Analysis of Reported Adverse Events Related to Over-the-Scope Clips: A MAUDE Database Analysis

被引:2
|
作者
Abbas, Daniyal [1 ]
Abdallah, Mohamed [2 ]
Ahmed, Khalid [3 ]
Abdalla, Abubaker O. [4 ]
Mcdonald, Nicholas [2 ]
Umar, Shifa [5 ]
Hanson, Brian J. [2 ]
Bilal, Mohammad [2 ,6 ]
机构
[1] East Carolina Univ, Dept Internal Med, Greenville, NC USA
[2] Univ Minnesota, Dept Med, Div Gastroenterol Hepatol & Nutr, Minneapolis, MN USA
[3] Wright Ctr Community Hlth, Dept Internal Med, Scranton, PA USA
[4] Emory Univ, Div Gastroenterol & Hepatol, Atlanta, GA USA
[5] Mayo Clin, Dept Gastroenterol & Hepatol, Rochester, MN USA
[6] Div Gastroenterol Hepatol & Nutr, 406 Harvard St SE, MMC36, Minneapolis, MN 55455 USA
来源
TECHNIQUES AND INNOVATIONS IN GASTROINTESTINAL ENDOSCOPY | 2023年 / 25卷 / 02期
关键词
Over-the-scope clip; Adverse events; MAUDE database; Endoscopic closure; DUODENAL PERFORATION; ENDOSCOPIC TREATMENT; CLOSURE; SYSTEM; DEVICE; EXPERIENCE; FISTULAS; OVESCO; TRACT;
D O I
10.1016/j.tige.2022.11.005
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
BACKGROUND AND AIMS: Over-the-scope clips (OTSCs) have been increasingly used in the management of bleeding, perforations, fistulae, and anastomotic leaks in the gastrointestinal tract. However, since their Food and Drug Administration (FDA) approval in 2010, data on the adverse events of OTSC systems have been lacking. The aim of our study was to report and analyze adverse events and device failures associated with OTSC systems using the FDA's Manufacturer and User Facility Device Experience (MAUDE) database.METHODS: We analyzed the FDA MAUDE database post-marketing surveillance data from January 2013 through May 2022 on the two OTSC systems available in the United States.RESULTS: Forty medical device reporting claims were found from January 2013 through May 2022. Adverse events were classified as device-related problems and patient-related adverse events. Forty device-related problems were reported, along with 20 device-related adverse events. Most device related problems were reported in the Padlock defect closure system (n = 23), followed by the Padlock pro clip system (n = 8) and Ovesco OTSC system (n = 9). The most common device-related problem was related to the failure of OTSC deployment (n = 13), followed by material protrusion or extrusion (n = 7). The most common patient adverse events were perforation (n = 4), esophageal laceration (n = 4), bleeding (n = 3), and luminal stenosis (n = 3). All 4 patients required surgical management of the perforation. No deaths related to the use of OTSCs were reported.CONCLUSION: Failure of OTSC deployment and perforation were the most common device-related problem and patient-related adverse event, respectively. As the use of OTSC systems continues to increase, it is important for endoscopists to be mindful of both the common and rare adverse events associated with OTSC use.
引用
收藏
页码:106 / 112
页数:7
相关论文
共 50 条
  • [41] Use of over-the-scope clips in the colon in clinical practice: results from a German administrative database
    Horenkamp-Sonntag, Dirk
    Liebentraut, Judith
    Engel, Susanne
    Skupnik, Christoph
    Albers, David
    Schumacher, Brigitte
    Koop, Herbert
    ENDOSCOPY, 2020, 52 (12) : 1103 - 1110
  • [42] Adverse events related to AtriCure EPi-Sense Coagulation Device-Analysis of the FDA MAUDE database
    Sheth, Aakash R.
    Al Yafeai, Zaki
    Dominic, Paari
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (02) : 151 - 153
  • [43] Adverse Events Associated With Corticosteroid-Eluting Sinus Stents: A MAUDE Database Analysis
    Narwani, Vishal
    Torabi, Sina J.
    Kasle, David A.
    Patel, Rahul A.
    Lerner, Michael Z.
    Manes, R. Peter
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2022, 166 (01) : 179 - 182
  • [44] Adverse Events Associated With ClariFix Posterior Nasal Nerve Cryoablation: A MAUDE Database Analysis
    Singh, Amrita K.
    Kasle, David A.
    Torabi, Sina J.
    Manes, R. Peter
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2021, 165 (04) : 597 - 601
  • [45] Analysis of Hydroxychloroquine Adverse Events Reported in the WHO Pharmacovigilance Database
    Ahmed, Nehad J.
    Alshehri, Ahmed M.
    Almalki, Ziyad S.
    Alahmari, Abdullah K.
    LATIN AMERICAN JOURNAL OF PHARMACY, 2023, 42 (01): : 25 - 30
  • [46] Adverse Events Associated With 3 Vitrectomy Platforms Reported to the US FDA MAUDE Database
    Hwang, Hannah H.
    Botsford, Benjamin W.
    Lenis, Tamara L.
    Sadiq, Mohammad Ali
    Mahrous, Abdallah
    Orlin, Anton
    Kiss, Szilard
    D'Amico, Donald J.
    Van Tassel, Sarah H.
    Kovacs, Kyle D.
    JOURNAL OF VITREORETINAL DISEASES, 2024, 8 (05) : 546 - 553
  • [47] Endoscopic management of iatrogenic EUS-related duodenal perforations with over-the-scope clips
    El Bacha, Hicham
    Prat, Frederic
    ENDOSCOPY INTERNATIONAL OPEN, 2020, 8 (01) : E59 - E63
  • [48] What Is the Real Mode of Failure? Trends in StarClose Adverse Events Reported to the MAUDE Database
    Amendola, Michael F.
    Redifer, Kaila
    JOURNAL OF VASCULAR SURGERY, 2016, 63 (06) : 91S - 91S
  • [49] Clinical Adverse Events and Device Failures Reported for the Captivator and Duette Endoscopic Mucosal Resection (EMR) Systems A MAUDE Database Analysis
    Hagenstein, Lauren D.
    Dubrouskaya, Katsiaryna
    Ramai, Daryl
    Adler, Douglas G.
    JOURNAL OF CLINICAL GASTROENTEROLOGY, 2023, 57 (05) : 490 - 493
  • [50] Over-the-scope Clips better help sting Ulcer Bleeding
    Klein, Friederike
    ZEITSCHRIFT FUR GASTROENTEROLOGIE, 2019, 57 (01):