Motion-robust, multi-slice, real-time MR thermometry for MR-guided thermal therapy in abdominal organs

被引:3
|
作者
Kim, Kisoo [1 ,3 ]
Diederich, Chris [2 ]
Narsinh, Kazim [1 ]
Ozhinsky, Eugene [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA USA
[2] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA USA
[3] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, 2340 Sutter St,S341, San Francisco, CA 94115 USA
关键词
MR thermometry; abdominal organ; motion; real-time MR imaging; MR-guided thermal therapy; MAGNETIC-RESONANCE THERMOMETRY; INTENSITY FOCUSED ULTRASOUND; IN-VIVO; ABLATION; HYPERTHERMIA; REFERENCELESS; GUIDANCE;
D O I
10.1080/02656736.2022.2151649
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop an effective and practical reconstruction pipeline to achieve motion-robust, multi-slice, real-time MR thermometry for monitoring thermal therapy in abdominal organs. Methods: The application includes a fast spiral magnetic resonance imaging (MRI) pulse sequence and a real-time reconstruction pipeline based on multi-baseline proton resonance frequency shift (PRFS) method with visualization of temperature imaging. The pipeline supports multi-slice acquisition with minimal reconstruction lag. Simulations with a virtual motion phantom were performed to investigate the influence of the number of baselines and respiratory rate on the accuracy of temperature measurement. Phantom experiments with ultrasound heating were performed using a custom-made motion phantom to evaluate the performance of the pipeline. Lastly, experiments in healthy volunteers (N= 2) without heating were performed to evaluate the accuracy and stability of MR thermometry in abdominal organs (liver and kidney). Results: The multi-baseline approach with greater than 25 baselines resulted in minimal temperature errors in the simulation. Phantom experiments demonstrated a 713ms update time for 3-slice acquisitions. Temperature maps with 30 baselines showed clear temperature distributions caused by ultrasound heating in the respiratory phantom. Finally, the pipeline was evaluated with physiologic motions in healthy volunteers without heating, which demonstrated the accuracy (root mean square error [RMSE]) of 1.23 +/- 0.18 degrees C (liver) and 1.21 7 +/- 0.17 degrees C (kidney) and precision of 1.13 +/- 0.11 degrees C (liver) and 1.16 +/- 0.15 degrees C (kidney) using 32 baselines. Conclusions: The proposed real-time acquisition and reconstruction pipeline allows motion-robust, multi-slice, real-time temperature monitoring within the abdomen during free breathing.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] MR-guided joint puncture and real-time MR-assisted contrast media application
    Trattnig, S
    Breitenseher, M
    Pretterklieber, M
    Kontaxis, G
    Rand, T
    Helbich, T
    Imhof, H
    ACTA RADIOLOGICA, 1997, 38 (06) : 1047 - 1049
  • [22] Real-Time Motion Monitoring Using Orthogonal Cine MRI during MR-Guided Radiation Therapy for Liver Cancer
    Jassar, H.
    Tai, A.
    Chen, X.
    Paulson, E. S.
    Ahunbay, E. E.
    Omari, E.
    Lathuiliere, F.
    Beriault, S.
    Cooper, D.
    Hebert, F.
    Savard, L.
    Cloake, S.
    Straza, M. W., Jr.
    Awan, M.
    Hall, W. A.
    Li, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : S80 - S81
  • [23] Real-Time Motion Tracking Based on Orthogonal Cine MRI During MR-Guided Radiation Therapy for Prostate Cancer
    Jassar, H.
    Tai, A.
    Chen, X.
    Keiper, T.
    Paulson, E. S.
    Lathuiliere, F.
    Beriault, S.
    Cooper, D.
    Hebert, F.
    Savard, L.
    Cloake, S.
    Hall, W. A.
    Lawton, C. A. F.
    Li, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E541 - E541
  • [24] Real-time MR-guided joint puncture and arthrography: preliminary results
    Hilfiker, PR
    Weishaupt, D
    Schmid, M
    Dubno, B
    Hodler, J
    Debatin, JF
    EUROPEAN RADIOLOGY, 1999, 9 (02) : 201 - 204
  • [25] Real-time MR-guided joint puncture and arthrography: preliminary results
    Paul R. Hilfiker
    Dominik Weishaupt
    Marius Schmid
    Benjamin Dubno
    Jürg Hodler
    Jörg F. Debatin
    European Radiology, 1999, 9 : 201 - 204
  • [26] Real-time MR-guided catheterization of the coronary arteries in an animal model
    Buecker, A
    Spuentrup, E
    Pfeffer, J
    Schmitz-Rode, T
    Schaeffter, T
    Guenther, RW
    RADIOLOGY, 2001, 221 : 416 - 416
  • [27] Implementation of Real-Time MR-Guided Interstitial Brachytherapy for Gynecological Cancer
    Chan, K. Y.
    Simeonov, A.
    Di Tomasso, A.
    O'Leary, G.
    Filici, A. Lopez
    Rink, A.
    Beiki-Ardakani, A.
    Borg, J.
    Croke, J. M.
    Fyles, T. W.
    Han, K.
    Milosevic, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E284 - E284
  • [28] MR-Guided intravascular procedures:: Real-time parameter control and automated slice positioning with active tracking coils
    Bock, M
    Volz, S
    Zühlsdorff, S
    Umathum, R
    Fink, C
    Hallscheidt, P
    Semmler, W
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (05) : 580 - 589
  • [29] Real Time Motion Tracking for MR-guided Spine SBRT (Pilot Study)
    Ding, Yao
    Tom, Martin C.
    Yeboa, Debra N.
    Cifter, Gizem
    Briere, Tina M.
    Han, Eun Young
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4155 - S4157
  • [30] Real-time temperature estimation with enhanced spatial resolution during MR-guided hyperthermia therapy
    Pacheco, Cesar C.
    Orlande, Helcio R. B.
    Colaco, Marcelo J.
    Dulikravich, George S.
    Varon, Leonardo A. B.
    Lamien, Bernard
    NUMERICAL HEAT TRANSFER PART A-APPLICATIONS, 2020, 77 (08) : 782 - 806