Unveiling Peripherally Inserted Central Catheter Fractures and Related Complications in the Neonatal Intensive Care Unit: A Concise Review

被引:2
|
作者
Balasundaram, Palanikumar [1 ]
Lucena, Michelle H. [2 ]
Jiang, Lan [3 ]
Nafday, Suhas [4 ]
机构
[1] Javon Bea Hosp Riverside, Mercy Hlth Syst, Dept Pediat, Div Neonatol, Rockford, IL 61114 USA
[2] Baylor Coll Med, Texas Childrens Hosp, Dept Pediat, Div Neonatol, Houston, TX USA
[3] Albert Einstein Coll Med, Childrens Hosp Montefiore, Dept Pediat, Bronx, NY USA
[4] Albert Einstein Coll Med, Jack D Weiler Hosp, Childrens Hosp Montefiore, Dept Pediat,Div Neonatol, Bronx, NY USA
关键词
neonatal intensive care unit (nicu); catheter jamming; fracture; picc line complication; peripherally inserted central catheter (picc); CENTRAL VENOUS CATHETER; PERCUTANEOUS RETRIEVAL; SILASTIC CATHETER; REMOVAL; EMBOLIZATION; RETENTION; CHILDREN;
D O I
10.7759/cureus.47572
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Peripherally inserted central catheters (PICCs) have become popular over tunneled catheters in neonatal intensive care units (NICUs) due to their ease of use and convenience. Although rare, a PICC fracture can be a severe and potentially fatal complication. This narrative review aims to identify factors predisposing neonates to PICC fracture and related complications, such as catheter jamming, and explore strategies for preventing and detecting this complication. A thorough search of PubMed and Google Scholar was conducted using relevant keywords to identify articles discussing PICC fracture in neonates. The review encompassed English-language literature on PICC fracture in neonates, with additional pertinent publications identified through citation searching. The incidence of PICC fracture in neonates varies from less than 1% to 10%, with a higher risk associated with prolonged catheterization, lower gestational age and lower birth weight, and the use of multi-lumen catheters. PICC fractures can occur during insertion, maintenance, or removal. Factors such as catheter duration, gestational age, birth weight, and catheter type increase the risk of PICC fracture. Excessive syringe pressure, securement failure, and excessive force during removal are contributing factors. Catheter fatigue and thin-walled catheter design are common causes of breakage. Preventive measures include proper training of healthcare providers, regular monitoring, early recognition, and prompt catheter removal upon fracture. Preventing and detecting PICC fractures is crucial for neonatal safety. Vigilance during insertion, maintenance, and removal, along with care to avoid excessive force during removal and high pressure during flushing, can help prevent catheter breakage. More research is required to improve prevention strategies for PICC fractures in neonates.
引用
收藏
页数:11
相关论文
共 50 条
  • [11] Assessment of the complications of peripherally inserted central catheter in neonates admitted to the intensive care unit: A center's experience in Iran
    Moradi, Raheleh
    Shariat, Mamak
    Farrokhzad, Nahid
    Abroon, Ameneh
    Kafi, Soheila
    Hamidpoor, Azadeh
    Hassani, Elaheh
    Charousaei, Hasti
    Nayeri, Fatemeh Sadat
    IRANIAN JOURNAL OF NURSING AND MIDWIFERY RESEARCH, 2022, 27 (06) : 505 - 508
  • [12] Limited Ultrasound Protocol for Upper Extremity Peripherally Inserted Central Catheter Monitoring: A Pilot Study in the Neonatal Intensive Care Unit
    Motz, Patrick
    von Arnim, Amelie von Saint Andre
    Likes, Maggie
    Chabra, Shilpi
    Traudt, Christopher
    Iyer, Ramesh S.
    Dighe, Manjiri
    JOURNAL OF ULTRASOUND IN MEDICINE, 2019, 38 (05) : 1341 - 1347
  • [13] Use of peripherally inserted central catheters and occurrence of bloodstream infections in a neonatal intensive care unit
    Yonekura, Marcia Yumi
    Abramczyk, Marcelo Luiz
    Lapchik, Milton Soibelmann
    VIGILANCIA SANITARIA EM DEBATE-SOCIEDADE CIENCIA & TECNOLOGIA, 2015, 3 (01): : 93 - 96
  • [14] Complications after peripherally inserted central catheter versus central venous catheter implantation in intensive care unit: propensity score analysis using a nationwide database
    Takahashi, Toshiaki
    Morita, Kojiro
    Uda, Kazuaki
    Matsui, Hiroki
    Yasunaga, Hideo
    Nakagami, Gojiro
    EXPERT REVIEW OF MEDICAL DEVICES, 2024, 21 (05) : 447 - 453
  • [15] Diagnostic accuracy of ultrasound for localising peripherally inserted central catheter tips in infants in the neonatal intensive care unit: a systematic review and meta-analysis
    Doyle, Shauna C.
    Bergin, Niamh M.
    Young, Rena
    England, Andrew
    McEntee, Mark F.
    PEDIATRIC RADIOLOGY, 2022, 52 (12) : 2421 - 2430
  • [16] Diagnostic accuracy of ultrasound for localising peripherally inserted central catheter tips in infants in the neonatal intensive care unit: a systematic review and meta-analysis
    Shauna C. Doyle
    Niamh M. Bergin
    Rena Young
    Andrew England
    Mark F. McEntee
    Pediatric Radiology, 2022, 52 : 2421 - 2430
  • [17] Peripherally inserted central catheter design and material for reducing catheter failure and complications (Review)
    Schults, Jessica A.
    Kleidon, Tricia
    Charles, Karina
    Young, Emily Rebecca
    Ullman, Amanda J.
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2024, (06):
  • [18] Factors associated with infection from the use of peripherally inserted central catheters in a Neonatal Intensive Care unit
    Duarte, Elysangela Dittz
    Pimenta, Adriano Marcal
    Noelly e Silva, Barbara Christina
    de Paula, Clarissa Moura
    REVISTA DA ESCOLA DE ENFERMAGEM DA USP, 2013, 47 (03) : 546 - 553
  • [19] Comparison of catheter-related large vein thrombosis in centrally inserted versus peripherally inserted central venous lines in the neurological intensive care unit
    Wilson, Thomas J.
    Stetler, William R., Jr.
    Fletcher, Jeffrey J.
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2013, 115 (07) : 879 - 882
  • [20] Impact of a Surgeon-Led Peripherally Inserted Central Venous Catheter Team on Peripherally Inserted Central Venous Catheter-Related Complications and Costs
    Pernar, Luise I. M.
    Wolf, Lindsay L.
    Seshadri, Anupamaa
    Patel, Vihas
    SURGICAL INFECTIONS, 2016, 17 (03) : 352 - 356