C.
    Koc, Bayram
    JOURNAL OF EVALUATION IN CLINICAL PRACTICE, 2016, 22 (01) : 40 - 45
  • [39] Comparison of office, ambulatory and home blood pressure in children and adolescents on the basis of normalcy tables
    G S Stergiou
    N Karpettas
    D B Panagiotakos
    A Vazeou
    Journal of Human Hypertension, 2011, 25 : 218 - 223
  • [40] HOME BLOOD-PRESSURE - VARIABILITY, COMPARISON WITH OFFICE READINGS AND PROPOSAL FOR REFERENCE VALUES
    DEGAUDEMARIS, R
    CHAU, NP
    MALLION, JM
    JOURNAL OF HYPERTENSION, 1994, 12 (07) : 831 - 838